GMK Primary Total Knee System 02.07.0034RPRecalls & reported problems
Medacta International SA · Resurfacing Patella size 2 · Class 2 · Implant
All models of GMK Primary Total Knee System (99)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Medacta GMK Primary Total Knee System patella component (a resurfacing patella used in knee replacement) has 42 adverse event reports naming its identifier, with 41 injuries and no deaths. Reports mostly describe the implant loosening or the patella dislocating, leading to pain, instability, and revision surgery; a few mention infection or a nickel allergy. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 42
- Deaths reported
- 0
- Injuries reported
- 41
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 42 adverse event reports that list this exact device (DI 07630030815751), received between Nov 10, 2017 and Aug 3, 2026 — classified as 41 injuries, 1 malfunction.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (14 reports).
1 other recall cover devices cleared under the same FDA submission (K090988); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Dislodged or Dislocated14
- Loosening of Implant Not Related to Bone-Ingrowth9
- Appropriate Device Problem Term/Code Not Available8
- Adverse Event Without Identified Device or Use Problem7
- Loose or Intermittent Connection3
- Break1
- Device Appears to Trigger Rejection1
- Device Slipped1
Patient problems
- Joint Dislocation11
- Pain10
- Implant Pain4
- Joint Laxity4
- Unspecified Infection4
- Joint Disorder3
- Aseptic loosening2
- Hypersensitivity/Allergic reaction2
Recent adverse event reports
The 25 most recent of 42 reports, in the reporter's own words as published by the FDA.
InjuryAug 3, 2026Adverse Event Without Identified Device or Use Problem
Revision surgery due to patella maltracking and dislocation after about 1 year and 4 months from primary. All knee implants were revised using competitors implants. The primary was performed using MyKnee patient match cutting block.
Patient problems: Joint Dislocation
MAUDE report 50114124 ↗InjuryMay 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A MEDWATCH REPORT WAS ANONYMOUSLY SUBMITTED TO FDA. AFTER UNDERGOING TKA ON BOTH KNEES (ON (B)(6) 2022 RIGHT KNEE, ON (B)(6) 2022 LEFT KNEE), THE PATIENT EXPERIENCED SEVERE PAIN, STIFFNESS, INSTABILITY, SWELLING, AND INFLAMMATION. DESPITE TWO KNEE MANIPULATIONS AND NUMEROUS PHYSIOTHERAPY SESSIONS, THE SYMPTOMS PERSISTED. ANOTHER SURGEON EXAMINED THE PATIENT AND DISCOVERED THAT SHE WAS SEVERELY ALLERGIC TO THE IMPLANT MATERIALS, PARTICULARLY NICKEL, AND THAT THE IMPLANTS WERE LOOSE AND MALPOSITIONED. CONSEQUENTLY, A REVISION SURGERY WAS NECESSARY. THE PATIENT EXPRESSED STRONG DISAPPOINTMENT WITH THE QUALITY OF THE MEDACTA DEVICES AND WITH THE LACK OF SUPPORT RECEIVED FROM THE COMPANY, DESPITE REPEATED ATTEMPTS TO CONTACT THEM VIA PHONE AND MAIL. THIS CASE IS RELATED TO THE LEFT KNEE, WHILE ANOTHER COMPLAINT HAS BEEN OPENED AND REFERS TO THE RIGHT KNEE. MEDWATCH CODE: MW5168966.
Patient problems: Aseptic loosening; Hypersensitivity/Allergic reaction; Implant Pain; Inflammation
MAUDE report 22051946 ↗InjuryMay 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A MEDWATCH REPORT WAS ANONYMOUSLY SUBMITTED TO FDA. AFTER UNDERGOING TKA ON BOTH KNEES ((B)(6) 2022 RIGHT KNEE, (B)(6) 2022 LEFT KNEE), THE PATIENT EXPERIENCED SEVERE PAIN, STIFFNESS, INSTABILITY, SWELLING, AND INFLAMMATION. DESPITE TWO KNEE MANIPULATIONS AND NUMEROUS PHYSIOTHERAPY SESSIONS, THE SYMPTOMS PERSISTED. ANOTHER SURGEON EXAMINED THE PATIENT AND DISCOVERED THAT SHE WAS SEVERELY ALLERGIC TO THE IMPLANT MATERIALS, PARTICULARLY NICKEL, AND THAT THE IMPLANTS WERE LOOSE AND MALPOSITIONED. CONSEQUENTLY, A REVISION SURGERY WAS NECESSARY. THE PATIENT EXPRESSED STRONG DISAPPOINTMENT WITH THE QUALITY OF THE MEDACTA DEVICES AND WITH THE LACK OF SUPPORT RECEIVED FROM THE COMPANY, DESPITE REPEATED ATTEMPTS TO CONTACT THEM VIA PHONE AND MAIL. THIS CASE IS RELATED TO THE RIGHT KNEE, WHILE THERE IS ANOTHER COMPLAINT RELATED TO THE LEFT KNEE. MEDWATCH CODE: MW5168965.
Patient problems: Aseptic loosening; Hypersensitivity/Allergic reaction; Implant Pain; Inflammation
MAUDE report 22051945 ↗InjuryApr 30, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT CAME IN REPORTING STIFFNESS AT ABOUT 5 MONTHS AFTER THE PRIMARY. REHABILITATION HAS BEEN PERFORMED IN ORDER TO RESTORE THE JOINT MOBILITY.
Patient problems: Loss of Range of Motion
MAUDE report 21933168 ↗InjuryDec 11, 2024Device Dislodged or Dislocated; Adverse Event Without Identified Device or Use Problem; Device Dislodged or Dislocated; Adverse Event Without Identified Device or Use Problem
THE PATIENT HAD A PRIMARY KNEE SURGERY ON (B)(6) 2017. ON (B)(6) 2020 THE PATIENT CAME IN REPORTING PAIN DUE TO A LOOSE FEMUR THAT WAS CAUSED FROM FALLING. THE SURGEON REVISED THE FEMORAL COMPONENT AND LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY. PRESENTLY, ON (B)(6) 2024, THE PATIENT CAME IN REPORTING INSTABILITY AND THE CAUSE IS UNKNOWN. THE REAL ISSUE WAS THE PATELLA DISLOCATING LATERAL AND HE WAS FOCUSED TO ADDRESS THAT. HE CUT THE PATELLA OUT AND DIDN'T PUT ANOTHER PATELLA IN. THE SURGEON ADDRESSED THE PATELLA TRACKING ISSUES WITH SUTURE AND TISSUE TIGHTENING AND RE-ALIGNMENT AND REVISED THE LINER WITH ONE OF THE SAME THICKNESS. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Joint Laxity
MAUDE report 20902528 ↗InjurySep 19, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
THE PATIENT HAD ANTERIOR KNEE PAIN AND THE CAUSE IS UNKNOWN. AT ABOUT 1 MONTH POST PRIMARY THE SURGEON REVISED THE PATIENT'S PATELLA AND REVISED THE INSERT. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Pain
MAUDE report 20254225 ↗InjuryJun 25, 2024Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
AT ABOUT 1 YEAR AND 3 WEEKS AFTER PRIMARY, THE PATIENT CAME IN DUE TO SIGNS OF INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON REVISED ALL MEDACTA HARDWARE AND REIMPLANTED PERMANENT PRODUCT. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Unspecified Infection
MAUDE report 19605508 ↗InjuryMar 14, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
REVISION FOR PATELLA DISLOCATION. AT 5 MONTHS POST PRIMARY PATELLA AND LINER REVISED. NO KNEE INSTABILITY. THE PATIENT HAD A CLINICAL HISTORY OF PATELLA MALTRACKING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18901047 ↗InjuryJan 10, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AT ABOUT 9 MONTHS AFTER THE PRIMARY, THE PATIENT CAME IN REPORTING PAIN DUE TO THE PATELLA IMPLANT MALTRACKING WITH IMPINGEMENT ON THE LATERAL SIDE OF THE FEMORAL IMPLANT. THE SURGEON REVISED THE FEMORAL COMPONENT AND INSERT AND THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Pain
MAUDE report 18482243 ↗InjuryMay 26, 2023Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
AT ABOUT 1 YEAR 11 MONTHS AFTER THE PRIMARY, THE PATIENT CAME IN REPORTING PAIN DUE TO A LOOSE PATELLA AND THE CAUSE OF THE LOOSE PATELLA IS UNKNOWN. THE SURGEON REVISED THE PATELLA AND THE INSERT AS PER STANDARD PROCEDURE. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Pain
MAUDE report 17009821 ↗InjuryMay 24, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AT ABOUT 1 YEAR AND 1 MONTH AFTER THE PRIMARY SURGERY, THE PATIENT CAME IN REPORTING PATELLA PAIN AND THE CAUSE WAS UNKNOWN. THE SURGEON REVISED THE PATELLA IMPLANT AND IMPLANTED A THICKER POLY (FROM 11MM TO 13MM). THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Implant Pain; Joint Laxity
MAUDE report 16991235 ↗InjuryFeb 10, 2023Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
AT ABOUT 3 YEARS AND 3 MONTHS AFTER THE PRIMARY SURGERY, THE PATIENT CAME IN REPORTING PAIN AND RADIOLUCENCY WAS NOTICED AROUND THE PATELLA. THE SURGEON REVISED THE PATELLA AND FOUND IT WAS LOOSE. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Implant Pain
MAUDE report 16349596 ↗InjuryJan 4, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated
AT ABOUT 1 MONTH AFTER THE PRIMARY, THE PATIENT CAME IN REPORTING PAIN DUE TO A PATELLAR DISLOCATION AND THE CAUSE IS UNKNOWN. THE SURGEON REVISED THE LINER, FEMUR AND PATELLA. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Joint Laxity
MAUDE report 16093804 ↗InjuryDec 16, 2022Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
REVISION SURGERY FOR PATELLA LOOSENING AT ABOUT 1 YEAR AND 3 MONTHS POST PRIMARY. THE SURGEON REVISED ALSO THE LINER FROM 12 MM TO 14 MM THICKNESS TO RESTORE THE KNEE STABILITY.
Patient problems: Joint Laxity
MAUDE report 15994324 ↗InjuryAug 24, 2022Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
AT ABOUT 1 YEAR AND 9 MONTH AFTER PRIMARY, REVISION SURGERY FOR PAIN WAS PERFORMED. DURING THE REVISION THE PATELLA IMPLANT WAS FOUND TO BE LOOSE. SURGERY COMPLETED SUCCESSFULLY BY SWAPPING PATELLA AND BY IMPLANTING A THICKER LINER.
Patient problems: Pain
MAUDE report 15282565 ↗InjuryJul 28, 2022Loosening of Implant Not Related to Bone-Ingrowth
THE PATIENT CAME IN REPORTING PAIN DUE TO A LOOSE PATELLA AND THE CAUSE OF THE LOOSE PATELLA IS UNKNOWN. AT ABOUT 2 YEARS FORM PRIMARY THE SURGEON REVISED THE PATELLA AND INSERT. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Pain
MAUDE report 15121612 ↗InjuryJul 18, 2022Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
AT ABOUT 3 YEARS AND 3 MONTHS POST PRIMARY THE SURGEON REVISED THE PATIENT KNEE FOR PATELLA LOOSENING. THE SURGERY WAS COMPLETED SUCCESSFULLY. IT WAS ALSO NOTED A PERIPHERAL METALLOSIS AND RISK OF THE PREMATURE WEAR OF THE FEMUR DUE TO THE PATELLA'S DISPLACEMENT.
Patient problems: Metal Related Pathology
MAUDE report 15037296 ↗InjuryJul 7, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated
AFTER THE PRIMARY SURGERY, IT WAS DISCOVERED THE PATELLA DISLOCATION FROM THE FEMUR. AT ABOUT 3 WEEKS AFTER THE PRIMARY SURGERY, THE SURGEON REVISED ALL THE IMPLANTS.
Patient problems: Joint Dislocation
MAUDE report 14942802 ↗InjuryJun 17, 2022Break; Material Rupture
THE PATIENT HAD A PRIMARY KNEE SURGERY ON (B)(6) 2021. ON (B)(6) 2021, THE PATIENT CAME IN REPORTING PAIN DUE TO FALLING AND RUPTURING THE PATELLA TENDON AND DAMAGING THE COLLATERAL LIGAMENTS. THE SURGEON REVISED ALL IMPLANTS EXCEPT THE PATELLA. THE SURGERY WAS COMPLETED SUCCESSFULLY. ON 18 MAY 2022, THE PATIENT CAME IN REPORTING PAIN DUE TO A PATELLA TENDON RUPTURE AND THE CAUSE IS UNKNOWN. THE SURGEON REMOVED THE PATELLA IMPLANT AND CLOSED THE PATIENT UP WITHOUT A IMPLANTING A NEW PATELLA IMPLANT. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Fall; Pain
MAUDE report 14733635 ↗InjuryApr 4, 2022Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
THE PATIENT CAME IN REPORTING PAIN DUE TO A LOOSE PATELLA AND THE CAUSE OF THE LOOSE PATELLA IS UNKNOWN. AT 4 YEARS AND 10 MONTHS AFTER THE PRIMARY SURGERY, THE SURGEON REVISED THE PATELLA AND THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Pain
MAUDE report 13995320 ↗InjuryNov 25, 2021Loose or Intermittent Connection; Appropriate Device Problem Term/Code Not Available
THE PATIENT CAME IN REPORTING PAIN DUE TO A LOOSE PATELLA AND THE CAUSE OF THE LOOSE PATELLA IS UNKNOWN. AT 6 YEARS AND 3 MONTHS AFTER THE PRIMARY SURGERY, THE SURGEON REVISED THE PATELLA AND THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Pain
MAUDE report 12874309 ↗InjurySep 23, 2021Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
THE PATIENT CAME IN REPORTING PAIN DUE TO A LOOSE PATELLA. THE CAUSE OF THE LOOSE PATELLA IS UNKNOWN. 3 YEARS AND 10 MONTHS AFTER THE PRIMARY SURGERY, THE SURGEON REVISED THE PATELLA AND POLY AND THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Pain
MAUDE report 12517835 ↗InjuryJun 2, 2021Appropriate Device Problem Term/Code Not Available
DISLOCATION OF THE RESURFACING PATELLA AFTER PRIMARY SURGERY PERFORMED ON (B)(6) 2020. REVISION SURGERY HAS NOT BEEN PLANNED YET.
Patient problems: Joint Dislocation
MAUDE report 11919335 ↗MalfunctionJun 2, 2021Device Dislodged or Dislocated; Appropriate Device Problem Term/Code Not Available
DISLOCATION OF THE RESURFACING PATELLA AFTER PRIMARY SURGERY PERFORMED ON (B)(6) 2021. REVISION SURGERY HAS NOT BEEN PLANNED YET.
Patient problems: Joint Dislocation
MAUDE report 11919187 ↗InjuryMar 25, 2021Appropriate Device Problem Term/Code Not Available
THE PATIENT CAME IN 1 YEAR AND 2 MONTHS AFTER THE PRIMARY SURGERY DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED ALL COMPONENTS WITH COMPETITOR HARDWARE. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Unspecified Infection
MAUDE report 11568009 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K090988. One submission often covers many models, so check the product and lot details in each record.
- Z-0266-2018 · Oct 24, 2017 · TerminatedThe GMK knee system: designed for cemented use in total knee arthroplasty. Sphere and Primary Inserts with part number ranges: 02.12.0110FL - 02.12.0620FL; 02.12.0110FR - 02.12.0
Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page