Mectacer Biolox Delta Femoral Ball Head 01.29.202Recalls & reported problems
Medacta International SA · Femoral Head Ø 28 Size M · Class 2 · Implant
All models of Mectacer Biolox Delta Femoral Ball Head (15)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Medacta Biolox Delta femoral ball head is the ceramic sphere used in hip replacements. The 35 reports mostly describe infection at the surgical site, sometimes within weeks of surgery and sometimes years later, along with pain, dislocation of the head from its liner, and wear; 35 injuries and no deaths were reported. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 36
- Deaths reported
- 0
- Injuries reported
- 36
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 36 adverse event reports that list this exact device (DI 07630030809132), received between Nov 15, 2018 and Aug 3, 2026 — classified as 36 injuries.
The most frequently reported specific device problem is “Device Appears to Trigger Rejection” (14 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Appears to Trigger Rejection14
- Adverse Event Without Identified Device or Use Problem9
- Appropriate Device Problem Term/Code Not Available8
- Device Dislodged or Dislocated3
- Inadequacy of Device Shape and/or Size1
- Migration1
- Naturally Worn1
Patient problems
- Unspecified Infection15
- Medical device site infection10
- Pain5
- Unequal Limb Length2
- Hematoma1
- Implant Pain1
- Joint Dislocation1
- Joint Laxity1
Recent adverse event reports
The 25 most recent of 36 reports, in the reporter's own words as published by the FDA.
InjuryAug 3, 2026Device Appears to Trigger Rejection
Revision surgery at about 20 days from primary due to infection.The surgeon performed a washout and revised the head and liner. The surgery was completed successfully.
Patient problems: Medical device site infection
MAUDE report 50114125 ↗InjuryApr 15, 2026Device Appears to Trigger Rejection
REVISION SURGERY WAS PERFORMED 4 DAYS AFTER THE PRIMARY PROCEDURE DUE TO A DISCOLORED AND RASH-LIKE INCISION SITE OF UNKNOWN CAUSE. THE SURGEON PERFORMED AN IRRIGATION AND DEBRIDEMENT (I&D) AND REVISED THE HEAD AND LINER, REPLACING THEM WITH COMPONENTS OF THE SAME SIZE. THE SURGERY WAS SUCCESSFULLY COMPLETED.
Patient problems: Medical device site infection
MAUDE report 24882482 ↗InjuryDec 20, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
THE PATIENT HAD PRIMARY SURGERY WITH A COMPETITOR. ON (B)(6) 2025 THE PATIENT WAS REVISED FOR AN UNKNOWN REASON AND MEDACTA HARDWARE WAS IMPLANTED. ON (B)(6) 2025 THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN WAS UNKNOWN. THE SURGEON PERFORMED A WASHOUT, REMOVED ALL IMPLANTS, AND IMPLANTED NEW HARDWARE. THE SURGERY WAS COMPLETED SUCCESSFULLY (MDR 2025-01232). PRESENTLY, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE LINER AND HEAD TO A LINER AND HEAD OF THE SAME SIZE. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Medical device site infection
MAUDE report 23858477 ↗InjuryJul 25, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
AT ABOUT 2 WEEKS FROM PRIMARY, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE COMPETITOR CUP AND LINER AND MEDACTA PROXIMAL BODY AND HEAD. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Medical device site infection
MAUDE report 22607935 ↗InjuryJul 16, 2025Device Appears to Trigger Rejection; Migration; Device Appears to Trigger Rejection; Migration
AFTER 2 YEARS AND 11 MONTHS, THE PATIENT CAME IN REPORTING PAIN DUE TO A SUBSIDED STEM THAT WAS CAUSED FROM AN INFECTION. THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE STEM, HEAD AND LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Medical device site infection
MAUDE report 22518249 ↗InjuryJun 19, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
ON THE (B)(6) 2025, THE PATIENT UNDERWENT A HIP REVISION DUE TO FEMUR FRACTURE. COMPETITOR STEM WAS REMOVED, AMIS-K IMPLANTED AND CONTEXTUALLY MEDACTA DM CONVERTER IMPLANTED. AT ABOUT 1 MONTH FROM PREVIOUS REVISION, THE PATIENT RETURNED TO THE HOSPITAL REPORTING A DISLOCATION. AFTER AN ATTEMPT TO CLOSE-REDUCE THE DISLOCATION WITHOUT SURGERY, THE SURGEON DECIDED TO OPEN THE PATIENT AND FOUND THE HEAD DISLOCATED FROM THE DM LINER AND THE DM LINER DISLOCATED FROM THE DM CONVERTER. IT IS POSSIBLE THAT THE HEAD DISLOCATED FROM THE DM LINER DURING THE CLOSE-REDUCTION MANEUVER BUT THIS CANNOT BE CONFIRMED. SURGEON REVISED THE HEAD, LINER AND ALSO THE DM CONVERTER ALONG WITH THE CUP AS HE WAS NOT SATISFIED WITH ITS POSITION. KERBOUL CAGE AND VERCACEM IMPLANTED.
Patient problems: Joint Dislocation
MAUDE report 22288958 ↗InjuryMar 25, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
AT ABOUT 3 WEEKS FROM PRIMARY, THE PATIENT CAME IN DUE TO SIGNS OF INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE HEAD AND LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Medical device site infection
MAUDE report 21687570 ↗InjuryMar 5, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
THE PATIENT CAME IN DUE TO SIGNS OF INFECTION AND THE PATHOGEN IS UNKNOWN. ABOUT 1 YEAR AND 9 MONTHS AFTER THE PRIMARY SURGERY, THE SURGEON REVISED THE COMPETITOR CUP AND LINER AND MEDACTA HEAD. THE SURGERY WAS COMPLETED.
Patient problems: Medical device site infection
MAUDE report 21523850 ↗InjuryOct 23, 2024Naturally Worn; Naturally Worn
AT ABOUT 10 YEARS AND 1 MONTH AFTER PRIMARY, THE PATIENT CAME IN FOR A POST-OP APPOINTMENT AND THE HEAD AND LINER SEEMED TO BE WORN. THE CAUSES WERE UNKNOWN. THE SURGEON REVISED THE HEAD AND LINER AND THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20512204 ↗InjuryOct 8, 2024Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. ABOUT 1 MONTH AND A HALF AFTER THE PRIMARY SURGERY, THE SURGEON PERFORMED A WASHOUT AND REVISED THE HEAD AND LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Medical device site infection
MAUDE report 20396114 ↗InjuryMay 24, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT HAD HIP INSTABILITY DUE TO A LEG LENGTH DISCREPANCY AND THE CAUSE IS UNKNOWN. AT ABOUT 2 YEARS AND 1 MONTH POST PRIMARY THE SURGEON REVISED THE MEDACTA HEAD AND LINER WITH A NEW MEDACTA HEAD AND LINER AND REVISED THE MEDACTA CUP WITH A COMPETITOR CUP. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Pain; Unequal Limb Length
MAUDE report 19392538 ↗InjuryMar 8, 2024Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
REVISION DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. AT ABOUT 1 MONTH POST PRIMARY THE SURGEON PERFORMED A WASHOUT AND REVISED THE HEAD. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Unspecified Infection
MAUDE report 18865035 ↗InjuryDec 13, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
AT ABOUT 2 MONTHS FROM THE PRIMARY, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE HEAD AND LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Medical device site infection
MAUDE report 18316743 ↗InjuryOct 31, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AT ABOUT 6 MONTHS AFTER THE PRIMARY, THE PATIENT CAME IN REPORTING PAIN AND INSTABILITY DUE TO A LEG LENGTH DISCREPANCY. THE SURGEON REVISED THE HEAD AND LINER AND THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Unequal Limb Length
MAUDE report 18039859 ↗InjurySep 13, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. AT ABOUT 7 MONTHS FROM PRIMARY THE SURGEON PERFORMED A WASHOUT AND REVISED THE HEAD AND LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Medical device site infection
MAUDE report 17734946 ↗InjuryAug 10, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated
THE PATIENT CAME IN REPORTING PAIN DUE TO A DISLOCATION OF THE HEAD FROM THE LINER AND THE CAUSE IS UNKNOWN. ABOUT 2 WEEKS AFTER THE PRIMARY SURGERY, THE SURGEON REVISED ALL COMPONENTS AND THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Implant Pain
MAUDE report 17496768 ↗InjuryMay 10, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN WAS UNKNOWN. AT ABOUT 3 WEEKS POST PRIMARY THE SURGEON PERFORMED A WASHOUT AND REVISED THE HEAD AND LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Unspecified Infection
MAUDE report 16900765 ↗InjuryMay 9, 2023Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
THE PATIENT CAME IN REPORTING ANTERIOR GROIN PAIN AND THE CAUSE IS UNKNOWN. BONE SCAN WAS GOOD AND THE PATIENT WAS NEGATIVE FOR ANY INFECTION. THERE WAS A SOFT TISSUE MASS THAT APPEARED ON ONE OF THE SCANS. ABOUT 9 MONTHS AFTER THE PRIMARY SURGERY, THE SURGEON REVISED THE MEDACTA CUP AND LINER WITH A COMPETITOR CUP AND LINER AND REVISED THE MEDACTA HEAD WITH A MEDACTA HEAD. DURING SURGERY, IT WAS FOUND THAT THE ANTERIOR CAPSULE AND THE SURROUNDING TISSUE WERE VERY THICK AND LEATHERY. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Pain
MAUDE report 16893408 ↗InjuryApr 18, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. AT ABOUT 1 MONTH AFTER THE PRIMARY SURGERY, THE SURGEON PERFORMED A WASHOUT AND REVISED THE HEAD AND LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Unspecified Infection
MAUDE report 16761028 ↗InjuryMar 10, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
AT ABOUT 1 YEAR AND 4 MONTHS AFTER THE PRIMARY SURGERY, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED SUCCESSFULLY THE HEAD AND LINER.
Patient problems: Unspecified Infection
MAUDE report 16521455 ↗InjuryDec 15, 2022Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
AT ABOUT 7 YEARS 8 MONTHS AFTER THE PRIMARY, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE HEAD AND LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Unspecified Infection
MAUDE report 15987317 ↗InjurySep 16, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AT ABOUT 6 MONTHS AFTER THE PRIMARY SURGERY, THE PATIENT CAME IN REPORTING PAIN AND INSTABILITY DUE TO A LEG LENGTH DISCREPANCY. THE SURGEON REVISED THE HEAD AND LINER AND THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Joint Laxity
MAUDE report 15438966 ↗InjuryJun 1, 2022Appropriate Device Problem Term/Code Not Available
THE PATIENT CAME IN REPORTING PAIN DUE TO DEVELOPING A HEMATOMA AND THE CAUSE IS UNKNOWN. AT 1 MONTH AFTER THE PRIMARY SURGERY, THE SURGEON REVISED THE HEAD AND LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Hematoma
MAUDE report 14560052 ↗InjuryApr 6, 2022Appropriate Device Problem Term/Code Not Available
THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AFTER ABOUT 1 MONTH FROM THE PRIMARY SURGERY, THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE MEDACTA HEAD AND COMPETITOR LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY
Patient problems: Unspecified Infection
MAUDE report 14021621 ↗InjuryMar 19, 2022Appropriate Device Problem Term/Code Not Available
THE PATIENT HAD A PRIMARY HIP SURGERY ON (B)(6) 2019. ON (B)(6) 2021, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN WAS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE HEAD AND LINER. PRESENTLY, ON (B)(6) 2022, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON REMOVED ALL COMPONENTS, PERFORMED A WASHOUT, AND REIMPLANTED NEW COMPONENTS. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Unspecified Infection
MAUDE report 13826438 ↗
Other Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented devices
- BIOLOX DELTA 1365-32-310Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented551 reports
- PINNACLE ALTRX 1221-36-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented481 reports
- BIOLOX DELTA 1365-32-320Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death314 reports
- BIOLOX DELTA 1365-36-720Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death189 reports
- DJO SURGICAL 400-03-362Encore Medical, L.P. · Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented187 reports
- Anthology 71365705Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented176 reports
- BIOLOX DELTA 1365-40-710Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented170 reports
- PINNACLE GRIPTION 1217-32-054Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented169 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page