GMK Sphere Total Knee System 02.12.0021RRecalls & reported problems

Medacta International SA · Femur Sphere cemented Right, size 1+ · Class 2 · Implant

All models of GMK Sphere Total Knee System (209)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07630030862458), received between Sep 4, 2019 and May 22, 2024 — classified as 4 injuries.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available2
  • Device Dislodged or Dislocated1
  • Loosening of Implant Not Related to Bone-Ingrowth1

Patient problems

  • Joint Dislocation1
  • Joint Laxity1
  • Pain1
  • Rupture1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 22, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    REVISION SURGERY DUE TO KNEE LUXATION (FEMORAL COMPONENT FROM THE LINER), AT ABOUT 1 MONTH AFTER PRIMARY. THE PATIENT HAD A DOMESTIC INCIDENT AND DURING REHABILITATION THE FEMORAL COMPONENT DISLOCATED. THE SURGEON REVISED ALL COMPONENTS TO HINGE COMPONENTS.

    Patient problems: Joint Dislocation

    MAUDE report 19371341 ↗
  • InjuryNov 22, 2023Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    REVISION SURGERY DUE TO KNEE INSTABILITY AND LOOSENING OF TIBIAL AND FEMORAL COMPONENTS, AT ABOUT 4 YEARS 9 MONTHS FROM THE PRIMARY. ALL COMPONENTS REVISED SUCCESSFULLY TO HINGE SYSTEM.

    Patient problems: Joint Laxity

    MAUDE report 18190074 ↗
  • InjurySep 22, 2020Appropriate Device Problem Term/Code Not Available

    THE PATIENT CAME IN REPORTING INSTABILITY DUE TO DEVELOPING A MEDIAL COLLATERAL LIGAMENT RUPTURE. THE CAUSE OF THE RUPTURE IS UNKNOWN. 11 MONTHS AFTER PRIMARY THE SURGEON REVISED THE FEMORAL COMPONENT, TIBIAL TRAY, AND INSERT. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Rupture

    MAUDE report 10567153 ↗
  • InjurySep 4, 2019Appropriate Device Problem Term/Code Not Available

    REVISION SURGERY PERFORMED AFTER 2 YEARS AND 2 MONTHS FROM THE PRIMARY DUE TO PAIN AND THE CAUSE OF PAIN IS UNKNOWN. THE SURGEON REVISED THE FEMORAL COMPONENT, TIBIAL TRAY AND INSERT (10 MM EXCHANGED WITH LINER 17 MM) SUCCESSFULLY.

    Patient problems: Pain

    MAUDE report 8959262 ↗

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →

Sources and updates