Skip to content

Masterloc Femoral Stems 01.39.202Recalls & reported problems

Medacta International SA · Masterloc Cementless Ti Coated LAT Stem # 2 · Class 2 · Implant

All models of Masterloc Femoral Stems (21)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07630030885624), received between Aug 30, 2018 and Oct 24, 2019 — classified as 3 injuries.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Fracture1
  • Inadequacy of Device Shape and/or Size1
  • Loosening of Implant Not Related to Bone-Ingrowth1
  • Unstable1

Patient problems

  • Bone Fracture(s)1
  • Joint Disorder1
  • Pain1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 24, 2019Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    THE PATIENT CAME IN COMPLAINING OF PAIN DUE TO NO BONE INGROWTH ON THE STEM. ABOUT 1 YEAR AND 9 MONTHS AFTER PRIMARY THE SURGEON REVISED SUCCESSFULLY THE STEM AND CERAMIC HEAD.

    Patient problems: Pain

    MAUDE report 9233222 ↗
  • InjuryJun 5, 2019Unstable

    REVISION SURGERY PERFORMED AFTER 6 MONTHS FROM THE PRIMARY DUE TO INSTABILITY CAUSED BY A SUBSIDED STEM. THE SURGEON REVISED THE STEM, LINER AND HEAD. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Joint Disorder

    MAUDE report 8669897 ↗
  • InjuryAug 30, 2018Fracture; Inadequacy of Device Shape and/or Size

    THE PATIENT CAME IN COMPLAINING OF PAIN 21 DAYS AFTER PRIMARY SURGERY. THE SURGEON DETERMINED THAT THE PATIENT HAD A FRACTURED FEMUR. THE SURGEON DETERMINED THAT THE STEM WAS UNDERSIZED IN THE PRIMARY. THE SURGEON REVISED THE STEM AND HEAD. THE SURGERY WAS COMPLETED SUCCESSFULLY

    Patient problems: Bone Fracture(s)

    MAUDE report 7832918 ↗

Other Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented devices

All Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page