GMK Primary Total Knee System 02.07.1202RRecalls & reported problems
Medacta International SA · Fixed tibial tray cemented right, size 2 · Class 2 · Implant
All models of GMK Primary Total Knee System (99)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Medacta GMK Primary knee implant, a cemented right tibial tray used in total knee replacement, appears in 61 adverse event reports with 60 injuries and no deaths. Most reports describe the implant loosening or sinking, sometimes with pain, joint instability or infection, and many patients needing a second (revision) surgery to remove or replace parts. Reports rose from 4 in the prior 12 months to 10 in the most recent 12 months. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 61
- Deaths reported
- 0
- Injuries reported
- 60
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 61 adverse event reports that list this exact device (DI 07630030819872), received between Aug 3, 2017 and Aug 13, 2026 — classified as 60 injuries, 1 malfunction.
The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (17 reports).
1 other recall cover devices cleared under the same FDA submission (K090988); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Loosening of Implant Not Related to Bone-Ingrowth17
- Appropriate Device Problem Term/Code Not Available12
- Adverse Event Without Identified Device or Use Problem8
- Loose or Intermittent Connection5
- Migration5
- Device Appears to Trigger Rejection4
- Device Slipped3
- Implant subsidence2
Patient problems
- Pain10
- Joint Disorder8
- Joint Laxity7
- Implant Pain6
- No Clinical Signs, Symptoms or Conditions5
- Unspecified Infection5
- Aseptic loosening3
- Bone Fracture(s)3
Recent adverse event reports
The 25 most recent of 61 reports, in the reporter's own words as published by the FDA.
InjuryAug 13, 2026Adverse Event Without Identified Device or Use Problem
Revision surgery after 2 years and 8 months from the primary due to pain.The surgeon revised the tibial tray, insert, femoral component, and patella from GMK-Sphere to GMK-Revision. The surgery was completed successfully.
Patient problems: Implant Pain
MAUDE report 50232825 ↗InjuryJul 22, 2026Device Appears to Trigger Rejection
REVISION SURGERY WAS PERFORMED 1 MONTH AND HALF AFTER THE PRIMARY DUE TO INFECTION. THE SURGEON REMOVED ALL IMPLANTS AND PLACED ANTIBIOTIC SPACER. DURING THE PRIMARY SURGICAL PROCEDURE THE MEDACTA REPRESENTATIVE EXPERIENCED A MEDICAL EVENT, LOST CONSCIOUSNESS, AND FELL INTO THE OPEN SURGICAL FIELD ON THE SURGICAL DRAPS BEFORE COLLAPSING ONTO THE FLOOR. FOLLOWING THE EVENT, THE SURGEON INITIATED THE DISINFECTION PROTOCOL DUE TO THE POTENTIAL CONTAMINATION RISK OF THE SURGICAL SITE.
Patient problems: Medical device site infection
MAUDE report 25904707 ↗InjuryJul 8, 2026Loosening of Implant Not Related to Bone-Ingrowth
THE PATIENT UNDERWENT PRIMARY SURGERY ON (B)(6) 2025. APPROXIMATELY ONE YEAR AFTER SURGERY, THE PATIENT EXPERIENCED PAIN IN THE JOINT, AND X-RAYS CONFIRMED TIBIAL LOOSENING. A REVISION SURGERY IS SCHEDULED FOR (B)(6), 2026.
Patient problems: Aseptic loosening
MAUDE report 25747736 ↗InjuryJun 24, 2026Loosening of Implant Not Related to Bone-Ingrowth
REVISION SURGERY DUE TO TIBIAL TRAY LOOSENING AFTER ABOUT 10 YEARS AND 1 MONTH FROM PRIMARY.
Patient problems: Aseptic loosening
MAUDE report 25607907 ↗InjuryMay 28, 2026Implant subsidence
REVISION SURGERY DUE TO TIBIAL TRAY SUBSIDENCE AFTER ABOUT 1 YEAR AND 2 MONTHS FROM PRIMARY. THE SURGEON REVISED THE CEMENTED S2 TIBIAL TRAY TO A FIXED TINBN COATED SIZE 2 TIBIAL TRAY, REVISED THE FLEX 2R - 11MM INSERT TO A FLEX 2R - 10MM INSERT, REVISED THE S3+R CEMENTED FEMORAL COMPONENT TO A S3+R CEMENTED TINBN FEMORAL COMPONENT, AND REVISED THE MOTO ECROSS D 29 TO A SAME SIZE PATELLA. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Implant Pain
MAUDE report 25313930 ↗InjuryMar 3, 2026Adverse Event Without Identified Device or Use Problem
AT ABOUT 1 YEAR 4 MONTHS AFTER THE PRIMARY, THE PATIENT CAME IN PRESENTING INSTABILITY IN FLEXION AND THE CAUSE IS UNKNOWN. THE SURGEON REVISED ALL COMPONENTS FROM GMK-SPHERIKA COMPONENTS TO GMK-REVISION COMPONENTS, EXCEPT THE PATELLA. THE 10MM LINER WAS REVISED TO A SEMI-CONSTRAINED 17MM LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Implant Pain; Joint Laxity
MAUDE report 24494896 ↗InjuryFeb 26, 2026Adverse Event Without Identified Device or Use Problem
AT ABOUT 1 YEAR 3 MONTHS AFTER THE PRIMARY, THE PATIENT CAME IN DUE TO STIFFNESS AND THE CAUSE IS UNKNOWN. THE SURGEON REVISED THE TIBIAL TRAY, INSERT, AND FEMORAL COMPONENT FROM GMK-SPHERE TO GMK-REVISION. THE SURGEON ALSO REVISED THE PATELLA TO A SAME SIZE PATELLA. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Loss of Range of Motion
MAUDE report 24454901 ↗InjuryJan 8, 2026Adverse Event Without Identified Device or Use Problem
AT4 YEARS 1 MONTH FROM PRIMARY, THE PATIENT CAME IN REPORTING INSTABILITY AND THE CAUSE IS UNKNOWN. THE SURGEON REVISED ALL COMPONENTS TO HINGE COMPONENTS. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Implant Pain; Joint Laxity
MAUDE report 24004375 ↗InjuryNov 26, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AT ABOUT 11 MONTHS FROM THE PRIMARY SURGERY, THE PATIENT CAME IN REPORTING PAIN AND INSTABILTY DUE TOLIMITED EXTENSION AND FLEXION. THE SURGEON REVISED THE TIBIAL TRAY TO REVISION AND UPSIZED THE LINER (FROM 11 MM TO 12MM) AND IMPLANTED TIBIAL AUGMENTS. THE SURGERY WAS COMPLETED SUCCESFULLY.
Patient problems: Joint Laxity
MAUDE report 23655143 ↗InjuryNov 18, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
THE PATIENT HAD PAIN DUE TO A LOOSE TIBIA AND FEMUR IMPLANTS. AT ABOUT 6 YEARS POST PRIMARY THE SURGEON REVISED THE TIBIA, INSERT, AND FEMUR. THE SURGERY WAS COMPLETED SUCCESSFULLY. LACK OF ADHERENCE TO CEMENT FOUND. NON MEDACTA CEMENT WAS USED DURING PRIMARY.
Patient problems: Aseptic loosening
MAUDE report 23583097 ↗InjuryAug 6, 2025Implant subsidence; Implant subsidence
4 YEARS AFTER THE PRIMARY PROCEDURE, THE PATIENT REPORTED PAIN CAUSED BY TIBIAL COMPONENT SUBSIDENCE. THE EVENT WAS LIKELY RELATED TO POOR BONE QUALITY AND EXCESSIVE BODY WEIGHT. THE SURGEON REVISED ALL COMPONENTS USING REVISION IMPLANTS. THE SURGERY WAS SUCCESSFULLY COMPLETED.
Patient problems: Pain
MAUDE report 22725446 ↗InjuryMay 19, 2025Migration; Migration
AT ABOUT 2 YEARS AND 3 MONTHS AFTER PRIMARY, THE PATIENT CAME IN REPORTING TIBIAL SUBSIDENCE WHICH CAUSED SUBSEQUENT TIBIAL LOOSENING CAUSED BY TRANSPORTING HEAVY LOADS. SURGEON REVISED ALL THE COMPONENTS SUCCESSFULLY.
Patient problems: Implant Pain
MAUDE report 22042951 ↗InjuryApr 30, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AT ABOUT 5 MONTHS FROM THE PRIMARY, THE PATIENT CAME IN REPORTING STIFFNESS. REHABILITATION HAS BEEN PERFORMED IN ORDER TO RESTORE THE JOINT MOBILITY.
Patient problems: Loss of Range of Motion
MAUDE report 21932364 ↗InjuryDec 11, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
THE PATIENT HAD PAIN AND WAS UNABLE TO FULLY EXTEND THEIR KNEE. HE DIDN'T LISTEN/FOLLOW RECOVERY INSTRUCTIONS AFTER SURGERY. AT ABOUT 8 MONTHS POST PRIMARY THE SURGEON DECIDED TO REVISE TIBIAL TRAY, TIBIAL INSERT, AND FEMORAL COMPONENT FROM A GMK SPHERIKA TO A GMK HINGE. IT IS UNKNOWN WHY THE SURGEON DECIDED TO REVISE EVERYTHING. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Loss of Range of Motion
MAUDE report 20900788 ↗InjuryOct 16, 2024Migration; Migration
REVISION SURGERY PERFORMED DUE TO TIBIAL SUBSIDENCE. THE GMK-SPHERE TIBIAL TRAY SIZE 2 IMPLANTED DURING PRIMARY WAS UNDERSIZED. THE SURGEON REVISED THE TIBIAL TRAY AND INSERT AND THE SURGERY WAS COMPLETED SUCCESSFULLY. A GMK REVISION TIBIAL TRAY SIZE 3 WITH EXTENSION STEM WERE SUCCESSFULLY IMPLANTED. PRIMARY SURGERY DATE UNKNOWN.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; No Clinical Signs, Symptoms or Conditions
MAUDE report 20460603 ↗InjurySep 6, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
AT ABOUT 3 YEARS AND 7 MONTHS AFTER PRIMARY, THE PATIENT HAS BEEN REVISED DUE TO FEMORAL AND TIBIAL TRAY MOBILIZATION. ALL IMPLANTS REVISED SUCCESSFULLY TO COMPETITOR'S SYSTEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20154931 ↗InjuryApr 9, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AT 2 YEARS AND 4 MONTHS FROM THE PRIMARY, THE PATIENT CAME IN REPORTING PAIN DUE TO A FRACTURED TIBIA AND THE CAUSE IS UNKNOWN. THE SURGEON CONVERTED THE PATIENT TO REVISION COMPONENTS. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Bone Fracture(s)
MAUDE report 19071420 ↗InjuryApr 9, 2024Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
AT 1 MONTH FROM THE PRIMARY, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE HEAD AND LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Medical device site infection
MAUDE report 19069665 ↗InjuryApr 9, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
AT ABOUT 3 YEARS AND 2 MONTHS FROM THE PRIMARY, THE PATIENT CAME IN REPORTING PAIN DUE TO A LOOSE TIBIAL COMPONENT. THE CAUSE OF THE LOOSE TIBIAL COMPONENT IS UNKNOWN. THE SURGEON REVISED THE TIBIAL TRAY, INSERT, AND FEMUR TO COMPETITOR COMPONENTS. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Implant Pain
MAUDE report 19069663 ↗InjuryFeb 29, 2024Migration or Expulsion of Device; Migration or Expulsion of Device
AT ABOUT 3 YEARS 8 MONTHS AFTER THE PRIMARY, REVISION SURGERY PERFORMED FOR TIBIAL TRAY SUBSIDENCE. A SEMI-CONSTRAINT (GMK REVISION) WAS SUCCESSFULLY IMPLANTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18805802 ↗InjuryFeb 23, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
AT 5 YEARS AND 11 MONTHS FROM THE PRIMARY, THE PATIENT CAME IN REPORTING INSTABILITY DUE TO LAXITY AND TIBIAL COMPONENT LOOSEING. THE CAUSE IS UNKNOWN. THE SURGEON REVISED THE TIBIAL TRAY AND INSERT (FROM 12 TO 14 MM). THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Joint Laxity; Pain
MAUDE report 18765738 ↗InjuryNov 3, 2023Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
ABOUT 2 YEARS AND 4 MONTHS AFTER THE PRIMARY SURGERY, TIBIAL TRAY AND INSERT REVISED DUE TO TIBIAL TRAY LOOSENING.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 18063840 ↗InjurySep 27, 2023Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
AT ABOUT 1 MONTH AFTER THE PRIMARY, THE PATIENT CAME IN REPORTING PAIN AND DISCOMFORT AND THE CAUSE IS UNKNOWN. THE SURGEON REVISED THE FEMUR, INSERT AND TIBIAL TRAY WITH REVISION COMPONENTS. THE SURGERY WAS COMPLETED SUCCESSFULLY. ANY ISSUE WITH THE IMPLANTS DETECTED.
Patient problems: Pain
MAUDE report 17823868 ↗InjurySep 1, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON THE (B)(6) 2022, THE PATIENT UNDERWENT A SURGERY DUE TO FLEXION DEFAULT.
Patient problems: Joint Laxity
MAUDE report 17670075 ↗InjuryJul 28, 2023Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
AT ABOUT 3 YEARS FROM THE PRIMARY, REVISION SURGERY DUE TO CEMENTED TIBIAL PLATE LOOSENING. THE SURGEON REVISED THE TIBIAL TRAY AND THE INSERT (FROM 12 TO 13 MM). TIBIAL COMPONENTS REVISED SUCCESSFULLY.
Patient problems: Joint Laxity
MAUDE report 17412888 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K090988. One submission often covers many models, so check the product and lot details in each record.
- Z-0266-2018 · Oct 24, 2017 · TerminatedThe GMK knee system: designed for cemented use in total knee arthroplasty. Sphere and Primary Inserts with part number ranges: 02.12.0110FL - 02.12.0620FL; 02.12.0110FR - 02.12.0
Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page