GMK Sphere Total Knee System 02.12.0411FLRecalls & reported problems

Medacta International SA · Tibial Insert Fixed SPHERE Flex #4/11 mm L · Class 2 · Implant

All models of GMK Sphere Total Knee System (209)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Medacta GMK Sphere tibial insert is a plastic liner used in total knee replacement. The 24 reports that name its identifier describe knee infections needing washout and liner replacement, and knee instability needing a thicker liner, with 24 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
24
Deaths reported
0
Injuries reported
24
Recalls naming it
0

Summary

The FDA's MAUDE database contains 24 adverse event reports that list this exact device (DI 07630030862601), received between Aug 18, 2017 and Aug 6, 2025 — classified as 24 injuries.

The most frequently reported specific device problem is “Device Appears to Trigger Rejection” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available7
  • Adverse Event Without Identified Device or Use Problem5
  • Device Appears to Trigger Rejection5
  • Unstable4
  • Loose or Intermittent Connection3
  • Break1
  • Loosening of Implant Not Related to Bone-Ingrowth1
  • Unintended Movement1

Patient problems

  • Joint Disorder6
  • Medical device site infection6
  • Joint Laxity4
  • Unspecified Infection3
  • Failure of Implant2
  • Pain2
  • Insufficient Information1
  • Wound Dehiscence1

Recent adverse event reports

The 24 most recent of 24 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 6, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    AT ABOUT 3 YEARS 4 MONTHS AFTER THE PRIMARY, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Medical device site infection

    MAUDE report 22726217 ↗
  • InjuryJun 5, 2024Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    REVISION FOR KNEE INFECTION AND THE PATHOGEN IS UNKNOWN. AT 2 YEARS AND 10 MONTHS POST PRIMARY THE SURGEON PERFORMED A WASHOUT AND REVISED THE LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Medical device site infection

    MAUDE report 19465945 ↗
  • InjuryNov 22, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT CAME IN REPORTING INSTABILITY AND THE CAUSE OF THE INSTABILITY IS UNKNOWN. AT ABOUT 9 MONTHS AFTER PRIMARY, THE SURGEON REVISED THE INSERT WITH A THICKER ONE (FROM 11 TO 13MM). THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Joint Laxity

    MAUDE report 18191307 ↗
  • InjuryAug 18, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    AT ABOUT 1 MONTH AFTER THE PRIMARY, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED SUCCESSFULLY THE LINER.

    Patient problems: Medical device site infection

    MAUDE report 17570816 ↗
  • InjuryJul 13, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT CAME IN REPORTING INSTABILITY AND THE CAUSE IS UNKNOWN. ABOUT 1 YEAR AND 10 MONTHS AFTER THE PRIMARY SURGERY, THE SURGEON REVISED THE LINER AND THE SURGERY WAS COMPLETED SUCCESSFULLY (FROM 11 MM TO 14 MM).

    Patient problems: Joint Laxity

    MAUDE report 17315592 ↗
  • InjuryMay 25, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    AT ABOUT 2 MONTHS AFTER THE PRIMARY SURGERY, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Medical device site infection

    MAUDE report 17001451 ↗
  • InjuryApr 13, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    AT ABOUT 2 WEEKS HE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Medical device site infection

    MAUDE report 16731991 ↗
  • InjuryMar 21, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. AT 1 MONTH FROM PRIMARY THE SURGEON REMOVED ALL COMPONENTS AND IMPLANTED AN ANTIBIOTIC SPACER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Medical device site infection

    MAUDE report 16583372 ↗
  • InjuryJul 15, 2022Loose or Intermittent Connection; Appropriate Device Problem Term/Code Not Available

    THE PATIENT HAD A PRIMARY KNEE SURGERY ON (B)(6) 2020. ON (B)(6) 2021, THE PATIENT CAME IN REPORTING INSTABILITY DUE TO LAXITY AND THE CAUSE WAS UNKNOWN. THE SURGEON REVISED THE GMK-SPHERE TIBIAL INSERT WITH A THICKER ONE TO GIVE THE PATIENT MORE STABILITY. THE SURGERY WAS COMPLETED SUCCESSFULLY. PRESENTLY, (B)(6) 2022, THE PATIENT CAME IN REPORTING INSTABILITY DUE TO LAXITY. THE SURGEON REVISED GMK-SPHERE TIBIAL INSERT. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Insufficient Information

    MAUDE report 15027950 ↗
  • InjuryJan 25, 2022Loose or Intermittent Connection; Appropriate Device Problem Term/Code Not Available

    THE PATIENT CAME IN REPORTING INSTABILITY DUE TO LAXITY. THE SURGEON REVISED THE 11MM POLY WITH A 17MM POLY 2 YEARS AND 6 MONTHS AFTER PRIMARY. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Joint Laxity

    MAUDE report 13342906 ↗
  • InjuryDec 10, 2021Break

    THE PATIENT CAME IN FOR A POST-OP APPOINTMENT AND POST-OP X-RAYS INDICATED THAT THE POLY HAD DISENGAGED FROM THE TIBIAL TRAY AND THE CAUSE IS UNKNOWN. THE SURGEON REVISED THE INSERT 11 MONTHS AFTER PRIMARY AND THE SURGERY WAS COMPLETED SUCCESSFULLY. THE LINER SCREW WAS NOT USED IN PRIMARY SURGERY.

    Patient problems: Failure of Implant

    MAUDE report 12975553 ↗
  • InjuryFeb 24, 2021Unstable

    THE PATIENT CAME IN REPORTING INSTABILITY 10 MONTHS AFTER THE PRIMARY. THE CAUSE OF THE INSTABILITY IS UNKNOWN. THE SURGEON REVISED THE TIBIAL INSERT FIXED SPHERE FLEX #4/11 MM L WITH A SPHERE TIBIAL INSERT - FLEX S4L 14 MM. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Joint Laxity

    MAUDE report 11372951 ↗
  • InjuryJan 27, 2021Adverse Event Without Identified Device or Use Problem

    THE PATIENT CAME IN 9MONTHS AFTER THE PRIMARY SURGERY REPORTING PAIN AND THE CAUSE OF THE PAIN IS UNKNOWN. THE SURGEON REVISED THE SPHERE INSERT FLEX LEFT 11MM S4 WITH A GMK-SPHERE TIBIAL INSERT - FLEX S4L - 14MM. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Pain

    MAUDE report 11235030 ↗
  • InjuryOct 1, 2020Appropriate Device Problem Term/Code Not Available

    THE PATIENT CAME IN, 2 YEARS AND 3 MONTHS AFTER PREVIOUS REVISION SURGERY, DUE TO SIGNS OF INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND SWAPPED THE POLY. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Unspecified Infection

    MAUDE report 10612479 ↗
  • InjuryAug 10, 2020Unstable; Appropriate Device Problem Term/Code Not Available

    THE PATIENT CAME IN REPORTING INSTABILITY 9 MONTHS AFTER THE PRIMARY SURGERY. THE CAUSE OF THE INSTABILITY IS UNKNOWN. THE SURGEON REVISED THE FEMORAL COMPONENT, PATELLA, AND POLY. THE SURGERY WAS COMPLETED SUCCESSFULLY. THE FEMORAL COMPONENT AND PATELLA WERE NOT MOBILIZED, THE NEW LINER IS 12MM.

    Patient problems: Joint Disorder

    MAUDE report 10384112 ↗
  • InjuryMar 16, 2020Appropriate Device Problem Term/Code Not Available

    THE PATIENT CAME IN DUE TO SIGNS OF INFECTION 1 YEAR AND 2 MONTHS AFTER THE PRIMARY, THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND POLY-SWAP AND THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Unspecified Infection

    MAUDE report 9834573 ↗
  • InjuryNov 13, 2019Appropriate Device Problem Term/Code Not Available

    REVISION SURGERY PERFORMED 24 DAYS AFTER THE PRIMARY DUE TO WOUND DEHISCENCE. THE SURGEON PERFORMED A WASHOUT AND POLY-SWAP AND INSERTED ANTIBIOTIC BEADS. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Wound Dehiscence

    MAUDE report 9314362 ↗
  • InjuryJul 9, 2019Appropriate Device Problem Term/Code Not Available

    THE PATIENT CAME IN, MORE THAN ONE YEAR AFTER PRIMARY SURGERY, COMPLAINING OF MID FLEXION INSTABILITY. THE SURGEON REVISED A 11MM POLY WITH A 14MM POLY AND THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Joint Disorder

    MAUDE report 8770629 ↗
  • InjuryApr 25, 2019Loosening of Implant Not Related to Bone-Ingrowth

    THE PATIENT CAME IN, 8 MONTHS AFTER PRIMARY SURGERY, COMPLAINING OF INSTABILITY DUE TO LAXITY. THE SURGEON REVISED THE 11MM POLY WITH A 13MM POLY AND THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Joint Disorder

    MAUDE report 8548593 ↗
  • InjuryMar 11, 2019Unstable

    THE PATIENT CAME IN COMPLAINING OF INSTABILITY 10 MONTHS AFTER PRIMARY SURGERY AND THE CAUSE OF INSTABILITY IS UNKNOWN. THE SURGEON REVISED THE 11MM POLY WITH A 17MM POLY. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Joint Disorder

    MAUDE report 8408123 ↗
  • InjuryJan 2, 2019Loose or Intermittent Connection

    THE PATIENT CAME IN DUE TO INSTABILITY CAUSED BY THE SCREW BACKING OUT OF THE POLY. THE TORQUE LIMITING DRIVER WAS NOT USED IN THE PRIMARY SURGERY. THE SURGEON REVISED THE POLY AND THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Joint Disorder

    MAUDE report 8208001 ↗
  • InjuryMay 8, 2018Unstable

    THE PATIENT CAME IN COMPLAINING OF INSTABILITY 1 YEAR AND 1 MONTH AFTER PRIMARY. THE PATIENT STRETCHED OUT HIS KNEE THROUGH REHAB. THE SURGEON SWAPPED THE 10MM POLY FOR A 12MM POLY. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Joint Disorder

    MAUDE report 7494714 ↗
  • InjuryMay 3, 2018Unintended Movement

    THE PATIENT CAME IN COMPLAINING OF PAIN. THE SURGEON DETERMINED THE SCREW BACKED OUT OF THE SPHERE POLY. THE TORQUE DRIVER WAS NOT USED IN THE PRIMARY SURGERY. THE SURGEON REVISE THE SCREW ON (B)(6) 2018.

    Patient problems: Failure of Implant; Pain

    MAUDE report 7483274 ↗
  • InjuryAug 18, 2017Adverse Event Without Identified Device or Use Problem

    THE PATIENT CAME IN DUE TO SIGNS OF INFECTION. THE PATHOGEN IS NOT AVAILABLE. THE SURGEON WASHED OUT THE KNEE AND REVISED THE INSERT. THE SURGERY WAS COMPLETED SUCCESSFULLY. X-RAYS AND EXPLANTS ARE NOT AVAILABLE.

    Patient problems: Unspecified Infection

    MAUDE report 6803617 ↗

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →