GMK Sphere Total Knee System 02.12.0412FRRecalls & reported problems

Medacta International SA · Tibial Insert Fixed SPHERE Flex #4/12 mm R · Class 2 · Implant

All models of GMK Sphere Total Knee System (209)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Medacta GMK Sphere total knee implant is a cemented metal-and-plastic knee replacement with a fixed tibial insert. The 29 reports mostly describe knee looseness or instability, pain, and infection years after the original surgery, often leading to revision surgery to swap the insert or kneecap part. There were no deaths, 28 injuries, and 1 malfunction. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
29
Deaths reported
0
Injuries reported
28
Recalls naming it
0

Summary

The FDA's MAUDE database contains 29 adverse event reports that list this exact device (DI 07630030826580), received between Jul 7, 2017 and May 15, 2026 — classified as 28 injuries, 1 malfunction.

The most frequently reported specific device problem is “Device Appears to Trigger Rejection” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem10
  • Appropriate Device Problem Term/Code Not Available6
  • Device Appears to Trigger Rejection3
  • Unstable3
  • Loose or Intermittent Connection2
  • Break1
  • Detachment of Device or Device Component1
  • Device Issue1

Patient problems

  • Joint Laxity8
  • Unspecified Infection6
  • Pain5
  • Failure of Implant2
  • Implant Pain2
  • Joint Disorder2
  • Loss of Range of Motion2
  • Medical device site infection2

Recent adverse event reports

The 25 most recent of 29 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 15, 2026Adverse Event Without Identified Device or Use Problem

    AT ABOUT 2 YEARS AND 11 MONTHS FROM THE PRIMARY, THE PATIENT CAME IN PRESENTING ANTERIOR KNEE PAIN AND INSTABILITY AND THE CAUSE IS UNKNOWN. THERE WAS NO INDICATION OF TRAUMA. THE SURGEON REVISED THE PATIENT`S NATURAL PATELLA AND THE INSERTFOR STABILITY(FROM 12 TO 17 MM). THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Implant Pain; Joint Laxity

    MAUDE report 25185356 ↗
  • InjuryJul 1, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. ABOUT 3 YEARS AFTHER THE PRIMARY SURGERY, THE SURGEON PERFORMED A WASHOUT AND REVISED THE LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Medical device site infection

    MAUDE report 22369025 ↗
  • InjuryFeb 6, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AT ABOUT 4 YEARS 8 MONTHS AFTER THE PRIMARY, THE PATIENT CAME IN REPORTING INSTABILITY DUE TO LAXITY AND THE CAUSE IS UNKNOWN. THE SURGEON REVISED THE POLY (12 TO 17 MM) AND RESURFACED THE PATIENT'S NATURAL PATELLA. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Joint Laxity

    MAUDE report 21322605 ↗
  • InjuryNov 14, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AT ABOUT 3 YEARS AND 2 MONTHS AFTER PRIMARY, THE PATIENT CAME IN REPORTING PATELLA PAIN AND INSTABILITY AND THE CAUSE IS UNKNOWN. THE SURGEON REVISED THE PATELLA IMPLANT AND REVISED THE INSERT (12MM) WITH A THICKER ONE (14MM). THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Implant Pain; Joint Laxity

    MAUDE report 20681572 ↗
  • InjuryJun 28, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY 7 DAYS AFTER SURGERY. THE PATIENT HAD THE WOUND OPEN, BUT THE REASON OF IT IS UNKNOWN. DURING REVISION THE SURGEON WASHED OUT THE WOUND AND NOTICED THAT THE KNEE WAS UNSTABLE, SO HE DECIDED TO SWAP THE LINER INSERTING A 2MM THICKER ONE.

    Patient problems: Joint Laxity

    MAUDE report 19633343 ↗
  • InjuryDec 5, 2023Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size

    AT ABOUT 5 YEARS AND 3 MONTHS AFTER PRIMARY, THE PATIENT CAME IN REPORTING PAIN DUE TO FEELING TIGHTNESS AND A LIMITED RANGE OF MOTION AND THE CAUSE IS UNKNOWN. THE SURGEON REVISED SUCCESSFULLY THE PATELLA AND POLY (FROM 12MM TO 11MM). THE SURGERY WAS COMPLETED SUCCESSFULLY. THE SURGEON THOUGHT THE PATELLA WAS UNDERSIZED, BUT THE CAUSE WAS UNKNOWN.

    Patient problems: Loss of Range of Motion

    MAUDE report 18266622 ↗
  • InjuryNov 21, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AT ABOUT 1 YEAR 1 MONTH AFTER THE PRIMARY, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Medical device site infection

    MAUDE report 18181958 ↗
  • InjuryJul 26, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    AT ABOUT 2 MONTHS FROM THE PRIMARY, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Unspecified Infection

    MAUDE report 17396714 ↗
  • InjuryJul 19, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT HAD PRIMARY KNEE SURGERY ON (B)(6) 2010. ON (B)(6) 2020, THE PATIENT CAME IN REPORTING INSTABILITY DUE TO A LOOSE PATELLA AND TIBIA. THE SURGEON REVISED THE FEMORAL COMPONENT, TIBIAL TRAY, INSERT, AND PATELLA AND ADDED AN EXTENSION STEM AND CONE. THE SURGERY WAS COMPLETED SUCCESSFULLY. PRESENTLY, ON (B)(6) 2023, THE PATIENT CAME IN REPORTING PAIN AND THE CAUSE IS UNKNOWN. THE SURGEON OPENED THE PATIENT AND FOUND NO IMPLANTS TO BE LOOSE. HE REVISED THE POLY (WITH ONE OF THE SAME THICKNESS; NO INSTABILITY REPORTED) AND THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Pain

    MAUDE report 17348647 ↗
  • InjurySep 9, 2022Break; Break

    GMK SPHERE IMPLANTED APPROXIMATELY 6 YEARS AND 10 MONTHS AGO (12 AUG 2015) FOR OSTEOARTHRITIS. THE PATIENT PRESENTED WITH A PAINFUL KNEE, AND RADIOGRAPHS REVEALED THAT THE INSERT SCREW HAD BROKEN AND WAS LOOSE IN THE JOINT. THE SURGEON REMOVED THE PART OF THE SCREW LOOSE IN THE JOINT AND IMPLANTED A NEW INSERT (WITHOUT THE SCREW). ONLY HALF OF THE SCREW WAS RECOVERED, THE OTHER HALF IS STILL LODGED IN THE TIBIAL TRAY.

    Patient problems: Pain

    MAUDE report 15392133 ↗
  • InjurySep 8, 2022Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    REVISION PERFORMED 1 YEAR AND 2 MONTHS AFTER THE PRIMARY SURGERY DUE TO INFECTION. ONLY INSERT REVISED.

    Patient problems: Unspecified Infection

    MAUDE report 15376910 ↗
  • InjuryJul 18, 2022Appropriate Device Problem Term/Code Not Available

    REVISION SURGERY AT ABOUT 1 YEAR AND 2 MONTHS POST PRIMARY FOR KNEE INSTABILITY. THE SURGEON EXPLANTED THE LINER THICK 12MM AND IMPLANTED A THICKER LINER OF 17MM.

    Patient problems: Joint Laxity

    MAUDE report 15037411 ↗
  • InjuryJul 14, 2022Appropriate Device Problem Term/Code Not Available

    THE PATIENT CAME IN REPORTING A STIFF KNEE. AT 5 MONTHS AFTER THE PRIMARY SURGERY, THE SURGEON REVISED THE 12MM POLY TO A 10MM POLY AND THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Loss of Range of Motion

    MAUDE report 15011706 ↗
  • InjuryJun 9, 2022Appropriate Device Problem Term/Code Not Available

    AT ABOUT 3 MONTHS AND A HALF AFTER THE PRIMARY, REVISION SURGERY WAS PERFORMED DUE TO INFECTION. THE SURGEON REVISED ALL COMPONENTS AND THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Unspecified Infection

    MAUDE report 14641376 ↗
  • InjuryApr 25, 2022Unstable

    THE PATIENT CAME IN REPORTING INSTABILITY AND THE CAUSE OF THE INSTABILITY IS UNKNOWN. AT 1 YEAR AND 4 MONTHS POST PRIMARY THE SURGEON REVISED THE INSERT WITH A HIGHER ONE AND THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Joint Laxity

    MAUDE report 14191937 ↗
  • MalfunctionDec 22, 2021Migration; Migration

    INLAY LOCKING SCREW UNSCREWING, NO REVISION HAS BEEN PERFORMED AT THE MOMENT. PRIMARY WAS PERFORMED IN (B)(6) 2017.

    Patient problems: Failure of Implant

    MAUDE report 13058735 ↗
  • InjuryJul 2, 2021Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY PERFORMED TO DO A WASH OUT OF AN INFECTED KNEE TWO WEEKS AFTER PRIMARY SURGERY. 18 DAYS AFTER PRIMARY SURGERY THE TIBIAL INSERT WAS EXPLANTED AND EXCHANGED WITH AN INSERT OF THE SAME SIZE.

    Patient problems: Unspecified Infection

    MAUDE report 12106248 ↗
  • InjuryJun 24, 2021Adverse Event Without Identified Device or Use Problem

    THE PATIENT CAME IN REPORTING ANTERIOR KNEE PAIN AND THE CAUSE OF THE ANTERIOR KNEE PAIN IS UNKNOWN. 3 YEARS AND 5 MONTHS AFTER PRIMARY THE SURGERY RESURFACED THE PATIENT'S NATURAL PATELLA AND REVISED THE POLY. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Pain

    MAUDE report 12056289 ↗
  • InjuryJun 23, 2021Unstable

    THE PATIENT CAME IN REPORTING INSTABILITY DUE TO LAXITY AND THE CAUSE OF THE LAXITY IS UNKNOWN. 1 YEAR AND 2 MONTHS AFTER PRIMARY THE SURGEON REVISED THE GMK-SPHERE TIBIAL INSERT - FLEX S4L - 12 MM WITH A GMK-SPHERE TIBIAL INSERT - FLEX S4R - 17MM TO GIVE THE PATIENT MORE STABILITY. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Joint Laxity

    MAUDE report 12047794 ↗
  • InjuryMay 27, 2021Appropriate Device Problem Term/Code Not Available

    18 DAYS AFTER THE PRIMARY, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE POLY AND THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Unspecified Infection

    MAUDE report 11893144 ↗
  • InjuryJan 14, 2021Loose or Intermittent Connection

    THE PATIENT CAME IN REPORTING INSTABILITY DUE TO LAXITY, 2 YEARS 11 MONTHS AFTER THE PRIMARY. THE SURGEON REVISED THE 12MM POLY WITH A 17MM POLY AND THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Joint Laxity

    MAUDE report 11168792 ↗
  • InjuryDec 28, 2020Adverse Event Without Identified Device or Use Problem

    THE PATIENT CAME IN (2 YEARS AND 9 MONTHS AFTER PRIMARY SURGERY) REPORTING PAIN AND IT WAS OBSERVED THAT THE PATIENT HAD DEVELOPED ARTHROFIBROSIS (FIBROTIC TISSUE AROUND THE JOINT) WHICH LIMITED THE PATIENT'S FLEXION. THE CAUSE OF THE ARTHROFIBROSIS IS UNKNOWN. THE SURGEON CLEANED THE FIBROTIC TISSUE AND DOWNSIZED THE SPHERE INSERT FLEX RIGHT 12MM S4 TO A SPHERE INSERT FLEX RIGHT 11MM S4 TO OPEN UP THE JOINT AND GIVE THE PATIENT MORE ROOM TO BEND THE KNEE. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Fibrosis; Pain

    MAUDE report 11075683 ↗
  • InjuryDec 19, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    THE PATIENT HAD A PRIMARY KNEE SURGERY ON (B)(6) 2019. ON (B)(6) 2019, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION. THE SURGEON PERFORMED A WASHOUT AND POLY-SWAP AND THE SURGERY WAS COMPLETED SUCCESSFULLY. PRESENTLY, ON (B)(6) 2019, THE PATIENT CAME IN COMPLAINING OF PAIN AND IT WAS OBSERVED THAT THE POLY HAD BECOME DISENGAGED, NO SCREW WAS USED. THE SURGEON REVISED THE POLY AND THE SURGERY WAS COMPLETED SUCCESSFULLY USING THE SCREW.

    Patient problems: Pain

    MAUDE report 9494236 ↗
  • InjuryJul 3, 2019Appropriate Device Problem Term/Code Not Available

    REVISION SURGERY PERFORMED AFTER 1 YEAR AND 9 MONTHS AFTER THE PRIMARY DUE TO KNEE STIFFNESS. THE SURGEON REVISED THE 12MM POLY WITH A 10 MM POLY AND THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Limited Mobility Of The Implanted Joint

    MAUDE report 8757900 ↗
  • InjuryMay 22, 2019Adverse Event Without Identified Device or Use Problem

    5 DAYS AFTER PRIMARY THE SURGEON PERFORMED A REVISION SURGERY OF THE KNEE FOR AN HEMATOMA. THE SURGEON PERFORMED AN I&D AND POLY-SWAP. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Hematoma

    MAUDE report 8631628 ↗

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →