GMK Sphere Total Knee System 02.12.0025LRecalls & reported problems
Medacta International SA · Femur Sphere cemented Left, size 5+ · Class 2 · Implant
All models of GMK Sphere Total Knee System (209)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 16
- Deaths reported
- 0
- Injuries reported
- 16
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 07630030862410), received between Apr 30, 2019 and May 30, 2025 — classified as 16 injuries.
The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Loosening of Implant Not Related to Bone-Ingrowth5
- Adverse Event Without Identified Device or Use Problem3
- Appropriate Device Problem Term/Code Not Available3
- Device Appears to Trigger Rejection2
- Device Dislodged or Dislocated1
- Malposition of Device1
- Microbial Contamination of Device1
Patient problems
- Unspecified Infection6
- Implant Pain2
- Joint Laxity2
- Medical device site infection2
- Pain2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Aseptic loosening1
Recent adverse event reports
The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.
InjuryMay 30, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
AT ABOUT 8 YEARS AND 2 MONTHS AFTER PRIMARY, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT, REMOVED ALL IMPLANTS AND IMPLANTED AN ANTIBIOTIC SPACER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Medical device site infection
MAUDE report 22114074 ↗InjuryApr 3, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
AT ABOUT 2 YEARS AND 8 MONTHS AFTER PRIMARY, THE PATIENT CAME IN REPORTING INSTABILITY. DURING THE REVISION, THE SURGEON OBSERVED THAT THE FEMORAL COMPONENT WAS LOOSE DUE TO LACK OF CEMENT ADHERENCE. THE SURGEON REVISED THE FEMUR, INSERT AND PATELLA. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Joint Laxity
MAUDE report 21760519 ↗InjuryFeb 6, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
REVISION SURGERY FOR KNEE LOOSENING AT ABOUT 6 YEARS AND 10 MONTHS POST-PRIMARY DUE TO FEMORAL LOOSENING. DURING THE REVISION, IT WAS FOUND THAT THE TIBIAL IMPLANT WAS LOOSE TOO. THERE WERE NO SIGNS OF CEMENT ON THE REVISED IMPLANTS.
Patient problems: Aseptic loosening
MAUDE report 21321420 ↗InjuryOct 28, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
AT ABOUT 2 YEARS AFTER THE PRIMARY, THE PATIENT HAS BEEN REVISED BECAUSE OF FEMORAL AND TIBIAL COMPONENTS MOBILIZATION. IT WAS REPORTED THAT THE PATIENT STARTED TO EXPERIENCE PROBLEMS IMMEDIATELY AFTER THE PRIMARY. DURING THE REVISION, THE FEMORAL IMPLANT CAME OFF IMMEDIATELY AND A LITTLE AMOUNT OF CEMENT REMAINED ON THE CONDYLE OF THE DEVICE. ALL IMPLANTS HAVE BEEN REVISED SUCCESSFULLY.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 20545183 ↗InjuryMay 7, 2024Malposition of Device; Malposition of Device
THE PATIENT CAME IN REPORTING PAIN AND REVISION SURGERY WAS PERFORMED AT ABOUT 5 MONTHS AFTER PRIMARY DUE TO THE FEMORAL COMPONENT MALPOSITIONING. PATIENT PRIMARY SURGERY WAS PERFORMED USING PATIENT SPECIFIC MYKNEE GUIDES AND, AT THE END OF THE SURGERY FROM THE POST-OP X-RAYS, IT WAS DETECTED THAT THE FEMORAL COMPONENT WAS IMPLANTED TOO MUCH IN FLEXION WITH RESPECT TO THE PLANNED CUT.
Patient problems: Implant Pain
MAUDE report 19257376 ↗InjuryOct 2, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
REVISION SURGERY FOR INFECTION AT 1 YEAR AND 11 MONTHS POST PRIMARY AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT, REMOVED ALL IMPLANTS, AND IMPLANTED AN ANTIBIOTIC SPACER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Medical device site infection
MAUDE report 17852274 ↗InjuryFeb 1, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT CAME IN REPORTING INSTABILITY AND THE CAUSE IS UNKNOWN. ABOUT 1 YEAR AND 2 MONTHS AFTER THE PRIMARY SURGERY, THE SURGEON REVISED ALL COMPONENTS AND SWITCH TO A REVISION SYSTEM, AND THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Pain
MAUDE report 16276069 ↗InjuryJan 13, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated
DURING REHAB AT 4 MONTHS FROM PRIMARY, PATELLAR DISLOCATION WAS EXPERIENCED. FEMUR, PATELLA IMPLANT AND INSERT REVISED SUCCESSFULLY.
Patient problems: Pain
MAUDE report 16157251 ↗InjuryOct 24, 2022Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
THE PATIENT CAME IN REPORTING PAIN DUE TO ASEPTIC LOOSENING AND THE CAUSE IS UNKNOWN. ABOUT 10 MONTHS AFTER THE PRIMARY SURGERY, THE SURGEON REVISED ALL COMPONENTS EXCEPT THE PATELLA AND THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Implant Pain
MAUDE report 15657369 ↗InjuryDec 13, 2021Loosening of Implant Not Related to Bone-Ingrowth
THE PATIENT CAME IN FOR A POST-OP APPOINTMENT 10 MONTHS AFTER THE PRIMARY SURGERY AND IT WAS DETERMINED THAT THE PATIENT NEEDED TO BE CONVERTED BECAUSE THE QUADRICEP WAS NOT STRONG ENOUGH AND WAS CAUSING THE KNEE TO HYPEREXTEND. ALL COMPONENTS WERE REVISED AND THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Joint Laxity
MAUDE report 12982990 ↗InjuryMay 19, 2021Adverse Event Without Identified Device or Use Problem
2 YEARS 3 MONTHS AFTER THE PRIMARY THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS ESCHERICHIA COLI. THE SURGEON PERFORMED A WASHOUT, REMOVED ALL COMPONENTS, AND IMPLANTED AN ANTIBIOTIC SPACER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Unspecified Infection
MAUDE report 11847282 ↗InjuryMar 15, 2021Adverse Event Without Identified Device or Use Problem
THE PATIENT CAME IN 1 YEAR AND 5 MONTHS AFTER THE PRIMARY SURGERY DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT, REMOVED ALL COMPONENTS AND IMPLANTED AN ANTIBIOTIC SPACER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Unspecified Infection
MAUDE report 11481956 ↗InjuryAug 27, 2020Appropriate Device Problem Term/Code Not Available
THE PATIENT CAME IN AFTER 5 MONTHS FROM THE LAST REVISION SURGERY DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REMOVED ALL COMPONENTS AND IMPLANTED NEW COMPONENTS THE SURGERY WAS COMPLETED SUCCESSFULLY. THIS PATIENT HAD 2 REVISION SURGERY PREVIOUSLY DUE TO AN INFECTION WHERE ONLY THE INSERT WAS SWAPPED (FOR BOTH CASES). THIS REVISION SURGERY HAPPENED 2 YEARS AND 7 MONTHS AFTER PRIMARY SURGERY.
Patient problems: Unspecified Infection
MAUDE report 10458489 ↗InjuryFeb 12, 2020Appropriate Device Problem Term/Code Not Available
THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON REMOVED ALL HARDWARE AND IMPLANTED AN ANTIBIOTIC SPACER ALMOST 1 YEAR AFTER PRIMARY. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Unspecified Infection
MAUDE report 9698515 ↗InjuryNov 5, 2019Appropriate Device Problem Term/Code Not Available
REVISION SURGERY PERFORMED AFTER 1 YEAR AFTER THE PRIMARY DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON REVISED THE FEMUR, TIBIA AND INSERT COMPONENTS AND IMPLANTED AN ARTICULATING SPACER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Unspecified Infection
MAUDE report 9278377 ↗InjuryApr 30, 2019Microbial Contamination of Device
ON APRIL 1ST WE WERE ALERTED OF A KNEE INFECTION, THE REVISION SURGERY WAS PERFORMED SUCCESSFULLY ON APRIL 3RD, ABOUT 1 YEAR AFTER PRIMARY. ALL HARDWARE WERE EXPLANTED AND AN ANTIBIOTIC SPACER WAS PUT IN. THE PATIENTS WAS FIRST NOTED TO HAVE AN INFECTION AT THE 4TH WEEK POST SURGERY. THE FIRST SWAB RESULTS CAME BACK AS A STAPH AUREUS INFECTION. THE PATIENT HAS HAD MULTIPLE SURGERIES SINCE TO WASH THE KNEE OUT AND TO TRY TO REMOVE THE INFECTION.
Patient problems: Unspecified Infection
MAUDE report 8564993 ↗
Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page