GMK Primary Total Knee System 02.07.1201LRecalls & reported problems
Medacta International SA · Fixed tibial tray cemented left, size 1 · Class 2 · Implant
All models of GMK Primary Total Knee System (99)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 07630030819841), received between Jan 4, 2018 and May 13, 2026 — classified as 5 injuries, 1 malfunction.
The most frequently reported specific device problem is “Device Appears to Trigger Rejection” (1 report).
1 other recall cover devices cleared under the same FDA submission (K090988); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Appears to Trigger Rejection1
- Device Slipped1
- Implant subsidence1
- Migration1
- Migration or Expulsion of Device1
- No Apparent Adverse Event1
Patient problems
- Pain2
- Joint Disorder1
- Joint Laxity1
- No Clinical Signs, Symptoms or Conditions1
- Unspecified Infection1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 13, 2026No Apparent Adverse Event
THE SURGEON EXPERIENCED DIFFICULTY INSERTING THE EFFICIENCY 10 MM LEFT INSERT INTO A SIZE 1 LEFT CEMENTED TIBIAL BASEPLATE, REQUIRING MULTIPLE ATTEMPTS. THE PATIENT ACHIEVED FULL EXTENSION WITH A ROM OF 120°, WITH NO TIGHTNESS ISSUES. DIFFICULTY WAS ALSO ENCOUNTERED WHEN INSERTING THE DEFINITIVE 10 MM POLYETHYLENE INSERT, LEADING TO DAMAGE TO ITS POSTERIOR ASPECT AFTER REPEATED ATTEMPTS. AN 11 MM INSERT WAS THEN TRIED, ALSO WITH MULTIPLE ATTEMPTS, AND WAS ULTIMATELY FORCEFULLY IMPACTED USING A TIBIAL IMPACTOR. THE PROCEDURE TIME WAS EXTENDED BY APPROXIMATELY 45 MINUTES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25160337 ↗InjuryJul 31, 2025Implant subsidence; Implant subsidence
6 MONTHS AFTER THE PRIMARY SURGERY, THE PATIENT PRESENTED WITH PAIN DUE TO COLLAPSE OF THE LATERAL TIBIAL BONE. DURING THE PRIMARY SURGERY, THE TIBIAL COMPONENT HAD BEEN SLIGHTLY UNDERSIZED, AND THE PATIENT HAD SOFT BONE. THE SURGEON REVISED THE TIBIAL TRAY AND INSERT, AND IMPLANTED A STEM EXTENSION ALONG WITH TIBIAL AUGMENTS. THE REVISION SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Pain
MAUDE report 22673493 ↗InjuryMay 23, 2024Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
AT ABOUT 1 YEAR 3 MONTHS AFTER THE PRIMARY, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON REMOVED ALL COMPONENTS AND IMPLANTED AN ANTIBIOTIC SPACER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Unspecified Infection
MAUDE report 19381855 ↗InjuryFeb 23, 2024Migration; Migration
ABOUT 7 YEARS AND 8 MONTHS AFTER THE PRIMARY SURGERY, REVISION SURGERY HAS BEEN PERFORMED DUE TO TIBIAL SUBSIDENCE.
Patient problems: Pain
MAUDE report 18765773 ↗InjuryDec 15, 2023Migration or Expulsion of Device; Migration or Expulsion of Device
REVISION SURGERY DUE TO TIBIAL TRAY FIXED CEMENTED SIZE 1 SUBSIDENCE AND KNEE INSTABILITY, AT ABOUT 2 YEARS 10 MONTHS AFTER THE PRIMARY. ALL COMPONENTS SUCCESSFULLY REVISED WITH COMPETITOR'S SYSTEM.
Patient problems: Joint Laxity
MAUDE report 18334828 ↗InjuryJan 4, 2018Device Slipped
THE PATIENT CAME IN COMPLAINING OF PAIN. THE SURGEON DETERMINED THE TIBIAL BASEPLATE HAD SUBSIDED. THE SURGEON REVISED THE TIBIAL BASEPLATE AND THE POLY. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Joint Disorder
MAUDE report 7163357 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K090988. One submission often covers many models, so check the product and lot details in each record.
- Z-0266-2018 · Oct 24, 2017 · TerminatedThe GMK knee system: designed for cemented use in total knee arthroplasty. Sphere and Primary Inserts with part number ranges: 02.12.0110FL - 02.12.0620FL; 02.12.0110FR - 02.12.0
Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page