Skip to content

Mectacer Biolox Option Heads 01.29.233HRecalls & reported problems

Medacta International SA · Head Biolox Option ø 40 · Class 2 · Implant

All models of Mectacer Biolox Option Heads (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07630030809354), received between Apr 6, 2023 and Dec 9, 2025 — classified as 4 injuries.

The most frequently reported specific device problem is “Device Appears to Trigger Rejection” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Appears to Trigger Rejection4

Patient problems

  • Medical device site infection3
  • Unspecified Infection2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 9, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    AT ABOUT 1 YEAR FROM THE PRIMARY, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT, REMOVED ALL IMPLANTS, AND IMPLANTED NEW HARDWARE. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Medical device site infection

    MAUDE report 23758549 ↗
  • InjuryOct 28, 2024Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    THE PATIENT HAD A PRIMARY HIP SURGERY ON (B)(6) 2021. ON (B)(6) 2022, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN WAS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE HEAD AND THE SURGERY WAS COMPLETED SUCCESSFULLY. ON (B)(6) 2023, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON REVISED THE HEAD, SLEEVE AND LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY. PRESENTLY, ON (B)(6) 2024, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE HEAD AND LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Medical device site infection; Unspecified Infection

    MAUDE report 20544237 ↗
  • InjuryFeb 6, 2024Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    AT ABOUT 1 MONTH AFTER THE PRIMARY, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE STEM, HEAD AND LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Medical device site infection

    MAUDE report 18648765 ↗
  • InjuryApr 6, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    THE PATIENT HAD A PRIMARY HIP SURGERY ON (B)(6) 2019. ON (B)(6) 2022, THE PATIENT CAME IN REPORTING PAIN DUE TO A DISLOCATION OF THE HEAD FROM THE LINER AND THE CAUSE IS UNKNOWN. THE SURGEON REVISED SUCCESSFULLY THE HEAD AND LINER. ON (B)(6) 2022, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN WAS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED SUCCESSFULLY THE HEAD AND LINER. PRESENTLY, ON (B)(6) 2023 THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REMOVED ALL COMPONENTS AND IMPLANTED PERMANENT HARDWARE. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Unspecified Infection

    MAUDE report 16689565 ↗

Other Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented devices

All Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page