Skip to content

GMK Sphere Total Knee System 02.12.0214FLRecalls & reported problems

Medacta International SA · Tibial Insert Fixed SPHERE Flex #2/14 mm L · Class 2 · Implant

All models of GMK Sphere Total Knee System (209)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 07630030826214), received between Jan 26, 2018 and Aug 28, 2024 — classified as 6 injuries, 1 malfunction.

The most frequently reported specific device problem is “Device Slipped” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Appropriate Device Problem Term/Code Not Available2
  • Device Slipped1
  • Packaging Problem1
  • Unstable1

Patient problems

  • No Code Available2
  • Pain2
  • Joint Laxity1
  • No Information1
  • Unspecified Infection1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 28, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AT ABOUT 3 YEARS AND 9 MONTHS FROM THE PRIMARY, THE PATIENT CAME IN REPORTING KNEE INSTABILITY AND THE CAUSE IS UNKNOWN. THE SURGEON UPSIZED THE POLY (FROM 12 TO 20 MM) AND THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Joint Laxity

    MAUDE report 20085910 ↗
  • InjuryFeb 1, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT CAME IN REPORTING INSTABILITY AND ANTERIOR KNEE PAIN AND THE CAUSE IS UNKNOWN. AT ABOUT 8 MONTHS FROM PRIMARY THE SURGEON REVISED THE INSERT AND RESURFACED THE PATIENT'S NATURAL PATELLA. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Pain

    MAUDE report 16276067 ↗
  • InjuryNov 3, 2020Appropriate Device Problem Term/Code Not Available

    PATIENT WAS EXPERIENCING INSTABILITY AND ANTERIOR KNEE PAIN FROM CHRONIC TENDINOPATHY. 11 MONTH AFTER PRIMARY THE SURGEON REVISED THE GMK-SPHERE TIBIAL INSERT - FLEX S2L -14MM WITH A GMK-SPHERE TIBIAL INSERT - FLEX S2L - 20MM. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Pain

    MAUDE report 10778506 ↗
  • InjuryOct 30, 2020Unstable

    THE PATIENT CAME IN REPORTING INSTABILITY. THE CAUSE OF THE INSTABILITY IS UNKNOWN. 1 YEAR AND 2 MONTHS AFTER PRIMARY THE SURGEON REVISED THE GMK-SPHERETIBIAL INSERT - FLEX S2L - 14 MM WITH A GMK-SPHERETIBIAL INSERT - FLEX S2L - 17MM. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Code Available

    MAUDE report 10761314 ↗
  • MalfunctionMay 29, 2019Packaging Problem; Packaging Problem

    DURING THE PACKAGING OPENING IT WAS NOTICED THAT THE BAG CONTAINING THE INSERT INSIDE THE BLISTER WAS OPEN. THE SURGEON DECIDED TO USE A NEW IMPLANT TO COMPLETE THE SURGERY. THE IMPLANT WAS IN THE HOSPITAL AS CONSIGNMENT.

    Patient problems: No Information

    MAUDE report 8649826 ↗
  • InjuryJan 26, 2018Appropriate Device Problem Term/Code Not Available

    REVISION SURGERY PERFORMED DUE TO SUPERFICIAL INFECTION AT THE INCISION SITE, THAT COMMUNICATE WITH THE JOINT, 22 DAYS AFTER PRIMARY. THE SURGEON WASHED OUT THE INFECTION AND REPLACED THE POLY WITH A SIZE2X14MMM LEFT POLY. THE SURGEON INSERTED THE POLY INCORRECTLY (LOT 163355), AT WHICH TIME THE LOCKING MECHANISM WAS DAMAGED. AN INSERT FROM ANOTHER LOT WAS OPENED AND INSERTED WITHOUT FURTHER DELAY.

    Patient problems: Unspecified Infection

    MAUDE report 7221025 ↗
  • InjuryJan 26, 2018Device Slipped

    INLAY LOCKING SCREW ARTHROSCOPIC REMOVAL DUE TO UNSCREWING. TORQUE LIMITING SCREWDRIVER WAS NOT USED IN PRIMARY SURGERY.

    Patient problems: No Code Available

    MAUDE report 7221024 ↗

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page