GMK Sphere Total Knee System 02.12.0006RRecalls & reported problems
Medacta International SA · Femur Sphere cemented Right, size 6 · Class 2 · Implant
All models of GMK Sphere Total Knee System (209)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 07630030825880), received between Oct 24, 2018 and Jun 9, 2026 — classified as 8 injuries.
The most frequently reported specific device problem is “Device Appears to Trigger Rejection” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Appears to Trigger Rejection3
- Unstable3
- Loosening of Implant Not Related to Bone-Ingrowth2
- Appropriate Device Problem Term/Code Not Available1
- Loose or Intermittent Connection1
Patient problems
- Medical device site infection3
- Joint Disorder1
- Joint Laxity1
- No Code Available1
- No Known Impact Or Consequence To Patient1
- Pain1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjuryJun 9, 2026Device Appears to Trigger Rejection
AT ABOUT 5 YEARS 1 MONTH AFTER THE PRIMARY, THE PATIENT CAME IN DUE TO SIGNS OF INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON REMOVED ALL MEDACTA IMPLANTS AND IMPLANTED ANTIBIOTIC SPACERS. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Medical device site infection
MAUDE report 25442545 ↗InjuryMar 3, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
AT 1 YEAR AND 3 MONTHS FROM MEDACTA PRIMARY SURGERY, THE PATIENT CAME IN DUE TO SIGNS OF INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON REMOVED ALL MEDACTA HARDWARE AND IMPLANTED AN ANTIBIOTIC SPACER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Medical device site infection
MAUDE report 21503937 ↗InjuryOct 15, 2024Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
FIRST STAGE KNEE REVISION FOR INFECTION AT ABOUT 10 YEARS AND 3 MONTHS POST PRIMARY. ALL IMPLANTS WERE REMOVED AND WASHOUT WAS PERFORMED. A COMPETITOR SPACER WAS IMPLANTED AS THE JOINT WAS INFECTED. AFTER A CHECK OF THE AGENT WITH THE HOSPITAL, NO FURTHER INFORMATION WILL BE AVAILABLE REGARDING THE 2ND STAGE REVISION OR THE PATIENT.
Patient problems: Medical device site infection
MAUDE report 20448872 ↗InjurySep 23, 2022Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
REVISION SURGERY AT ABOUT 2 YEARS AND 5 MONTHS POST PRIMARY FOR FEMUR AND TIBIA LOOSENING. ALL THE DEVICES HAVE BEEN REVISED SUCCESSFULLY.
Patient problems: Pain
MAUDE report 15473602 ↗InjuryJun 18, 2021Unstable
THE PATIENT CAME IN REPORTING MID-FLEXION INSTABILITY 1 YEAR AND 9 MONTHS AFTER THE PRIMARY SURGERY. THE SURGEON REVISED ALL IMPLANTS AND CONVERTED THE PATIENT FROM A SPHERE KNEE TO A HINGE KNEE. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Joint Laxity
MAUDE report 12024588 ↗InjuryAug 26, 2020Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
THE PATIENT CAME IN REPORTING INSTABILITY DUE TO A LOOSE TIBIA AND FEMUR. THE SURGEON OBSERVED THAT THERE WAS NO CEMENT PRESENT ON THE IMPLANTS. THE SURGEON REVISED THE FEMUR, TIBIA, AND INSERT 7 YEARS AND 1 MONTH AFTER PRIMARY. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10453190 ↗InjuryJun 24, 2020Loose or Intermittent Connection; Unstable; Appropriate Device Problem Term/Code Not Available
THE PATIENT CAME IN REPORTING INSTABILITY DUE TO A LOOSE FEMORAL COMPONENT. THE SURGEON REVISED ALL IMPLANTS AND CONVERTED THE GMK SPHERE KNEE TO A GMK HINGE KNEE 1 YEAR AND 11 MONTHS AFTER PRIMARY. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Joint Disorder
MAUDE report 10188722 ↗InjuryOct 24, 2018Unstable
CEMENT DIDN'T ADHERE TO IMPLANTS. THE TYPE OF CEMENT USED IS PALACOS R. ADDITIONAL COMMENT PERFORMED BY SURGEON: APPARENTLY A SURGEON HAS REVISED TWO OTHER PATIENTS OF MINE OVER THE LAST YEAR OR SO THAT WERE BOTH MEDACTA KNEES. IN BOTH OF THOSE CASES THERE WAS NO ADHERENCE OF CEMENT TO THE METAL COMPONENTS. FROM WHAT I UNDERSTAND, ONE OF THE KNEES WAS REVISED BECAUSE OF MID FLEXION INSTABILITY AND THE OTHER ONE HAD INCOMPETENCE OF THE MEDIAL COLLATERAL LIGAMENT.
Patient problems: No Code Available
MAUDE report 7999378 ↗
Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page