GMK Primary Total Knee System 02.07.1205RRecalls & reported problems
Medacta International SA · Fixed tibial tray cemented right, size 5 · Class 2 · Implant
All models of GMK Primary Total Knee System (99)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This is a fixed tibial tray used in Medacta's GMK primary total knee replacement. The 61 reports mostly describe knees that became loose or wobbly after surgery and needed repeat operations, along with infections, pain, and joint problems; the injuries list shows 60 injuries and no deaths. Report numbers stayed flat, with 6 in the last 12 months and 5 in the prior12 months. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 61
- Deaths reported
- 0
- Injuries reported
- 60
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 61 adverse event reports that list this exact device (DI 07630030819933), received between Aug 30, 2018 and May 20, 2026 — classified as 60 injuries, 1 malfunction.
The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (18 reports).
1 other recall cover devices cleared under the same FDA submission (K090988); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Loosening of Implant Not Related to Bone-Ingrowth18
- Loose or Intermittent Connection15
- Appropriate Device Problem Term/Code Not Available12
- Device Appears to Trigger Rejection10
- Adverse Event Without Identified Device or Use Problem6
- Defective Device1
- Implant subsidence1
- Inadequacy of Device Shape and/or Size1
Patient problems
- Pain13
- Medical device site infection11
- Unspecified Infection10
- Joint Disorder7
- Failure of Implant5
- Insufficient Information3
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
- Implant Pain2
Recent adverse event reports
The 25 most recent of 61 reports, in the reporter's own words as published by the FDA.
InjuryMay 20, 2026Device Appears to Trigger Rejection
REVISION SURGERY DUE TO INFECTION AFTER ABOUT 1 YEAR AND 2 MONTHS FROM PRIMARY. THE SURGEON REMOVED ALL GMK-SPHERIKA IMPLANTS AND IMPLANTED ANTIBIOTIC SPACERS. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Medical device site infection
MAUDE report 25230406 ↗InjuryMay 15, 2026Adverse Event Without Identified Device or Use Problem
THE PATIENT CAME IN DUE TO SIGNS OF INFECTION AND THE PATHOGEN IS UNKNOWN. AT ABOUT 1 YEAR AND 1 MONTH POST PRIMARY THE SURGEON REVISED ALL COMPONENTS (TIBIAL TRAY, LINER, AND FEMORAL COMPONENT) AND IMPLANTED ANTIBIOTIC SPACERS. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Medical device site infection
MAUDE report 25185358 ↗InjuryApr 23, 2026Device Appears to Trigger Rejection
REVISION SURGERY DUE TO INFECTION AFTER ABOUT 7 MONTHS FROM PRIMARY SURGERYAND THE PATHOGEN IS UNKNOWN. THE SURGEON REMOVED THE FEMORAL COMPONENT, INSERT, AND TIBIAL TRAY, THEN IMPLANTED ANTIBIOTIC SPACERS. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Medical device site infection
MAUDE report 24964782 ↗InjuryMar 3, 2026Adverse Event Without Identified Device or Use Problem
AT ABOUT 3 YEARS 7 MONTHS AFTER THE PRIMARY, THE PATIENT CAME IN PRESENTING TIGHTNESS AND THE CAUSE IS UNKNOWN. THE SURGEON REVISED FEMORAL COMPONENT, TIBIAL TRAY AND INSERT (SIZE 5 TO SIZE 4). THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Loss of Range of Motion
MAUDE report 24494907 ↗InjuryFeb 24, 2026Loosening of Implant Not Related to Bone-Ingrowth; Implant subsidence
REVISION SURGERY WAS PERFORMED 5 YEARS AND 7 MONTHS AFTER THE PRIMARY SURGERY DUE TO MEDIAL TIBIAL SUBSIDENCE AND FEMORAL LOOSENING. BOTH COMPONENTS WERE LOOSE AT THE CEMENT MANTLE. ADDITIONALLY, THE INSERT SHOWED SIGNS OF SCORING AND PITTING.
Patient problems: Aseptic loosening
MAUDE report 24429836 ↗InjuryDec 2, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AT 1 MONTH FROM THE PRIMARY, THE PATIENT CAME IN DUE TO AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON REMOVED ALL PRIMARY IMPLANTS AND IMPLANTED AN ANTIBIOTIC SPACER TO TREAT THE INFECTION. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Medical device site infection
MAUDE report 23690770 ↗InjuryOct 30, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT CAME IN REPORTING INSTABILITY FOR TWO TIMES AND THE CAUSE IS UNKNOWN. THE FIRST EPISODE OCCURRED ON (B)(6) 2024, AT ABOUT 2 YEARS AFTER THE PRIMARY SURGERY WITH MEDACTA PRODUCTS. THE SURGEON REVISED ONLY THE LINER (FROM 14 MM TO 17 MM). THE SECOND EPISODE OCCURRED ON (B)(6) 2025 THE SURGEON REVISED THE INSERT, TRAY, AND FEMUR FROM GMK-REVISION TO GMK-HINGE. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Implant Pain; Joint Laxity
MAUDE report 23423786 ↗InjuryApr 28, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
THE PATIENT HAD A PRIMARY KNEE SURGERY ON (B)(6) 2023. ON (B)(6) 2023, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE POLY AND THE SURGERY WAS COMPLETED SUCCESSFULLY. ON (B)(6) 2025, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED ALL COMPONENTS. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Medical device site infection
MAUDE report 21917680 ↗InjuryMar 26, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
AT ABOUT 4 YEARS 9 MONTHS FROM PRIMARY, THE PATIENT CAME IN DUE TO SIGNS OF INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON REVISED ALL MEDACTA HARDWARE AND REIMPLANTED PERMANENT PRODUCT. THE SURGERY WAS COMPLETED SUCCESSFUL.
Patient problems: Unspecified Infection
MAUDE report 21695441 ↗InjuryFeb 25, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
THE PATIENT HAD A PRIMARY KNEE SURGERY ON (B)(6) 2023. ON (B)(6) 2023, THE PATIENT HAD AN INFECTION AND THE PATHOGEN IS UNKNOWN (B)(4). THE SURGEON PERFORMED A WASHOUT AND REVISED THE INSERT. THE SURGERY WAS COMPLETED SUCCESSFULLY. ON (B)(6) 2025, THE PATIENT HAD AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON REVISED ALL MEDACTA HARDWARE AND REIMPLANTED PERMANENT PRODUCT. THE SURGERY WAS COMPLETED SUCCESSFULLY.)
Patient problems: Medical device site infection
MAUDE report 21458161 ↗InjuryNov 14, 2024Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
REVISION SURGERY AT ABOUT 1 YEAR AND 4 MONTHS POST PRIMARY DUE TO INFECTION, THE PATHOGEN IS UNKNOWN. THE SURGEON REVISED ALL MEDACTA HARDWARE AND REIMPLANTED PERMANENT PRODUCT. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Medical device site infection
MAUDE report 20684039 ↗InjurySep 17, 2024Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
REVISION DUE TO SIGNS OF INFECTION AND THE PATHOGEN IS UNKNOWN. AT ABOUT 4 MONTHS FROM THE PRIMARY SURGERY THE SURGEON REMOVED ALL MEDACTA HARDWARE AND IMPLANTED AN ANTIBIOTIC SPACER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Medical device site infection
MAUDE report 20237262 ↗InjuryJul 1, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
AT ABOUT 6 YEARS 2 MONTHS AFTER THE PRIMARY, THE PATIENT CAME IN REPORTING PAIN AND THE CAUSE IS UNKNOWN. THE SURGEON REVISED ALL COMPONENTS EXCEPT THE PATELLA TO REVISION COMPONENTS. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Pain
MAUDE report 19642078 ↗InjuryApr 25, 2024Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
AT ABOUT 2 MONTHS AFTER THE PRIMARY, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON REMOVED ALL COMPONENTS AND IMPLANTED AN ANTIBIOTIC SPACER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Medical device site infection
MAUDE report 19178698 ↗InjuryMar 26, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
THE PATIENT HAD A PRIMARY KNEE SURGERY ON (B)(6) 2020. ON (B)(6) 2023, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE LINER AND PATELLA. THE SURGERY WAS COMPLETED SUCCESSFULLY. PRESENTLY, ON (B)(6) 2024, THE PATIENT CAME IN REPORTING PAIN DUE TO ASEPTIC LOOSENING AND PAIN AND THE CAUSE IS UNKNOWN. THE SURGEON REVISED ALL COMPONENTS TO HINGE COMPONENTS AND THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 18977237 ↗InjuryFeb 5, 2024Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
AT ABOUT 2 YEARS AND 6 MONTHS FROM THE PRIMARY, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT, REMOVED ALL IMPLANTS, AND IMPLANTED AN ANTIBIOTIC SPACER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Unspecified Infection
MAUDE report 18639657 ↗InjuryDec 5, 2023Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
REVISION SURGERY DUE TO LOOSENING OF CEMENTED TIBIAL AND FEMORAL COMPONENTS, AT ABOUT 1 YEAR 4 MONTHS FROM THE PRIMARY.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 18266627 ↗InjuryNov 2, 2023Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
REVISION SURGERY DUE TO SECONDARY PATELLA RESURFACING AND INLAY SWAP (AS PER STANDARD PROCEDURE), AT ABOUT 2 YEARS 3 MONTHS FROM THE PRIMARY. DURING INTRAOPERATIVE CHECK OF PROSTHESIS FIXATION, AFTER LITTLE HAMMERING, FEMORAL COMPONENT CEMENTED AND TIBIAL TRAY FIXED CEMENTED WERE EASY TO TAKE OUT. NO CEMENT ON BACKSURFACE. THE SURGEON REVISED ALL COMPONENTS TO COMPETITOR'S SYSTEM AND THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Pain
MAUDE report 18054070 ↗InjuryOct 27, 2023Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
AT ABOUT 11 YEARS 6 MONTHS AFTER THE PRIMARY, REVISION SURGERY DUE TO A LOOSE CEMENTED TIBIAL TRAY FIXED. THE SURGEON OBSERVED THAT THE COBALT CEMENT SEEMED TO BE THE LOOSENING FACTOR. THE SURGEON REVISED ALL COMPONENTS TO REVISION COMPONENTS AND THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18020319 ↗InjuryAug 4, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. ABOUT 1 YEAR AND 11 MONTHS AFTER THE PRIMARY SURGERY, THE SURGEON PERFORMED A WASHOUT, REMOVED ALL IMPLANTS, AND IMPLANTED AN ANTIBIOTIC SPACER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Medical device site infection
MAUDE report 17457135 ↗InjuryJul 25, 2023Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
ON (B)(6) 2022 THE PATIENT HAD A PRIMARY KNEE SURGERY. ON (B)(6) 2022, THE PATIENT REPORTED ANTERIOR KNEE PAIN. THE SURGEON RESURFACED THE NATURAL PATIENT'S PATELLA. PRESENTLY ON (B)(6) 2023, REVISION SURGERY DUE TO CEMENTED TIBIAL TRAY OVERHANGING. PAIN DESCRIBED BY THE PATIENT. NO INFECTION, NO LOOSENING OF THE IMPLANTS. THE SURGEON REVISED SUCCESSFULLY THE TIBIAL COMPONENTS.
Patient problems: Pain
MAUDE report 17388872 ↗InjuryJul 5, 2023Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
AT ABOUT 7 MONTHS AFTER THE PRIMARY, THE PATIENT CAME IN REPORTING INSTABILITY DUE TO BEING IN EXTREME VALGUS AND THE CAUSE IS UNKNOWN. THE SURGEON REVISED ALL COMPONENTS AND ADDED AUGMENTS. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Joint Laxity
MAUDE report 17254289 ↗InjuryMay 8, 2023Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
AT ABOUT 2 YEARS AND 8 MONTHS AFTER THE PRIMARY SURGERY, THE PATIENT CAME IN REPORTING PAIN DUE TO ASEPTIC LOOSENING. THE SURGEON REVISED THE CEMENTED TIBIAL TRAY AND LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Pain
MAUDE report 16884698 ↗InjuryApr 18, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
AT ABOUT 1 MONTH AFTER THE PRIMARY SURGERY, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT, REMOVED ALL COMPONENTS AND IMPLANTED AN ANTIBIOTIC SPACER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Medical device site infection
MAUDE report 16761476 ↗InjuryMar 30, 2023Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
AT ABOUT 7 MONTHS AFTER THE PRIMARY SURGERY, THE PATIENT CAME IN REPORTING PAIN DUE TO A LOOSE TIBIAL TRAY AND THE CAUSE IS UNKNOWN. THE SURGEON REVISED SUCCESSFULLY ALL COMPONENTS TO GMK-REVISION.
Patient problems: Implant Pain
MAUDE report 16644025 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K090988. One submission often covers many models, so check the product and lot details in each record.
- Z-0266-2018 · Oct 24, 2017 · TerminatedThe GMK knee system: designed for cemented use in total knee arthroplasty. Sphere and Primary Inserts with part number ranges: 02.12.0110FL - 02.12.0620FL; 02.12.0110FR - 02.12.0
Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.