GMK Hinge Total Knee System 02.09.0417HRecalls & reported problems
Medacta International SA · Fixed tibial insert # 4/17mm · Class 2 · Implant
All models of GMK Hinge Total Knee System (55)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 9
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 07630030825217), received between Oct 12, 2017 and Jun 9, 2026 — classified as 9 injuries.
The most frequently reported specific device problem is “Device Appears to Trigger Rejection” (4 reports).
2 other recalls cover devices cleared under the same FDA submission (K130299); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Appears to Trigger Rejection4
- Adverse Event Without Identified Device or Use Problem1
- Break1
- Material Rupture1
- Microbial Contamination of Device1
- Migration1
Patient problems
- Medical device site infection3
- Unspecified Infection2
- Damage to Ligament(s)1
- Insufficient Information1
- Joint Disorder1
- Joint Laxity1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
InjuryJun 9, 2026Device Appears to Trigger Rejection
REVISION SURGERY DUE TO INFECTION AFTER ABOUT 1 MONTH FROM FIRST REVISION SURGERY. THE PATIENT HAD PRIMARY SURGERY USING COMPETITOR IMPLANTS AND WAS REVISED TO MEDACTA IMPLANTS ON (B)(6) 2026. THE SURGEON PERFORMED A WASHOUT AND REVISED THE GMK-HINGE SIZE 4 - 17MM INSERT TO A GMK-HINGE SIZE 3 - 12MM AT HIS DISCRETION. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Medical device site infection
MAUDE report 25442527 ↗InjuryMay 12, 2026Device Appears to Trigger Rejection
AT ABOUT 1 YEAR AND 8 MONTHS FROM THE PRIMARY, THEPATIENT CAME IN DUE TO SIGNS OF INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT, I&D, AND REVISED THE SIZE4 - 17MM INSERT TO A SIZE4 - 20MM INSERT AT HIS DISCRETION. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Medical device site infection
MAUDE report 25148955 ↗InjuryAug 29, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
AT ABOUT 3 YEARS AND 4 MONTHS THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT, REMOVED ALL IMPLANTS, AND IMPLANTED AN ANTIBIOTIC SPACER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Medical device site infection
MAUDE report 22923098 ↗InjuryMar 3, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
AT ABOUT 2 MONTHS FROM PRIMARY THE SURGEON REVISED THE PATIENT KNEE FOR INFECTION. THE SURGEON SWAPPED THE LINER SUCCESSFULLY.
Patient problems: Unspecified Infection
MAUDE report 16478723 ↗InjuryMar 1, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AT ABOUT 4 YEARS 10 MONTHS AFTER THE PRIMARY, THE PATIENT CAME IN REPORTING PAIN DUE TO STIFFNESS. THE SURGEON RESURFACED THE PATIENT'S NATURAL PATELLA AND DOWNSIZED THE INSERT (17 TO 14 MM). THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Joint Laxity
MAUDE report 16459845 ↗InjuryDec 21, 2022Migration; Migration
REVISION SURGERY AT ABOUT 4 MONTHS FOR YOKE SCREW (POST SCREW) LOOSENING. A NEW PIVOT WITH PIVOT SCREW WAS IMPLANTED SUCCESSFULLY USING THE TORQUE WRENCH.
Patient problems: Insufficient Information
MAUDE report 16022747 ↗InjuryJul 23, 2021Material Rupture
THE PATIENT HAD A PRIMARY AND WAS IMPLANTED WITH COMPETITOR IMPLANTS. ON (B)(6) 2019, THE PATIENT CAME IN AND HAD THE COMPETITOR COMPONENTS REVISED TO MEDACTA FOR REASONS UNKNOWN. THE SURGERY WAS COMPLETED SUCCESSFULLY. ON (B)(6) 2021, 1 YEAR AND 11 MONTHS AFTER HINGE IMPLANTATION THE PATIENT CAME IN REPORTING PAIN DUE TO A RUPTURED QUAD TENDON. THE CAUSE OF THE RUPTURED QUAD TENDON IS UNKNOWN. THE SURGEON PERFORMED A QUAD TENDON REPAIR AND REVISED THE POLY. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Damage to Ligament(s)
MAUDE report 12217045 ↗InjuryAug 25, 2020Microbial Contamination of Device
THE PATIENT CAME IN 2 MONTHS AFTER THE PRIMARY SURGERY DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Unspecified Infection
MAUDE report 10447978 ↗InjuryOct 12, 2017Break; Break
REVISION DUE TO TIBIAL COMPONENT (ASEPTIC) LOOSENING AND BREAKAGE OF THE INSERT SCREW. THE SURGEON REPLACED THE TIBIAL COMPONENT WITH INLAY AND STEM AND THE 2 WEDGES. THE SURGERY WAS COMPLETED SUCCESSFULLY. THE SURGEON THINKS THAT THE SCREW BREAKAGE IS A CONSEQUENCE OF A PATIENT FALL.
Patient problems: Joint Disorder
MAUDE report 6944514 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K130299. One submission often covers many models, so check the product and lot details in each record.
- Z-2661-2017 · May 19, 2017 · TerminatedMedacta GMK HINGE, Tibial Insert Fixed, SIZE 2, HEIGHT 23 mm, REF 02.09.0223H, STERILE EO
- Z-0275-2016 · Apr 17, 2015 · TerminatedGMK Hinge knee prosthesis; Model Number: 02.09.TA410; Orthopedics and Orthopedic Surgery
Other Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer devices
- ELEOS™ 25001208EOnkos Surgical, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER3 deaths193 reports
- ELEOS™ THSMWRS01MOnkos Surgical, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER190 reports
- "UNITED" USTAR II Hip System 1115-1413United Orthopedic Corp. · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer3 recalls3 reports
- "UNITED" USTAR II Hip System 1115-1213United Orthopedic Corp. · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer3 recalls0 reports
- "UNITED" USTAR II Hip System 1115-1417United Orthopedic Corp. · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer3 recalls0 reports
- "UNITED" USTAR II Hip System 1115-1217United Orthopedic Corp. · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer3 recalls0 reports
- Endo-Model Modular Knee Prosthesis System 15-3816/12WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer2 recalls6 reports
- Endo-Model Modular Knee Prosthesis System 15-3816/11WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer2 recalls6 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page