GMK Sphere Total Knee System 02.12.0005LRecalls & reported problems
Medacta International SA · Femur Sphere cemented Left, size 5 · Class 2 · Implant
All models of GMK Sphere Total Knee System (209)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 17
- Deaths reported
- 0
- Injuries reported
- 17
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 07630030825859), received between Sep 6, 2017 and Mar 10, 2026 — classified as 17 injuries.
The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Loosening of Implant Not Related to Bone-Ingrowth5
- Appropriate Device Problem Term/Code Not Available4
- Adverse Event Without Identified Device or Use Problem3
- Device Appears to Trigger Rejection2
- Inadequacy of Device Shape and/or Size1
- Insufficient Device Problem Information1
- Loose or Intermittent Connection1
- Malposition of Device1
Patient problems
- Joint Laxity4
- Unspecified Infection4
- Pain3
- Joint Disorder2
- Loss of Range of Motion2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Implant Pain1
- Inadequate Osseointegration1
Recent adverse event reports
The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.
InjuryMar 10, 2026Malposition of Device
REVISION SURGERY WAS PERFORMED 6 MONTHS AFTER THE PRIMARY PROCEDURE DUE TO MALPOSITIONING OF THE FEMORAL COMPONENT. THE SURGEON OBSERVED THAT THE FEMORAL COMPONENT HAD BEEN IMPLANTED WITH MALROTATION DURING THE PRIMARY SURGERY. DURING THE REVISION PROCEDURE, THE NATIVE PATELLA WAS REVISED, THE 12 MM INSERT WAS REPLACED WITH A 14 MM SEMI-CONSTRAINED INSERT, AND THE GMK-SPHERE SIZE 5 FEMORAL COMPONENT WAS REPLACED WITH A GMK-REVISION SIZE 5 FEMORAL COMPONENT. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Implant Pain; Loss of Range of Motion
MAUDE report 24558209 ↗InjuryMar 11, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AT ABOUT 1 YEAR AND 6 MONTHS FROM PRIMARY, THE PATIENT CAME IN REPORTING PAIN AND STIFFNESS. THE SURGEON REVISED THE INSERT AND FEMUR WITH A SC 17MM LINER AND A REVISION FEMUR BUT NO ISSUES WERE REPORTED WITH THE IMPLANTS. SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Loss of Range of Motion
MAUDE report 21570572 ↗InjuryFeb 25, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
ON (B)(6) 2016, THE PATIENT HAD KNEE PRIMARY SURGERY. ON (B)(6) 2023, THE PATIENT HAD AN ADDITIONAL SURGERY PERFORMED DUE TO INLAY LOCKING SCREW UNSCREWING (LEFT KNEE). THE INLAY LOCKING SCREW WAS REMOVED ARTHROSCOPICALLY. ON (B)(6) 2023, THE PATIENT UNDERWENT REVISION SURGERY OF THE LEFT KNEE DUE TO INFECTION. ONLY THE LINER WAS REVISED. ON (B)(6) 2023, THE PATIENT UNDERWENT A TOTAL KNEE REVISION SURGERY DUE TO INFECTION. THE EXACT DATE OF THE SURGERY, THE HOSPITAL, AND THE SURGEON ARE UNKNOWN.
Patient problems: Medical device site infection
MAUDE report 21458564 ↗InjuryJan 8, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
THE PATIENT CAME IN REPORTING INSTABILITY DUE TO LOOSE COMPONENTS AND THE CAUSE IS UNKNOWN. THE SURGEON REVISED THE FEMUR AND INSERT AND THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Joint Laxity
MAUDE report 21100349 ↗InjuryOct 8, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
REVISION SURGERY AT ABOUT 3 YEARS AND 1 MONTH POST PRIMARY DUE TO DISSATISFACTION POSTOPERATIVE, CHRONIC KNEE PAIN, NO DETECTED INFECTION. ACCORDING TO SURGEON DUE TO INSTABILITY OF THE KNEE. REMOVAL OF COMPONENTS AND CHANGE TO GMK HINGE.
Patient problems: Joint Laxity
MAUDE report 20396521 ↗InjuryAug 21, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
AT ABOUT 6 YEARS 11 MONTHS AFTER THE PRIMARY, THE PATIENT CAME IN REPORTING PAIN, DUE TO A LOOSE FEMORAL COMPONENT. THE SURGEON REVISED THE FEMORAL COMPONENT, POLY AND TIBIAL TRAY. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 20020770 ↗InjuryNov 2, 2023Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
REVISION SURGERY DUE TO LAXITY, AT ABOUT 3 YEARS 4 MONTHS FROM THE PRIMARY. THE PATIENT CAME IN REPORTING PAIN AND DISCOMFORT. DURING REVISION, AFTER LITTLE ITTING WITH A BONE TAMP, FEMORAL COMPONENT CEMENTED WAS EASY TO TAKE OUT. NO CEMENT ON BACKSURFACE. THE SURGEON REVISED SUCCESSFULLY THE FEMUR AND INSERT (10 TO 14 MM). THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Joint Laxity; Pain
MAUDE report 18054068 ↗InjurySep 13, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
THE PATIENT HAD A PRIMARY KNEE SURGERY ON (B)(6) 2020. ON (B)(6) 2023, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED SUCCESSFULLY THE POLY. ON (B)(6) 2023, THE PATIENT CAME IN REPORTING PAIN AFTER FALLING DIRECTLY ON THE KNEE, WHICH CAUSED THE INSERT TO DISENGAGE FROM THE BASEPLATE. THE SURGEON REVISED THE POLY AND THE SURGERY WAS COMPLETED SUCCESSFULLY. PRESENTLY, ON (B)(6) 2023, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT, REMOVED ALL IMPLANTS, AND IMPLANTED SUCCESSFULLY NEW COMPONENTS.
Patient problems: Unspecified Infection
MAUDE report 17734995 ↗InjurySep 6, 2023Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
AT 2 YEARS AND 9 MONTHS FROM THE PRIMARY, THE PATIENT CAME IN REPORTING PAIN DUE TO A LOOSE FEMORAL COMPONENT AND TIBIAL TRAY. THE SURGEON REVISED THE FEMORAL COMPONENT, INSERT, TIBIAL TRAY AND PATELLA COMPONENT. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Pain
MAUDE report 17689781 ↗InjurySep 26, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AT 1 DAY AFTER THE PRIMARY SURGERY, THE PATIENT REPORTED PAIN AND LAXITY. THE PCL (POSTERIOR CRUCIATE LIGAMENT) AND MCL (MEDIAL COLLATERAL LIGAMENT) WERE FOUND TO BE COMPROMISED, THEREFORE THE SURGEON DECIDED TO REVISE THE FEMORAL COMPONENT AND THE LINER TO SWITCH TO A CONSTRAINED PROSTHESIS.
Patient problems: Joint Laxity
MAUDE report 15484042 ↗InjuryApr 25, 2022Appropriate Device Problem Term/Code Not Available
THE PATIENT CAME IN REPORTING PAIN DUE TO LOOSE IMPLANTS AND THE CAUSE IS UNKNOWN. THE SURGEON REVISED ALL COMPONENTS WITH REVISION COMPONENTS 2 YEARS AND 1 MONTH AFTER PRIMARY. THE SURGERY WAS COMPLETED SUCCESSFULLY. THERE WAS NO CEMENT ADHERED TO THE IMPLANTS. THE CAUSE OF THIS IS UNKNOWN.
Patient problems: Pain
MAUDE report 14191915 ↗InjuryNov 7, 2021Appropriate Device Problem Term/Code Not Available
LEFT KNEE REVISION SURGERY DUE TO INFECTION WAS PERFORMED 1 YEAR AFTER THE PRIMARY SURGERY. HINGE SYSTEM IMPLANTED.
Patient problems: Unspecified Infection
MAUDE report 12758076 ↗InjuryApr 15, 2021Loose or Intermittent Connection
REVISION SURGERY PERFORMED DUE TO FEMUR AND TIBIAL TRAY LOOSENING 3 YEARS AND 10 MONTHS AFTER THE PRIMARY SURGERY.
Patient problems: Inadequate Osseointegration
MAUDE report 11670242 ↗InjuryDec 27, 2019Loosening of Implant Not Related to Bone-Ingrowth
ABOUT 1 YEAR AND 2 MONTHS AFTER PRIMARY THE PATIENT CAME IN COMPLAINING OF INSTABILITY. THE SURGEON OBSERVED THAT ALL THE COMPONENTS WERE LOOSE AND THAT THE CEMENT DID NOT HOLD DURING THE PRIMARY SURGERY. THE SURGEON REVISED THE FEMORAL COMPONENT, TIBIAL TRAY, POLY, AND PATELLA. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Joint Disorder
MAUDE report 9523296 ↗InjuryJun 12, 2019Appropriate Device Problem Term/Code Not Available
THE PATIENT CAME IN DUE TO SIGNS OF INFECTION 2 YEARS AND 10 MONTHS AFTER PRIMARY, THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND EXPLANTED THE FEMORAL COMPONENT, TIBIAL TRAY, POLY AND PATELLA. THE SURGEON IMPLANTED A TEMPORARY SPACER AND THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Unspecified Infection
MAUDE report 8691457 ↗InjuryApr 20, 2018Inadequacy of Device Shape and/or Size; Appropriate Device Problem Term/Code Not Available
REVISION SURGERY PERFORMED BECAUSE THE FEMORAL COMPONENT WAS TOO LARGE LIMITING THE ABILITY TO FLEX, 1 YEAR AND 2 MONTHS AFTER PRIMARY. THE FEMUR WAS EXPLANTED AND A SMALLER SIZE SPHERE WAS IMPLANTED ON ITS PLACE.
Patient problems: Joint Disorder
MAUDE report 7446081 ↗InjurySep 6, 2017Insufficient Device Problem Information
THE PATIENT CAME IN DUE TO SIGNS OF INFECTION. THE PATHOGEN IS CONFIRMED AS (B)(6). THE SURGEON REMOVED ALL MEDACTA HARDWARE AND IMPLANTED AN ANTIBIOTIC SPACER. THE SURGERY WAS COMPLETED SUCCESSFULLY. X-RAYS AND EXPLANTS ARE NOT AVAILABLE.
Patient problems: Unspecified Infection
MAUDE report 6844647 ↗
Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page