Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07630030835360), received between Oct 4, 2017 and May 16, 2024 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

1 other recall cover devices cleared under the same FDA submission (K121115); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Device Packaging Compromised1
  • Fracture1

Patient problems

  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 16, 2024Fracture; Fracture

    REVISION SURGERY DUE TO ROD BREAKAGE (5.5 X 480 MM TI ROD) APPROXIMATELY 6 MONTHS AFTER THE LAST SURGERY. THE THORACIC PART OF THE PRIMARY RODS WAS LEFT IN THE PATIENT SINCE THERE WERE LAMINA BANDS ATTACHED TO IT. THE SURGEON THEN ADDED A ROD FROM THE COMPETITOR WHICH HAS A CONNECTOR ATTACHED TO IT TO THE THORACIC ROD. ADDITIONALLY, HE ADDED THE TWO PARALLEL RODS TO ADD MORE STABILITY. RODS AND SIDE-TO-SIDE CONNECTORS ARE FROM COMPETITORS. THE PATIENT HAD UNDERGONE 3 PREVIOUS SURGERIES TO EXTEND THE CONSTRUCT AND 2 PREVIOUS SURGERIES DUE TO INFECTION (NO MEDACTA DEVICES REVISED).

    Patient problems: Insufficient Information

    MAUDE report 19330278 ↗
  • InjuryNov 10, 2021Break

    THE PATIENT CAME IN FOR A POST-OP APPOINTMENT AND X-RAYS REVEALED THAT THE ROD WAS BROKEN. THE SURGEON HAD TO REMOVE THE ROD AND 3 SCREWS 2 YEARS AND 1 MONTH AFTER PRIMARY. THE SURGERY WAS COMPLETED SUCCESSFULLY. THE PATIENT WAS NOT IN ANY PAIN ONCE THE ROD BROKE. THE PATIENT WAS WORKING OUT 2-3 TIMES A DAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12786545 ↗
  • MalfunctionOct 4, 2017Device Packaging Compromised; Device Packaging Compromised

    THE INTERNAL PACKAGING WAS DAMAGED, THEREFORE THE STERILITY OF THE PRODUCT WAS COMPROMISED. THE SURGEON USED OTHER RODS TO COMPLETE THE SURGERY SUCCESSFULLY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6913759 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K121115. One submission often covers many models, so check the product and lot details in each record.

  • Z-1023-2017 · Dec 7, 2016 · Terminated
    Medacta International SA, M.U.S.T. Rod, REF 03.50.433, Size 5.5mm, Length 480mm, Type Straight, Mat Co-Cr-Mo The rods act as a connector between the different pedicle screws to

Other Orthosis, Spinal Pedicle Fixation devices

All Orthosis, Spinal Pedicle Fixation devices →