Mpact Double Mobility Shells 01.32.156MBRecalls & reported problems
Medacta International SA · Double Mobility Acetabular shell Ø56 · Class 2 · Implant
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 07630030860997), received between Jul 21, 2017 and Oct 23, 2024 — classified as 8 injuries.
The most frequently reported specific device problem is “Osseointegration Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Appropriate Device Problem Term/Code Not Available2
- Osseointegration Problem2
- Device Slipped1
- Failure to Osseointegrate1
- Malposition of Device1
- Positioning Problem1
Patient problems
- Inadequate Osseointegration4
- Joint Disorder2
- Joint Laxity1
- Pain1
- Unspecified Infection1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjuryOct 23, 2024Osseointegration Problem; Osseointegration Problem
THE PATIENT HAD HIP INSTABILITY AND THE CAUSE WAS UNKNOWN. AT ABOUT 5 YEARS AND 1 MONTH POST PRIMARY UPON OPENING THE PATIENT, IT WAS OBSERVED THAT THE CUP WAS LOOSE AND THE CAUSE OF THE LOOSE CUP IS UNKNOWN. THE SURGEON REVISED THE CUP, HEAD, AND LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Inadequate Osseointegration; Joint Laxity
MAUDE report 20512683 ↗InjuryJun 12, 2024Osseointegration Problem; Osseointegration Problem
AT ABOUT 4 YEARS AND 4 MONTHS POST PRIMARY REVISION SURGERY FOR MPACT DM LOOSENING. ACETABULAR COMPONENTS AND HEAD WERE REVISED SUCCESSFULLY.
Patient problems: Inadequate Osseointegration
MAUDE report 19517427 ↗InjurySep 19, 2023Malposition of Device; Malposition of Device
AT ABOUT 1 YEAR AND 2 MONTHS FROM THE PRIMARY, THE PATIENT CAME IN REPORTING PAIN DUE TO AN ANTEVERTED CUP AND THE CAUSE IS UNKNOWN. THE SURGEON REVISED THE MEDACTA CUP AND LINER WITH COMPETITOR COMPONENTS AND REVISED THE MEDACTA HEAD WITH A MEDACTA HEAD. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Pain
MAUDE report 17773622 ↗InjuryJul 7, 2023Failure to Osseointegrate; Failure to Osseointegrate
AT ABOUT 1 YEAR AFTER THE PRIMARY, THE PATIENT CAME IN REPORTING PAIN DUE TO A LOOSE CUP AND THE CAUSE IS UNKNOWN. THE SURGEON REVISED SUCCESSFULLY THE MEDACTA CUP, LINER AND HEAD.
Patient problems: Inadequate Osseointegration
MAUDE report 17276824 ↗InjuryAug 27, 2020Appropriate Device Problem Term/Code Not Available
THE PATIENT CAME IN AFTER 1 YEAR AND 5 MONTHS FROM THE PRIMARY FOR A POST-OP APPOINTMENT AND THE SURGEON OBSERVED THAT THE ACETABULAR SHELL WAS NOT WELL FIXED WITHIN THE ACETABULUM. THE SURGEON DECIDED TO REVISE THE PATIENT AND REVISED THE SHELL, LINER, AND FEMORAL CERAMIC HEAD. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Joint Disorder
MAUDE report 10458657 ↗InjuryOct 31, 2019Positioning Problem
THE PATIENT CAME IN COMPLAINING OF PAIN, AND THE CAUSE OF THE PAIN WAS A DISLOCATION. POST-OP X-RAYS INDICATED THAT THE CUP WAS PLACED TOO LATERAL WHICH CAUSED A DISLOCATION. THE SURGEON REVISED THE CUP 1 DAY AFTER PRIMARY SURGERY. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Joint Disorder
MAUDE report 9260078 ↗InjuryJun 14, 2019Device Slipped
REVISION SURGERY PERFORMED AFTER 1,5 MONTH DUE TO PAIN CAUSED BY A LOOSE CUP. THE SURGEON REVISED THE 56MM MPACT DM CUP WITH A 58MM MPACT TWO HOLE CUP AND ALSO REVISED THE HEAD AND LINER. TE PATIENT WAS DIABETIC. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Inadequate Osseointegration
MAUDE report 8699599 ↗InjuryJul 21, 2017Appropriate Device Problem Term/Code Not Available
THE PATIENT WAS REVISED FOR INFECTION. THE SURGEON REVISED ALL MEDACTA COMPONENTS AND IMPLANTED A SPACER. THE SURGERY WAS COMPLETED SUCCESSFULLY. X-RAYS AND EXPLANTS ARE NOT AVAILABLE.
Patient problems: Unspecified Infection
MAUDE report 6731941 ↗
Other Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented devices
- PINNACLE 1217-25-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED735 reports
- NA 1365-52-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED683 reports
- APEX 1246-03-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED3 deaths613 reports
- PINNACLE 1217-20-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED2 deaths590 reports
- PINNACLE 1217-30-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 death484 reports
- NA 1365-53-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED404 reports
- PINNACLE ALTRX 1221-36-054Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED388 reports
- PINNACLE ALTRX 1221-32-050Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED386 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.