Evolis Total Knee System 3031.0205Recalls & reported problems
Medacta International SA · Cemented tibial baseplate size 5 · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07630030845284), received between Aug 20, 2018 and Jan 17, 2020 — classified as 2 injuries.
The most frequently reported specific device problem is “Loose or Intermittent Connection” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K081023); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Loose or Intermittent Connection2
- Improper or Incorrect Procedure or Method1
Patient problems
- Joint Disorder1
- No Code Available1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJan 17, 2020Loose or Intermittent Connection
REVISION SURGERY PERFORMED ON (B)(6) 2019 DUE TO A TIBIAL LOOSENING . PRIMARY SURGERY TOOK PLACE IN 2004, EXACT DATE UNKNOWN. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Joint Disorder
MAUDE report 9601766 ↗InjuryAug 20, 2018Loose or Intermittent Connection; Improper or Incorrect Procedure or Method
REVISION DUE TO TIBIAL TRAY LOOSENING AND SUBSIDENCE 5 YEARS AFTER PRIMARY. THE PATIENT HAVING A HIGH BMI; NO EXTENSION STEM WAS IMPLANTED IN THE PRIMARY CASE. THE FEMORAL COMPONENT WAS LEFT IN SITU. THE SALES REPRESENTATIVE WAS INFORMED ABOUT THE EVENT BY THE SURGEON ON 28 JUNE 2018. DUE TO HUMAN ERROR, THE SALES REPRESENTATIVE INFORMED THE HEADQUARTER ON 10 AUGUST 2018.
Patient problems: No Code Available
MAUDE report 7797182 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K081023. One submission often covers many models, so check the product and lot details in each record.
- Z-2095-2010 · Sep 11, 2009 · TerminatedEvolis-Femoral Sizer Posterior Reference The femoral sizer posterior reference is a manual orthopedic surgical instrument that is used to determine the trial size and set the exte
Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page