Adverse event reports
7
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 07630030810701), received between Jul 6, 2018 and Feb 6, 2025 — classified as 6 injuries, 1 malfunction.

The most frequently reported specific device problem is “Migration or Expulsion of Device” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available2
  • Migration or Expulsion of Device2
  • Device Appears to Trigger Rejection1
  • Device Dislodged or Dislocated1
  • Malposition of Device1
  • Osseointegration Problem1

Patient problems

  • Pain3
  • Failure of Implant2
  • Implant Pain1
  • Inadequate Osseointegration1
  • Insufficient Information1
  • Joint Dislocation1
  • No Code Available1
  • Unspecified Infection1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 6, 2025Malposition of Device; Malposition of Device

    AT ABOUT 1 YEAR AND 7 MONTHS FROM PRIMARY, THE PATIENT CAME IN REPORTING PAIN DUE TO A MALPOSITIONED CUP AND THE CAUSE IS UNKNOWN. THE SURGEON REVISED THE CUP AND LINER WITH COMPETITOR COMPONENTS AND REVISED THE MEDACTA HEAD WITH A MEDACTA HEAD. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Implant Pain

    MAUDE report 21320606 ↗
  • InjuryNov 16, 2023Osseointegration Problem; Appropriate Device Problem Term/Code Not Available; Osseointegration Problem; Appropriate Device Problem Term/Code Not Available

    REVISION SURGERY DUE TO CUP LOOSENING, AT ABOUT 7 YEARS 7 MONTHS FROM THE PRIMARY. THE SURGEON REVISED SUCCESSFULLY CUP, LINER AND HEAD. THE PATIENT EXPERIENCED PAIN AND SUBLUXING FEELING OF THE RIGHT HIP SINCE THE PRIMARY SURGERY IN 2016. EPISODES OF JOINT LUXATION NOT CONFIRMED. THE SURGEON REPORTED SIGNS OF THE STEM IMPINGEMENT ON THE CUP IN END OF RANGE HIP FLEXION.

    Patient problems: Inadequate Osseointegration

    MAUDE report 18146682 ↗
  • InjurySep 6, 2022Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    THE PATIENT HAD A PRIMARY KNEE SURGERY ON (B)(6) 2022. ON (B)(6) 2022, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE POLY AND THE SURGERY WAS COMPLETED SUCCESSFULLY. A COMPLAINT WAS FILLED. ON (B)(6) 2022, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT, REMOVED ALL IMPLANTS, AND IMPLANTED AN ANTIBIOTIC SPACER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Unspecified Infection

    MAUDE report 15361263 ↗
  • MalfunctionMay 21, 2021Migration or Expulsion of Device

    THE SURGEON NOTICED THAT THE CUP MIGRATED FROM THE PATIENT ACETABULAR AFTER THE FINAL REPOSITIONING DURING THE SURGERY. THE SURGEON REPLACED IT WITH A NEW MPACT MULTIHOLE CUP AND LINER. THE WAS 1 HOUR SURGERY TIME DELAY. TOTAL SURGERY TIME: 4 HOURS. THE SAME SIZE WAS USED FINALLY. IT IS UNKNOWN IF ADDITIONAL ANESTHESIA WAS NECESSARY.

    Patient problems: Insufficient Information

    MAUDE report 11862214 ↗
  • InjuryDec 10, 2019Appropriate Device Problem Term/Code Not Available

    THE IMPACT CUP WAS REMOVED BECAUSE OF THE PATIENT-REPORTED PAIN 2.5 YEARS AFTER PRIMARY SURGERY. DURING THE REVISION SURGERY, IT WAS CONFIRMED THAT THE CUP WAS PERFECTLY INTEGRATED, AND THE INFECTION SAMPLES WERE NEGATIVE.

    Patient problems: Pain

    MAUDE report 9443412 ↗
  • InjuryAug 30, 2018Migration or Expulsion of Device

    THE PATIENT CAME IN COMPLAINING OF PAIN 1 YEAR AND 9 MONTHS AFTER PRIMARY SURGERY. THE SURGEON DETERMINED THAT THE CUP HAD MIGRATED OVER TIME. THE CAUSE OF THE MIGRATION IS UNKNOWN. THE SURGEON REVISED THE CUP, LINER AND HEAD. THE SURGEON REVISED THE CUP AND LINER WITH ANOTHER COMPANY'S PRODUCT. THE SURGEON REVISED THE HEAD WITH ANOTHER MEDACTA HEAD. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Failure of Implant; No Code Available; Pain

    MAUDE report 7832924 ↗
  • InjuryJul 6, 2018Device Dislodged or Dislocated

    WE WERE INFORMED ON 08 JUNE 2018 THAT BEFORE PRIMARY SURGERY THE HIP WAS FOUND DISLOCATED: THE SURGEON WAS ABLE TO PUT THE HIP IN PLACE IN PRE-OP WITH LITTLE EFFORT. THEN, DURING THE PRIMARY SURGERY, AN MPACT 2 HOLE SHELL WAS USED, BUT NO SCREWS WERE USED SINCE THE DOCTOR DETERMINED IT FELT SOLID. THE NEXT DAY THE PATIENT CONTINUED WITH PAIN: THE SHELL WAS FOUND MISPLACED. ON (B)(6) 2018 WE WERE INFORMED THAT THE SURGEON REVISED THE CUP, HEAD, LINER AND STEM. THE SURGERY WAS COMPLETED SUCCESSFULLY

    Patient problems: Failure of Implant; Joint Dislocation; Pain

    MAUDE report 7666247 ↗

Other Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented devices

All Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented devices →