GMK Sphere Total Knee System 02.12.0021LRecalls & reported problems
Medacta International SA · Femur Sphere cemented Left, size 1+ · Class 2 · Implant
All models of GMK Sphere Total Knee System (209)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07630030862519), received between Nov 16, 2018 and May 25, 2021 — classified as 2 injuries.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Insufficient Device Problem Information1
- Loose or Intermittent Connection1
Patient problems
- Joint Disorder1
- Loss of Range of Motion1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryMay 25, 2021Insufficient Device Problem Information
THE PATIENT CAME IN REPORTING KNEE PAIN AND STIFFNESS. THE CAUSE OF THE KNEE PAIN AND STIFFNESS IS UNKNOWN. 5 MONTHS AFTER PRIMARY THE SURGEON REVISED THE FEMORAL COMPONENT AND INSERT AND ADDED AN EXTENSION STEM. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Loss of Range of Motion
MAUDE report 11877173 ↗InjuryNov 16, 2018Loose or Intermittent Connection
REVISION SURGERY PERFORMED AFTER 2 YEARS FROM THE PRIMARY DUE TO TIBIAL PLATEAU LOOSENING, ALL HARDWARE WERE REMOVED, THE FEMUR WAS WELL FIXED. THE CEMENT WAS NOT ADHERENT TO THE INTERNAL SURFACES OF BOTH COMPONENTS REMOVED.
Patient problems: Joint Disorder
MAUDE report 8077523 ↗
Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page