TL 04.1MM RN,S, SLACTIVE TIZR 6, 8, 10, 12, 14, 16MM DENTAL IMPLANTS SLActive Roxolid TL (REF 033.561S)Recalls & reported problems

Institut Straumann AG · Standard Plus Implant, D 4.1mm RN, SLActive® 8mm, Roxolid®, Loxim™ · Class 2 · Implant

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Straumann SLActive Roxolid dental implant appears in 10,097 adverse event reports, all of them injury reports, with no deaths. The reports mostly describe implants that fail to bond with the jawbone or later lose that bond, with some mentioning looseness, pain, peri-implantitis and inflammation. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
10,182
Deaths reported
0
Injuries reported
10,182
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10,182 adverse event reports that list this exact device (DI 07630031700889), received between Aug 28, 2019 and Sep 28, 2026 — classified as 10,182 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (5,983 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 152 reportsJan 2024: 103 reportsFeb 2024: 131 reportsMar 2024: 120 reportsApr 2024: 168 reportsMay 2024: 136 reportsJun 2024: 149 reportsJul 2024: 155 reportsAug 2024: 118 reportsSep 2024: 122 reportsOct 2024: 169 reportsNov 2024: 256 reportsDec 2024: 109 reportsJan 2025: 136 reportsFeb 2025: 121 reportsMar 2025: 167 reportsApr 2025: 30 reportsMay 2025: 0 reportsJun 2025: 385 reportsJul 2025: 142 reportsAug 2025: 130 reportsSep 2025: 63 reportsOct 2025: 117 reportsNov 2025: 116 reportsDec 2025: 163 reportsJan 2026: 146 reportsFeb 2026: 126 reportsMar 2026: 159 reportsApr 2026: 130 reportsMay 2026: 143 reportsJun 2026: 107 reportsJul 2026: 128 reportsAug 2026: 129 reportsSep 2026: 82 reportsOct 2026: 0 reportsNov 2026: 0 reports202420252026385
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate5,983
  • Loss of Osseointegration3,921
  • Adverse Event Without Identified Device or Use Problem232
  • Difficult to Insert38
  • Fracture35
  • Separation Failure9
  • Unintended Movement7
  • Difficult to Remove6

Patient problems

  • No Clinical Signs, Symptoms or Conditions4,453
  • Fibrosis3,518
  • Pain1,924
  • Tissue Damage1,168
  • Unspecified Infection1,045
  • Inflammation475
  • Hemorrhage/Blood Loss/Bleeding391
  • Swelling/ Edema316

Recent adverse event reports

The 25 most recent of 10,182 reports, in the reporter's own words as published by the FDA.

  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50921876 ↗
  • InjurySep 28, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2008 in FDI 14. On (b)(6) 2026, loss of osseointegration was verified. Patient presented with fair oral hygiene and inadequate bone quality/quantity. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50921574 ↗
  • InjurySep 28, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2019 in FDI 37. On (b)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Peri-implantitis, Mobility and Bleeding. No further patient complications were reported.

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Unspecified Infection

    MAUDE report 50921483 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2025 in FDI 27. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50921364 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. On (b)(6) 2026, non-osseointegration was verified. No product was returned to the manufacturer. At the event the patient experienced: Infection. No further patient complications were reported.

    Patient problems: Unspecified Infection

    MAUDE report 50921239 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6)2026 in the patient's mouth. Details of surgery: Primary stability not achieved. On (b)(6)2026, non-osseointegration was verified. No product was returned to the manufacturer. At the event the patient experienced: Pain and Mobility. No further patient complications were reported.

    Patient problems: Fibrosis; Pain

    MAUDE report 50921222 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. On (b)(6) 2026, non-osseointegration was verified. No product was returned to the manufacturer. At the event the patient experienced: Peri-implantitis. No further patient complications were reported.

    Patient problems: Unspecified Infection

    MAUDE report 50921215 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. On (b)(6) 2026, non-osseointegration was verified. No product was returned to the manufacturer. At the event the patient experienced: Infection. No further patient complications were reported.

    Patient problems: Unspecified Infection

    MAUDE report 50921214 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. Details of surgery: Primary stability not achieved. On (b)(6) 2026, non-osseointegration was verified. No product was returned to the manufacturer. At the event the patient experienced: Infection. No further patient complications were reported.

    Patient problems: Unspecified Infection

    MAUDE report 50921213 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 46. Details of surgery: Implant surface not completely covered with bone. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50921180 ↗
  • InjurySep 28, 2026Loss of Osseointegration

    The clinician reports the implant was inserted on (b)(6) 2019 in FDI 26. On (b)(6) 2026, loss of osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. At the event the patient experienced: Peri-implantitis, Pain and Mobility. No further patient complications were reported.

    Patient problems: Fibrosis; Pain; Unspecified Infection

    MAUDE report 50921038 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 16. Details of surgery: Primary stability not achieved, Implant surface not completely covered with bone and Sinus augmentation. On (b)(6) 2026, non-osseointegration was verified. Patient presented with inadequate bone quality/quantity. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility and Swelling. No further patient complications were reported.

    Patient problems: Fibrosis; Swelling/ Edema

    MAUDE report 50920681 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6)2026 in FDI 44. On (b)(6)2026, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50920636 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2025 in FDI 26. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50920632 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 11. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50920579 ↗
  • InjurySep 28, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6)2021 in the patient's mouth. On (b)(6)2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50919927 ↗
  • InjurySep 28, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2021 in the patient's mouth. On (b)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50919924 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6)2026 in FDI 14. On (b)(6)2026, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Increased Sensitivity. No further patient complications were reported.

    Patient problems: Hyperresponsive to stimuli

    MAUDE report 50919486 ↗
  • InjurySep 25, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2026(b)(6) in FDI 44. On 2026(b)(6), non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Pain and Swelling. No further patient complications were reported.

    Patient problems: Pain; Swelling/ Edema

    MAUDE report 50901721 ↗
  • InjurySep 21, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 37. On (b)(6) 2026, non-osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50845101 ↗
  • InjurySep 21, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. On (b)(6) 2026, loss of osseointegration was verified. Patient presented with bone type III and inadequate bone quality/quantity. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50844917 ↗
  • InjurySep 21, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2026(b)(6) in the patient's mouth. On 2026(b)(6), non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50844340 ↗
  • InjurySep 21, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. On (b)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Peri-implantitis. No further patient complications were reported.

    Patient problems: Unspecified Infection

    MAUDE report 50844339 ↗
  • InjurySep 21, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2025 in FDI 26. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50844169 ↗
  • InjurySep 21, 2026Loss of Osseointegration

    The clinician reports the implant was inserted 2021(b)(6) in FDI 37. On 2026(b)(6), loss of osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50844168 ↗

FDA clearance and classification

Cleared through 510(k) K122855 on Dec 12, 2012 (decision: Substantially Equivalent). Submitted by Straumann USA as “Tl 04.1mm Rn,S, Slactive Tizr 6, 8, 10, 12, 14, 16mm Dental Implants”. The FDA review took 85 days. FDA record ↗

The FDA regulates product code DZE (implant, endosseous, root-form) under 21 CFR 872.3640 as a Class II device. 21 CFR 872.3640 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the TL 04.1MM RN,S, SLACTIVE TIZR 6, 8, 10, 12, 14, 16MM DENTAL IMPLANTS SLActive Roxolid TL

Has the TL 04.1MM RN,S, SLACTIVE TIZR 6, 8, 10, 12, 14, 16MM DENTAL IMPLANTS SLActive Roxolid TL been recalled?

No FDA recall record in our data names this exact device identifier (DI 07630031700889), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the TL 04.1MM RN,S, SLACTIVE TIZR 6, 8, 10, 12, 14, 16MM DENTAL IMPLANTS SLActive Roxolid TL?

The FDA's MAUDE database lists 10,182 adverse event reports for this device, received between Aug 28, 2019 and Sep 28, 2026. In those reports, the most common specific device problems are “Failure to Osseointegrate” (5,983), “Loss of Osseointegration” (3,921) and “Difficult to Insert” (38); the most common patient problems are “Fibrosis” (3,518), “Pain” (1,924) and “Tissue Damage” (1,168). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the TL 04.1MM RN,S, SLACTIVE TIZR 6, 8, 10, 12, 14, 16MM DENTAL IMPLANTS SLActive Roxolid TL?

None of the reports are classified as deaths, 10,182 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the TL 04.1MM RN,S, SLACTIVE TIZR 6, 8, 10, 12, 14, 16MM DENTAL IMPLANTS SLActive Roxolid TL FDA approved?

It was cleared through 510(k) K122855 on Dec 12, 2012. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the TL 04.1MM RN,S, SLACTIVE TIZR 6, 8, 10, 12, 14, 16MM DENTAL IMPLANTS SLActive Roxolid TL MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the TL 04.1MM RN,S, SLACTIVE TIZR 6, 8, 10, 12, 14, 16MM DENTAL IMPLANTS SLActive Roxolid TL, and is it still sold?

The FDA's device database lists Institut Straumann AG as the company responsible for it (brand TL 04.1MM RN,S, SLACTIVE TIZR 6, 8, 10, 12, 14, 16MM DENTAL IMPLANTS, model SLActive Roxolid TL). Its record shows it as in commercial distribution.