GMK Primary Total Knee System 02.07.2002RRecalls & reported problems
Medacta International SA · Femur STD cemented right, size 2 · Class 2 · Implant
All models of GMK Primary Total Knee System (99)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07630030820182), received between Aug 27, 2020 and Nov 22, 2021 — classified as 2 injuries.
The most frequently reported specific device problem is “Loose or Intermittent Connection” (1 report).
1 other recall cover devices cleared under the same FDA submission (K090988); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Appropriate Device Problem Term/Code Not Available2
- Loose or Intermittent Connection1
Patient problems
- Joint Disorder1
- Unspecified Infection1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryNov 22, 2021Appropriate Device Problem Term/Code Not Available
AT 9 YEARS AND 9 MONTHS AFTER THE PRIMARY SURGERY, THE PATIENT CAME IN REPORTING INSTABILITY AND THE CAUSE OF THE INSTABILITY IS UNKNOWN. THE SURGEON REVISED ALL IMPLANTS AND THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Unspecified Infection
MAUDE report 12851834 ↗InjuryAug 27, 2020Loose or Intermittent Connection; Appropriate Device Problem Term/Code Not Available
THE PATIENT CAME IN AFTER 9 YEARS AND 2 MONTHS FORM THE PRIMARY SURGERY FOR A POST-OP APPOINTMENT. X-RAYS INDICATED A LOOSE FEMORAL COMPONENT AND TIBIAL TRAY. THE SURGEON REVISED THE FEMORAL COMPONENT, TIBIAL TRAY, AND INSERT. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Joint Disorder
MAUDE report 10458656 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K090988. One submission often covers many models, so check the product and lot details in each record.
- Z-0266-2018 · Oct 24, 2017 · TerminatedThe GMK knee system: designed for cemented use in total knee arthroplasty. Sphere and Primary Inserts with part number ranges: 02.12.0110FL - 02.12.0620FL; 02.12.0110FR - 02.12.0
Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page