Skip to content

STRAUMANN DENTAL IMPLANT SYSTEM SLA Titanium BL (REF 021.4414)Recalls & reported problems

Institut Straumann AG · Bone Level Implant, D 4.1 mm, RC, SLA 14 mm, Ti · Class 2 · Implant

All models of STRAUMANN DENTAL IMPLANT SYSTEM (128)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Straumann SLA Titanium dental implant is a bone level implant. It has 197 adverse event reports, with no deaths and 197 injuries. Most describe the implant failing to bond with the bone or losing its bond, and some mention infection and gum disease around the implant. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
205
Deaths reported
0
Injuries reported
205
Recalls naming it
0

Summary

The FDA's MAUDE database contains 205 adverse event reports that list this exact device (DI 07630031700346), received between Sep 12, 2019 and Sep 14, 2026 — classified as 205 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (118 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 3 reportsMay 2024: 2 reportsJun 2024: 2 reportsJul 2024: 2 reportsAug 2024: 1 reportsSep 2024: 5 reportsOct 2024: 3 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 0 reportsJun 2025: 9 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 2 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 3 reportsJan 2026: 2 reportsFeb 2026: 1 reportsMar 2026: 5 reportsApr 2026: 3 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 8 reportsOct 2026: 0 reports2024202520269
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate118
  • Loss of Osseointegration81
  • Fracture7

Patient problems

  • No Clinical Signs, Symptoms or Conditions101
  • Tissue Damage37
  • Unspecified Infection37
  • Fibrosis30
  • Pain21
  • Hemorrhage/Blood Loss/Bleeding15
  • Inflammation12
  • Abscess11

Recent adverse event reports

The 25 most recent of 205 reports, in the reporter's own words as published by the FDA.

  • InjurySep 14, 2026Loss of Osseointegration

    Patient problems: Fibrosis

    MAUDE report 50759553 ↗
  • InjurySep 14, 2026Loss of Osseointegration

    Patient problems: Fibrosis

    MAUDE report 50759548 ↗
  • InjurySep 14, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2012 in ADA 12. On (b)(6) 2026, loss of osseointegration was verified. Patient presented with fair oral hygiene. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50757603 ↗
  • InjurySep 14, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2012 in ADA 11. On (b)(6) 2026, loss of osseointegration was verified. Patient presented with fair oral hygiene. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50757599 ↗
  • InjurySep 7, 2026Loss of Osseointegration

    Patient problems: Abscess; Unspecified Infection

    MAUDE report 50653838 ↗
  • InjurySep 7, 2026Loss of Osseointegration

    Patient problems: Abscess; Unspecified Infection

    MAUDE report 50653837 ↗
  • InjurySep 7, 2026Loss of Osseointegration

    Patient problems: Abscess; Unspecified Infection

    MAUDE report 50652946 ↗
  • InjurySep 7, 2026Loss of Osseointegration

    Patient problems: Abscess; Unspecified Infection

    MAUDE report 50652938 ↗
  • InjuryJul 9, 2026Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2009 IN FDI 21. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFECTION, PERI-IMPLANTITIS, SWELLING AND BLEEDING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Swelling/ Edema; Unspecified Infection

    MAUDE report 25767633 ↗
  • InjuryMay 21, 2026Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2014 IN FDI 11. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFECTION AND PERI-IMPLANTITIS. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Unspecified Infection

    MAUDE report 25245664 ↗
  • InjuryApr 27, 2026Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2023 IN FDI 35. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED AND IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25006086 ↗
  • InjuryApr 27, 2026Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN FDI 44. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED AND IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25005997 ↗
  • InjuryApr 14, 2026Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2013 IN FDI 15. DETAILS OF SURGERY: SINUS AUGMENTATION. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH POOR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS, PAIN, MOBILITY AND BLEEDING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Pain; Unspecified Infection

    MAUDE report 24875880 ↗
  • InjuryMar 30, 2026Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2013 IN FDI 15. DETAILS OF SURGERY: SINUS AUGMENTATION. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH POOR ORAL HYGIENE. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS, PAIN, MOBILITY AND BLEEDING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Pain; Unspecified Infection

    MAUDE report 24734424 ↗
  • InjuryMar 26, 2026Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2022, IN FDI 43. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24704675 ↗
  • InjuryMar 26, 2026Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2022 IN FDI 24. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24704612 ↗
  • InjuryMar 26, 2026Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2022 IN FDI 35. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24704610 ↗
  • InjuryMar 16, 2026Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2015 IN THE PATIENT'S MOUTH. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24609010 ↗
  • InjuryFeb 24, 2026Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2014 IN THE PATIENT'S MOUTH. DETAILS OF SURGERY: SINUS AUGMENTATION. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS AND FISTULA. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fistula; Unspecified Infection

    MAUDE report 24431964 ↗
  • InjuryJan 29, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN FDI 46. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24213047 ↗
  • InjuryJan 26, 2026Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2023 IN FDI 28. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24181789 ↗
  • InjuryDec 29, 2025Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2023 IN FDI 31. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23914512 ↗
  • InjuryDec 29, 2025Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2021 IN FDI 12. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23914503 ↗
  • InjuryDec 8, 2025Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2023 IN FDI 14. ON (B)(6) 2025 LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23749099 ↗
  • InjurySep 29, 2025Fracture; Fracture

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2016 IN THE PATIENT'S MOUTH. ON (B)(6) 2025, FRACTURE OF THE IMPLANT WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23173088 ↗

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices

All IMPLANT, ENDOSSEOUS, ROOT-FORM devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page