GMK Primary Total Knee System 02.07.0412PSFRecalls & reported problems

Medacta International SA · Tibial insert PS fixed 12mm, size 4 · Class 2 · Implant

All models of GMK Primary Total Knee System (99)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
9
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 07630030817748), received between Aug 30, 2017 and Jul 16, 2025 — classified as 9 injuries, 1 malfunction.

The most frequently reported specific device problem is “Detachment Of Device Component” (2 reports).

1 other recall cover devices cleared under the same FDA submission (K090988); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem3
  • Appropriate Device Problem Term/Code Not Available2
  • Detachment Of Device Component2
  • Unstable2
  • Device Operates Differently Than Expected1
  • Inadequacy of Device Shape and/or Size1
  • Loose or Intermittent Connection1
  • Patient-Device Incompatibility1

Patient problems

  • Pain3
  • Joint Disorder2
  • Bone Fracture(s)1
  • Failure of Implant1
  • Joint Laxity1
  • Muscle/Tendon Damage1
  • No Known Impact Or Consequence To Patient1
  • Unspecified Infection1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 16, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AFTER 11 YEARS AND 6 MONTHS FROM THE PRIMARY, THE PATIENT CAME IN REPORTING PAIN AND THE CAUSE OF THE PAIN IS UNKNOWN. THE SURGEON REVISED TO A THICKER POLY AND THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Joint Laxity

    MAUDE report 22518829 ↗
  • InjuryApr 9, 2021Patient-Device Incompatibility; Appropriate Device Problem Term/Code Not Available

    THE PATIENT HAD A PRIMARY SURGERY AND WAS IMPLANTED WITH COMPETITOR IMPLANTS. ON (B)(6) 2021, THE PATIENT HAD THE COMPETITOR IMPLANTS REVISED TO MEDACTA IMPLANTS FOR REASONS UNKNOWN. PRESENTLY, ON (B)(6) 2021, THE PATIENT CAME IN REPORTING PAIN DUE TO A TORN PATELLA TENDON. THE CAUSE OF THE TORN PATELLA TENDON IS UNKNOWN. THE SURGEON REPAIRED THE PATELLA TENDON AND REVISED THE INSERT P.S. 12MM S4 WITH A INSERT SEMICONSTRAINED 12MM S4. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Muscle/Tendon Damage

    MAUDE report 11640108 ↗
  • MalfunctionDec 21, 2020Unstable

    REVISION SURGERY PERFORMED DUE TO KNEE INSTABILITY.

    Patient problems: Failure of Implant

    MAUDE report 11046850 ↗
  • InjuryFeb 28, 2020Adverse Event Without Identified Device or Use Problem

    TWO YEARS AFTER THE PREVIOUS REVISION SURGERY, THE PATIENT CAME IN REPORTING PAIN. THE CAUSE OF THE PAIN IS UNKNOWN. THE SURGEON REVISED ALL THE COMPONENTS. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Pain

    MAUDE report 9769048 ↗
  • InjuryFeb 13, 2020Adverse Event Without Identified Device or Use Problem

    THE PATIENT CAME IN, 7 YEARS AFTER PRIMARY SURGERY, DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PLANS TO PERFORM A WASHOUT AND POLY-SWAP. DURING SURGERY IT WAS DISCOVERED THAT THE KNEE WAS WELL FIXED AND WELL BALANCED SO THERE WAS NO NEED TO SWAP THE LINER. IT TURNS OUT THE PATIENT FRACTURED HIS PATELLA SO THE DOCTOR FIXED THAT AND SAID EVERYTHING ELSE WAS GOOD. INFECTION WAS NOT CONFIRMED AND THEY DID NOT REMOVE ANY IMPLANTS AT ALL.

    Patient problems: Bone Fracture(s); Unspecified Infection

    MAUDE report 9706225 ↗
  • InjurySep 26, 2019Appropriate Device Problem Term/Code Not Available

    THE PATIENT CAME IN, 1 YEAR AND 8 MONTHS AFTER PRIMARY SURGERY, COMPLAINING OF INSTABILITY AND THE CAUSE OF THE INSTABILITY IS UNKNOWN. THE SURGEON REVISED THE MEDACTA INSERT 12MM S4 WITH A MEDACTA INSERT 14MM S4. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Joint Disorder

    MAUDE report 9122746 ↗
  • InjuryMar 28, 2019Unstable

    ON (B)(6) WE WERE INFORMED ABOUT THE REVISION SURGERY PERFORMED THE DAY AFTER DUE TO KNEE INSTABILITY (6 YEARS AND 5 MONTHS AFTER PRIMARY), 12 MM PS INLAY HAS BEEN REVISED WITH A 17MM PS INLAY. AFTER THE REVISION THE KNEE WAS STABLE.

    Patient problems: Joint Disorder

    MAUDE report 8460367 ↗
  • InjuryMay 25, 2018Detachment Of Device Component; Loose or Intermittent Connection; Device Operates Differently Than Expected

    ON (B)(6) 2018 WE WERE INFORMED ABOUT A LOOSE INLAY-FIXATION SCREW OF A PRIMARY-PS KNEE; THE PATIENT HAD A TRAUMATIC EVENT. IT IS NOT KNOW IF THE SCREW WAS LOOSE BEFORE THE TRAUMATIC EVENT. REVISION SURGERY PERFORMED ON (B)(6) 2018 DURING SURGERY THEY RECOGNIZED THAT THE KNEE WAS VERY UNSTABLE AND THE LATERAL COLLATERAL LIGAMENT WAS COMPLETELY MISSING; IT IS UNKNOWN IF IT WAS MISSING BEFORE OR IF IT WAS RELATED TO THE TRAUMA. ALSO THE TIBIA PLATEAU WAS SUNKEN. THEY FINALLY IMPLANTED A GMK-HINGE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7544738 ↗
  • InjuryMay 17, 2018Detachment Of Device Component

    THE PATIENT CAME IN COMPLAINING OF PAIN 2 YEARS AND 7 MONTHS AFTER PRIMARY. THE SCREW DISASSOCIATED FROM THE POLY INSERT. THE SURGEON REMOVED THE SCREW ARTHROSCOPICALLY. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Pain

    MAUDE report 7522287 ↗
  • InjuryAug 30, 2017Inadequacy of Device Shape and/or Size

    THE PATIENT CAME IN COMPLAINING OF KNEE PAIN. THE SURGEON SWAPPED THE POLY FROM A SIZE 4 12 MM TO A SIZE 4 10 MM. THE SURGERY WAS COMPLETED SUCCESSFULLY. X-RAYS ARE AVAILABLE. EXPLANTS ARE NOT AVAILABLE.

    Patient problems: Pain

    MAUDE report 6829704 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K090988. One submission often covers many models, so check the product and lot details in each record.

  • Z-0266-2018 · Oct 24, 2017 · Terminated
    The GMK knee system: designed for cemented use in total knee arthroplasty. Sphere and Primary Inserts with part number ranges: 02.12.0110FL - 02.12.0620FL; 02.12.0110FR - 02.12.0

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →