Skip to content

GMK Hinge Total Knee System 02.09.0314HRecalls & reported problems

Medacta International SA · Fixed tibial insert # 3/14mm · Class 2 · Implant

All models of GMK Hinge Total Knee System (55)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
11
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 07630030825132), received between Sep 7, 2017 and Feb 6, 2026 — classified as 11 injuries.

The most frequently reported specific device problem is “Migration” (2 reports).

2 other recalls cover devices cleared under the same FDA submission (K130299); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem4
  • Migration2
  • Appropriate Device Problem Term/Code Not Available1
  • Break1
  • Device Appears to Trigger Rejection1
  • Loose or Intermittent Connection1
  • Material Rupture1
  • Microbial Contamination of Device1

Patient problems

  • Unspecified Infection4
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
  • Damage to Ligament(s)1
  • Hyperextension1
  • Implant Pain1
  • Medical device site infection1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 6, 2026Migration

    THE PATIENT HAD PRIMARY LEFT KNEE SURGERY ON (B)(6) 2025. ON (B)(6) 2025, THE PATIENT HAD PAIN DUE TO AN MCL INJURY. AT ABOUT 1 YEAR AFTER THE PRIMARY, THE SURGEON CONVERTED THE PATIENT FROM A SPHERE KNEE TO A HINGE KNEE BY REVISING THE TIBIAL TRAY, INSERT AND FEMORAL COMPONENT. THE SURGERY WAS COMPLETED SUCCESSFULLY. THE HINGE YOKE SCREW WAS INSERTED USING A MANUAL MEDACTA SCREWDRIVER AND FORCED INTO PLACE WITH A MALLET INSTEAD OF THE PRESCRIBED 6 NM TORQUE LIMITER. ON (B)(6) 2026, THE PATIENT PRESENTED DUE TO A GMK HINGE YOKE SCREW THAT HAD UNSCREWED, CAUSING KNEE INSTABILITY. THE SURGEON REVISED THE LINER, IMPLANTING A LINER OF THE SAME SIZE AND SCREWED THE YOKE SCREW WITH THE CORRECT TORQUE DRIVER. NO PROBLEMS WERE DETECTED ON THE OTHER IMPLANTS DURING THE REVISION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24276573 ↗
  • InjuryMar 11, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AT ABOUT 6 MONTHS FROM PRIMARY, THE PATIENT CAME IN REPORTING PAIN AND THE CAUSE IS UNKNOWN. THE SURGEON REVISED THE FEMORAL COMPONENT AND INSERT. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Implant Pain

    MAUDE report 21574226 ↗
  • InjuryMar 5, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ABOUT 8 MONTHS AFTER THE PRIMARY SURGERY (PRIMARY GMK-HINGE), DURING A FOLLOW-UP VISIT, X-RAYS WERE TAKEN AND A HYPEREXTENSION OF ABOUT 12 DEGREES WAS OBSERVED. PATIENT IS UNCOMFORTABLE AND UNABLE TO WALK PROPERLY. HE WILL NEED A REVISION SURGERY. ADDITIONAL INFO: MYKNEE CASE CODE: (B)(4). MYKNEE BLOCK WAS UTILIZED FOR THE TIBIAL RESECTION BUT NOT FOR THE FEMUR.

    Patient problems: Hyperextension

    MAUDE report 21525700 ↗
  • InjuryJan 29, 2025Migration; Migration

    REVISION SURGERY DUE TO GMK HINGE POST SCREW UNSCREWING AND SUBSEQUENT MIGRATION AT 2 MONTHS AFTER PRIMARY. IT HAS BEEN OBSERVED THAT THE THREAD OF THE SCREW WAS DAMAGED. TORQUE LIMITER WAS USED. THE SURGEON REVISED THE INSERT SUCCESSFULLY.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 21264315 ↗
  • InjuryJan 25, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AT ABOUT 9 MONTHS AFTER THE PRIMARY, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Medical device site infection

    MAUDE report 18578852 ↗
  • InjuryApr 14, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    THE PATIENT CAME IN DUE TO SIGNS OF INFECTION AND THE PATHOGEN IS UNKNOWN. AT 13 DAYS FROM THE PRIMARY SURGERY THE SURGEON PERFORMED A WASHOUT AND POLY SWAP AND THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Unspecified Infection

    MAUDE report 16741306 ↗
  • InjuryAug 2, 2022Break; Loose or Intermittent Connection

    LUXATION OF THE HINGE INLAY, WHEN THE PATIENT STOOD UP FROM A CHAIR. DURING THE REVISION, THE SCREW OF THE INLAY WAS BROKEN. THE SCREW OF THE HINGE MECHANISM WAS CORRECTLY FIXATED. REVISION SURGERY PERFORMED 3 MONTHS AFTER THE PRIMARY SURGERY; ONLY INSERT REVISED WITH A HIGHER ONE (FROM 14 MM TO 20 MM). A BIG LAXITY WAS OBSERVED DURING REVISION.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 15147140 ↗
  • InjuryJul 18, 2022Material Rupture

    THE PATIENT CAME IN REPORTING PAIN DUE TO A RUPTURED EXTENSOR MECHANISM. THE SURGEON REPAIRED THE EXTENSOR MECHANISM AND SWAPPED THE POLY. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Damage to Ligament(s)

    MAUDE report 15040192 ↗
  • InjuryJul 30, 2019Appropriate Device Problem Term/Code Not Available

    THE PATIENT CAME IN, ONE MONTH AFTER PRIMARY SURGERY, DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND POLY-SWAP AND THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Unspecified Infection

    MAUDE report 8841138 ↗
  • InjuryApr 19, 2018Microbial Contamination of Device

    THE PATIENT CAME IN DUE TO SIGNS OF INFECTION 1 MONTH AND TWO WEEKS AFTER PRIMARY. THE PATHOGEN IS UNAVAILABLE. THE SURGEON WASHED OUT THE KNEE AND SWAPPED THE LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Unspecified Infection

    MAUDE report 7442494 ↗
  • InjurySep 7, 2017Adverse Event Without Identified Device or Use Problem

    THE PATIENT CAME IN DUE TO SIGNS OF INFECTION. THE PATHOGEN IS UNKNOWN. THE SURGEON REVISED THE INSERT. THE SURGERY WAS COMPLETED SUCCESSFULLY. X-RAYS AND EXPLANTS ARE NOT AVAILABLE.

    Patient problems: Unspecified Infection

    MAUDE report 6847606 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K130299. One submission often covers many models, so check the product and lot details in each record.

  • Z-2661-2017 · May 19, 2017 · Terminated
    Medacta GMK HINGE, Tibial Insert Fixed, SIZE 2, HEIGHT 23 mm, REF 02.09.0223H, STERILE EO
  • Z-0275-2016 · Apr 17, 2015 · Terminated
    GMK Hinge knee prosthesis; Model Number: 02.09.TA410; Orthopedics and Orthopedic Surgery

Other Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer devices

All Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page