M.U.S.T. Polyaxial Pedicle Screw 03.50.018Recalls & reported problems

Medacta International SA · Pedicle screw 6x45mm + Nut (1x) - Sterile · Class 2 · Implant

All models of M.U.S.T. Polyaxial Pedicle Screw (25)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07630030834219), received between Feb 1, 2018 and Jul 5, 2019 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

1 other recall cover devices cleared under the same FDA submission (K121115); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Break1
  • Material Twisted/Bent1

Patient problems

  • No Known Impact Or Consequence To Patient1
  • Pain1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 5, 2019Material Twisted/Bent; Material Twisted/Bent

    PLIF SURGERY WAS PERFORMED AT L4-L5. THE SURGEON WAS NOT ABLE TO INSERT THE SET SCREWS INTO THE PEDICLE SCREW HEAD AT THE LEFT SIDE OF L4, SO THE SURGEON DISENGAGED THE SET SCREW, AND DISCOVERED CROSS-THREADING ON 2 SET SCREWS. EVEN THOUGH THE SURGEON REPLACED THE NEW SET SCREW, CROSS-THREADING REOCCURRED, SO THE SURGEON HAD TO REPLACE THE NEW PEDICLE SCREW. THE SAME EVENT ALSO OCCURRED AT THE RIGHT SIDE OF L4. DUE TO THIS EVENT THE SURGERY WAS PROLONGED ABOUT 20 MINUTES.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8764349 ↗
  • InjuryFeb 1, 2018Break; Break

    REVISION SURGERY PERFORMED TWO YEARS AFTER PRIMARY DUE TO PAIN CAUSED PROBABLY BY A PEDICLE SCREW BREAKAGE.

    Patient problems: Pain

    MAUDE report 7233960 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K121115. One submission often covers many models, so check the product and lot details in each record.

  • Z-1023-2017 · Dec 7, 2016 · Terminated
    Medacta International SA, M.U.S.T. Rod, REF 03.50.433, Size 5.5mm, Length 480mm, Type Straight, Mat Co-Cr-Mo The rods act as a connector between the different pedicle screws to

Other Orthosis, Spinal Pedicle Fixation devices

All Orthosis, Spinal Pedicle Fixation devices →

Sources and updates