- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07630030834295), received between Apr 10, 2018 and Jan 26, 2023 — classified as 3 injuries, 1 malfunction.
The most frequently reported specific device problem is “Osseointegration Problem” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K121115); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Osseointegration Problem2
- Break1
- Connection Problem1
Patient problems
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Device Embedded In Tissue or Plaque1
- Inadequate Osseointegration1
- Pain1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryJan 26, 2023Osseointegration Problem; Osseointegration Problem
AT ABOUT 2 YEARS AND 1 MONTH AFTER THE PRIMARY SPINE SURGERY, THE REVISION SURGERY WAS PERFORMED BECAUSE THE S SCREW (7X45) LOOSENED, CAME OUT FROM THE BONE, AND BONE FUSION DID NOT OCCUR. THE SURGERY WAS COMPLETED SUCCESSFULLY USING A 8X50 SCREW.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 16246868 ↗InjurySep 9, 2022Osseointegration Problem; Osseointegration Problem
ABOUT 1 YEAR AND 6 MONTHS AFTER THE PRIMARY SURGERY, A REVISION SURGERY WAS PERFORMED DUE TO L3/4/5 SCREWS LOOSENING AND L4/5 NOT FUSION. IT IS UNKNOWN WHICH SCREW CAME LOOSE. THE L3-5 SCREW WAS REMOVED TO REPLACE THE CAGE.
Patient problems: Inadequate Osseointegration
MAUDE report 15390881 ↗MalfunctionJul 4, 2018Break
SET SCREW (SPECIFIC LOT UNKNOWN) HAS BEEN BROKEN DURING TIGHTENING, SMALL PIECES REMAINED IN THE PATIENT'S WOUND. THE SET SCREW WAS REPLACED WITH A NEW ONE.WHEN THE SURGEON TIGHTENED THE SET SCREW AFTER IMPLANTATION OF THE PEDICLE SCREW 7X40 AT L5, HE DISCOVERED THE SET SCREW WAS BROKEN. HE REMOVED THE BROKEN PIECE OF THE SET SCREW, BUT SMALL PIECES (2 MM) REMAINED IN THE PATIENT BODY. THE SET SCREW WAS REPLACED WITH A NEW ONE, THE SURGEON DECIDED TO NOT RETRIEVE THE SMALL PIECES THAT REMAINED THE PATIENT.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 7662373 ↗InjuryApr 10, 2018Connection Problem
REVISION SURGERY 1 YEAR AND 1 MONTH AFTER PRIMARY SURGERY (LEVELS WERE L3-4) DUE TO PAIN, THE VERTEBRAE DIDN'T FUSE, THE SURGEON REVISED L3-4. THE SCREWS AND RODS (NOT LOOSE) WERE REMOVED AND REPLACED. THE INTERBODY REMAINED IN THE PATIENT.
Patient problems: Pain
MAUDE report 7412276 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K121115. One submission often covers many models, so check the product and lot details in each record.
- Z-1023-2017 · Dec 7, 2016 · TerminatedMedacta International SA, M.U.S.T. Rod, REF 03.50.433, Size 5.5mm, Length 480mm, Type Straight, Mat Co-Cr-Mo The rods act as a connector between the different pedicle screws to
Other Orthosis, Spinal Pedicle Fixation devices
- NA 04.614.508Synthes GmbH · ORTHOSIS, SPINAL PEDICLE FIXATION277 reports
- CREO 1134.0010Globus Medical, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION112 reports
- CREO 1119.0010Globus Medical, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION80 reports
- VIPER 186731745Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION43 reports
- CD HORIZON® Spinal System 1553201090Medtronic Sofamor Danek, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION38 reports
- EXPEDIUM 179902000Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION37 reports
- CD HORIZON® Spinal System 7540220Medtronic Sofamor Danek, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION36 reports
- EXPEDIUM VERSE 199721640Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION35 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page