Skip to content

GMK Hinge Total Knee System 02.09.0114HRecalls & reported problems

Medacta International SA · Fixed tibial insert # 1/14mm · Class 2 · Implant

All models of GMK Hinge Total Knee System (55)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07630030824999), received between Aug 28, 2024 and Jun 26, 2026 — classified as 2 injuries.

The most frequently reported specific device problem is “Device Appears to Trigger Rejection” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K130299); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Appears to Trigger Rejection1
  • Device Dislodged or Dislocated1
  • Fracture1

Patient problems

  • Joint Laxity1
  • Medical device site infection1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 26, 2026Device Appears to Trigger Rejection

    AT ABOUT 3 MONTHS FROM THE PRIMARY, THE PATIENT CAME IN DUE TO SIGNS OF INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE 14MM GMK-HINGE SIZE 1 INSERT TO A SAME SIZE INSERT, AND REVISED THE 14MM TINBN COATED POST EXTENSION TO A SAME SIZE EXTENSION. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Medical device site infection

    MAUDE report 25632984 ↗
  • InjuryAug 28, 2024Fracture; Device Dislodged or Dislocated; Fracture; Device Dislodged or Dislocated

    THE PATIENT CAME IN FOR A POST-OP APPOINTMENT AND A POST-OP X-RAY SHOWED A BROKEN POLY SCREW AND THE CAUSE IS UNKNOWN. ALSO, THE FEMUR WAS FOUND DISSOCIATED FROM THE BASEPLATE (POST WAS OUT OF ITS SEAT). ABOUT 1 YEAR AND 2 MONTHS AFTER THE PRIMARY SURGERY, THE SURGEON DECIDED TO REVISE THE PATIENT AND REVISED THE INSERT (SAME REF, DIFFERENT LOT). THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Joint Laxity

    MAUDE report 20083675 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K130299. One submission often covers many models, so check the product and lot details in each record.

  • Z-2661-2017 · May 19, 2017 · Terminated
    Medacta GMK HINGE, Tibial Insert Fixed, SIZE 2, HEIGHT 23 mm, REF 02.09.0223H, STERILE EO
  • Z-0275-2016 · Apr 17, 2015 · Terminated
    GMK Hinge knee prosthesis; Model Number: 02.09.TA410; Orthopedics and Orthopedic Surgery

Other Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer devices

All Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer devices →

Sources and updates