Skip to content

GMK Hinge Total Knee System 02.09.2603LRecalls & reported problems

Medacta International SA · Femoral Component Left, size 3 · Class 2 · Implant

All models of GMK Hinge Total Knee System (55)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07630030825439), received between Mar 9, 2018 and Mar 2, 2026 — classified as 5 injuries.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K130299); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available2
  • Detachment of Device or Device Component1
  • Loose or Intermittent Connection1
  • Loosening of Implant Not Related to Bone-Ingrowth1

Patient problems

  • Bone Fracture(s)2
  • Implant Pain1
  • Insufficient Information1
  • Unspecified Infection1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 2, 2026Loosening of Implant Not Related to Bone-Ingrowth

    AT ABOUT 7 MONTHS FROM THE PRIMARY, THE PATIENT CAME REPORTING KNEE PAIN AND INSTABILITY. A BONE SCAN REVEALED THAT, IN THE GMK HINGE SYSTEM, BOTH THE TIBIAL AND FEMORAL COMPONENTS WERE LOOSE, WHILE THE STEMS APPEARED WELL FIXED. THE SURGEON DISASSOCIATED THE STEMS FROM THE FEMORAL AND TIBIAL COMPONENTS IN SITU AND REVISED BOTH THE FEMUR AND TIBIA. TO RECOUPLE THE NEWLY REVISED FEMUR, THE SURGEON ADDED AN OFFSET TO THE STEM LEFT IN THE BODY. THE TIBIA WAS RECOUPLED TO THE ORIGINAL STEM WITHOUT THE USE OF AN OFFSET. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THE SURGEON DID NOT FOLLOW THE CORRECT PROCEDURE FOR THE GMK HINGE SYSTEM. THE SURGEON WAS REMINDED OF THE PROPER INDICATIONS FOR USE AND THE CORRECT SURGICAL TECHNIQUE FOR THE GMK HINGE SYSTEM AND PROVIDED THEM WITH THE IFU AND SURGICAL TECHNIQUE DOCUMENTATION.

    Patient problems: Implant Pain

    MAUDE report 24481879 ↗
  • InjuryJun 26, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IN THE SAME DAY AFTER THE KNEE SURGERY PERFORMED ON THE (B)(6) 2024, A POST-OP PERIPROSTHETIC BONE FRACTURE WAS OBSERVED. A FOLLOW UP VISIT WAS DONE ON THE (B)(6) 2024. ACCORDING TO THE REVISION SURGEON, THE CAUSES OF THE POSTOPERATIVE FRACTURE WERE THE PATIENT'S OSTEOPOROSIS AND FRAGILITY OF THE ANTERIOR FEMORAL CORTEX, EXACERBATED BY A FALSE ROUTE DURING REAMING. THE IMPLANTS WERE NOT INVOLVED AT THIS STAGE. REVISION OF FEMORAL COMPONENT AND EXTENSION STEM PERFORMED SUCCESSFULLY.

    Patient problems: Bone Fracture(s)

    MAUDE report 19614851 ↗
  • InjuryDec 20, 2021Detachment of Device or Device Component

    ABOUT 2 MONTHS AFTER THE PREVIOUS REVISION SURGERY(PREVIOUSLY ALL IMPLANTS WERE REVISED DUE TO THE TIBIA TENDON HAD COME OFF), THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT, REMOVED ALL COMPONENTS AND IMPLANTED AN ANTIBIOTIC SPACER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Unspecified Infection

    MAUDE report 13042794 ↗
  • InjurySep 23, 2021Loose or Intermittent Connection

    THE PATIENT CAME IN REPORTING PAIN DUE TO A LOOSE FEMUR AND THE CAUSE OF THE LOOSE FEMUR IS UNKNOWN. 2 YEARS AND 10 MONTHS AFTER THE PRIMARY SURGERY, THE SURGEON REVISED THE FEMORAL COMPONENT AND POLY. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Insufficient Information

    MAUDE report 12517862 ↗
  • InjuryMar 9, 2018Appropriate Device Problem Term/Code Not Available

    THE PATIENT CAME IN COMPLAINING OF PAIN. THE PATIENT FRACTURED HER FEMUR. THE CAUSE OF THE FRACTURE IS UNKNOWN. TRAUMA WAS NOT REPORTED. THE SURGEON REVISED THE IMPLANTS. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Bone Fracture(s)

    MAUDE report 7327060 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K130299. One submission often covers many models, so check the product and lot details in each record.

  • Z-2661-2017 · May 19, 2017 · Terminated
    Medacta GMK HINGE, Tibial Insert Fixed, SIZE 2, HEIGHT 23 mm, REF 02.09.0223H, STERILE EO
  • Z-0275-2016 · Apr 17, 2015 · Terminated
    GMK Hinge knee prosthesis; Model Number: 02.09.TA410; Orthopedics and Orthopedic Surgery

Other Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer devices

All Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page