GMK Hinge Total Knee System 02.09.0420HRecalls & reported problems

Medacta International SA · Fixed tibial insert # 4/20mm · Class 2 · Implant

All models of GMK Hinge Total Knee System (55)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07630030825224), received between Sep 19, 2024 and Aug 12, 2026 — classified as 3 injuries.

The most frequently reported specific device problem is “Device Appears to Trigger Rejection” (2 reports).

2 other recalls cover devices cleared under the same FDA submission (K130299); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Appears to Trigger Rejection2
  • Migration1

Patient problems

  • Medical device site infection2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 12, 2026Device Appears to Trigger Rejection

    Revision surgery due to infection after about 7 months from first revision surgery. The surgeon performed a washout and revised the 20mm - Size 4 GMK-Hinge insert to a same size insert. The patient ha primary surgery with competitor implants.

    Patient problems: Medical device site infection

    MAUDE report 50222759 ↗
  • InjuryNov 21, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    ON (B)(6) 2025 THE SURGEONPERFORMED A KNEE REVISION (NON-MEDACTA) FOR INSTABILITY AND IMPLANTED A HINGE PROSTHESIS. ON (B)(6) 2025 THE PATIENT PRESENTED WITH AN INFECTIONAND THE SURGEONPERFORMED LAVAGE AND POLYETHYLENE EXCHANGE ONLY. THE PATIENT IS CURRENTLY RECEIVING ANTIBIOTIC THERAPY WITH TAZOCYCLINE AND DAPTOMYCIN.

    Patient problems: Medical device site infection

    MAUDE report 23615335 ↗
  • InjurySep 19, 2024Migration; Migration

    DURING A CHECK-UP, THE INSERT SCREW WAS FOUND UNSCREWED AND MIGRATED. PRIMARY SURGERY PERFORMED 10 YEARS AGO. THE SURGEON IS CONSIDERING REMOVING AND REPLACING THE SCREW.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 20254284 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K130299. One submission often covers many models, so check the product and lot details in each record.

  • Z-2661-2017 · May 19, 2017 · Terminated
    Medacta GMK HINGE, Tibial Insert Fixed, SIZE 2, HEIGHT 23 mm, REF 02.09.0223H, STERILE EO
  • Z-0275-2016 · Apr 17, 2015 · Terminated
    GMK Hinge knee prosthesis; Model Number: 02.09.TA410; Orthopedics and Orthopedic Surgery