GMK Primary Total Knee System 02.07.0210PSFRecalls & reported problems
Medacta International SA · Tibial insert PS fixed 10mm, size 2 · Class 2 · Implant
All models of GMK Primary Total Knee System (99)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07630030816482), received between Dec 19, 2018 and Feb 8, 2021 — classified as 3 injuries.
The most frequently reported specific device problem is “Device Slipped” (1 report).
1 other recall cover devices cleared under the same FDA submission (K090988); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Appropriate Device Problem Term/Code Not Available1
- Device Slipped1
- Loose or Intermittent Connection1
- Material Fragmentation1
- Unstable1
Patient problems
- Joint Laxity1
- No Code Available1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryFeb 8, 2021Loose or Intermittent Connection; Appropriate Device Problem Term/Code Not Available
THE PATIENT CAME IN AFTER 5 YEARS AND 8 MONTHS FROM THE PRIMARY SURGERY REPORTING PAIN AND INSTABILITY. THE CAUSE OF THE PAIN WAS A LOOSE PATELLA AND THE CAUSE OF LOOSE PATELLA WAS LINER INSTABILITY. THE SURGEON REVISED THE INSERT (TO A THICKER ONE) AND PATELLA SUCCESSFULLY.
Patient problems: Joint Laxity
MAUDE report 11292054 ↗InjuryMar 27, 2020Material Fragmentation; Material Fragmentation
INLAY FIXATION SCREW BREAKAGE DURING PRIMARY SURGERY, THE SCREW BODY REMAINED IN THE TIBIAL TRAY. THE TORQUE-LIMITER SCREWDRIVER USED WAS THE CORRECT ONE. INLAY PS SUBSTITUTED WITH UC THAT NOT NEED THE SCREW.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9887601 ↗InjuryDec 19, 2018Device Slipped; Unstable
THE PATIENT CAME IN DUE TO INSTABILITY, 4 YEARS AND 7 MONTHS AFTER PRIMARY SURGERY, CAUSED BY THE SCREW BACKING OUT OF THE INSERT. THE TORQUE LIMITING DRIVER WAS NOT USED IN THE PRIMARY SURGERY. THE SURGEON REVISED THE INSERT AND THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: No Code Available
MAUDE report 8178707 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K090988. One submission often covers many models, so check the product and lot details in each record.
- Z-0266-2018 · Oct 24, 2017 · TerminatedThe GMK knee system: designed for cemented use in total knee arthroplasty. Sphere and Primary Inserts with part number ranges: 02.12.0110FL - 02.12.0620FL; 02.12.0110FR - 02.12.0
Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page