M.U.S.T. Polyaxial Pedicle Screw 03.52.323Recalls & reported problems
Medacta International SA · Enh. Cann. Pedicle Screw 6x40mm + Nut (1x) Sterile · Class 2 · Implant
All models of M.U.S.T. Polyaxial Pedicle Screw (25)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 07630030857584), received between Jul 31, 2018 and Sep 30, 2026 — classified as 6 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Fracture” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture3
- Appropriate Device Problem Term/Code Not Available2
- Adverse Event Without Identified Device or Use Problem1
- Difficult to Insert1
- Loose or Intermittent Connection1
- Loosening of Implant Not Related to Bone-Ingrowth1
- Material Deformation1
Patient problems
- No Known Impact Or Consequence To Patient3
- Implant Pain2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Failure of Implant1
- Medical device site infection1
- Pain1
- Unspecified Infection1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
InjurySep 30, 2026Fracture
During revision surgery for an unknown cause, the screw fractured during extraction at the junction between the shaft and the screw head.
Patient problems: Implant Pain
MAUDE report 50946479 ↗InjuryJul 17, 2026Adverse Event Without Identified Device or Use Problem
FEW DAYS AFTER THE PRIMARY LUMBAR FUSION WITH POSTERIOR SUPPLEMENTATION PERFORMED ON (B)(6) 2026, THE PATIENT DEVELOPED AN INFECTION. THE PATIENT RECEIVED TWO OTHER SURGERIES FOR SURGICAL CLEANING, ON (B)(6) 2026. NO DEVICES REVISED.
Patient problems: Medical device site infection
MAUDE report 25854818 ↗MalfunctionJul 9, 2024Fracture; Fracture
PRIMARY SURGERY WAS PERFORMED AT L4-L5. ABOUT 2 YEARS AND 9 MONTHS AFTER THE PRIMARY SURGERY, THE SURGEON WANTED TO EXTEND THE CONSTRUCT IN L3. DURING SURGERY, THE SURGEON DISCOVERED THAT THE HEAD OF THE PEDICLE SCREW BROKE AT THE RIGHT SIDE OF L4. THE SURGEON REMOVED THE BROKEN SCREW AND REPLACED IT WITH A LARGER PEDICLE SCREW (7MM X 40MM). THE ISSUE CAUSED 15 MINUTES DELAY. XRAYS ARE NOT AVAILABLE. NO TRAUMATIC EVENT OCCURRED. PRIMARY SURGERY PERFOMED WITH MYSPINE TECHNOLOGY CASE CODE: (B)(4).
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 19701565 ↗InjuryMar 1, 2023Fracture; Fracture
THE PATIENT CAME IN REPORTING PAIN DURING A POST-OP APPOINTMENT. THE SURGEON TOOK X-RAYS AND DECIDED TO SCHEDULE AN ADDITIONAL SURGERY AT ABOUT 3 YEARS 7 MONTHS AFTER THE PRIMARY TO EXTEND THE PATIENT'S CONSTRUCT. DURING THE ADDITIONAL SURGERY, IT WAS OBSERVED THAT THE L5 MUST MC SCREW WAS BROKEN (TULIP HEAD CAME OFF). THE SURGEON COULD NOT REMOVE THE REST OF THE SCREW AND PROCEEDED TO EXTEND THE PATIENTS CONSTRUCT. THERE WAS NO DELAY AND THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Implant Pain
MAUDE report 16459487 ↗InjuryMay 22, 2020Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
THE PATIENT HAD A MYSPINE MC SURGERY IN (B)(6) 2020 AND WAS REVISED ONE DAY AFTER DUE TO PAIN, A CAGE HAS BEEN REMOVED. INITIALLY, A DECOMPRESSION WAS ON ONE SIDE, DURING THE REVISION SURGERY IT WAS DONE ALSO ON THE SECOND SIDE. ON (B)(6) OF 2020 THE PATIENT UNDERWENT REVISION SURGERY AGAIN DUE TO PEDICLE SCREWS LOOSENING. THE PATIENT HAS OSTEOPOROTIC BONE.
Patient problems: Pain
MAUDE report 10081873 ↗InjuryDec 13, 2019Appropriate Device Problem Term/Code Not Available
1 MONTH POST PRIMARY THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE SET SCREWS AND ROD. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Unspecified Infection
MAUDE report 9465003 ↗InjuryNov 26, 2019Loose or Intermittent Connection
THE ENH. POLY-AXIAL PEDICLE SCREW - CANNULATED 6X40MM WAS INSERTED, BUT THE SCREW LOOSENING WAS CONFIRMED AFTER INSERTING THE SCREW COMPLETELY INTO THE BONE. ONE POSSIBLE REASON WAS THE PEDICLE SCREW THREAD WAS NOT FULLY DESIGNED NEAR THE PEDICLE SCREW TULIP, SO THE THREAD COULD NOT BE FIXED AT THE CORTICAL BONE COMPLETELY AND IT WAS NOT ABLE TO OBTAIN THE FIXING FORCE. PEDICLE SCREW WAS REPLACED WITH THE BACKUP PEDICLE SCREW. DUE TO THIS EVENT THE SURGERY WAS PROLONGED ABOUT 20 MINUTES. PATIENT BONE QUALITY WAS NOT GOOD.
Patient problems: Failure of Implant
MAUDE report 9374815 ↗MalfunctionNov 14, 2018Material Deformation; Material Deformation
THE NURSE WAS NOT ABLE TO ENGAGE THE PEDICLE SCREW WITH THE SCREWDRIVER. AFTER MANY ATTEMPTS, DEBRIS WAS GENERATED ON THE PEDICLE SCREW THREAD. THE SURGERY WAS FINISHED WITH THE BACK UP IMPLANT. NO CRITICAL DELAY. THE SAME SCREWDRIVER WAS ABLE TO ENGAGE WITH THE BACK UP IMPLANT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8069031 ↗MalfunctionSep 28, 2018Appropriate Device Problem Term/Code Not Available
THE SURGEON WAS NOT ABLE TO TIGHTEN THE SET SCREW INTO THE PEDICLE SCREW TULIP AT L5 EVEN THOUGH THE HE USED WITH THE REDUCER. HE TRIED TO REMOVE THE ROD AND INSERT THE NEW SET SCREW, BUT HE WAS NOT ABLE TO TIGHTEN THE NEW SET SCREW AS WELL. DISASSEMBLING THE PERCUTANEOUS TOWER AND RE-TRYING AGAIN HE WAS ABLE TO TIGHTEN THE SECOND SET SCREW. DUE TO THIS EVENT THE SURGERY WAS PROLONGED ABOUT 1 HOUR.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7918477 ↗MalfunctionJul 31, 2018Difficult to Insert
WHEN THE SURGEON WAS TRYING TO INSERT THE SET SCREW INTO THE PEDICLE SCREW HEAD WHICH WAS ALREADY IMPLANTED AT L5, IT WAS NOT ABLE TO INSERT, EVEN THOUGH THE SURGEON USED THE REDUCER. THEN HE TRIED TO INSERT WITHOUT THE ROD, BUT IT WAS NOT ABLE. SINCE IT WAS DOUBTFUL WHETHER THE ISSUE WAS WITH PEDICLE SCREW HEAD OR PERCUTANEOUS TOWER, THE PEDICLE SCREW WAS REPLACED WITH A MORE LARGE SIZE. THE SURGERY WAS PERFORMED BY PPS AND FINISHED SUCCESSFULLY. DUE TO THIS EVENT THE SURGERY WAS PROLONGED ABOUT 30 MINUTES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7734728 ↗
Other Appliance, Fixation, Spinal Intervertebral Body devices
- CD HORIZON® SOLERA® VOYAGER™ Spinal System 6550017Medtronic Sofamor Danek, Inc. · Appliance, fixation, spinal intervertebral body131 reports
- EVEREST® Spinal System 2901-10001VB Spine LLC · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY119 reports
- ZEVO™ Anterior Cervical Plate System 7710015Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall1 report
- XLP 7565050Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- Elegance® Anterior Cervical Plate 53430344Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- ATLANTIS® Anterior Cervical Plate System 876-170Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- XLP 7565060Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- Elegance® Anterior Cervical Plate 53430113Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page