This Straumann Roxolid dental implant is placed in the jawbone. Its 656 reports mostly describe the implant failing to bond with the surrounding bone, or losing that bond over time. Listed patient problems include scar-like tissue build-up, pain and tissue damage. The reports include 656 injuries and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 656
- Deaths reported
- 0
- Injuries reported
- 656
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 656 adverse event reports that list this exact device (DI 07630031706560), received between Sep 9, 2019 and Sep 10, 2026 — classified as 656 injuries.
The most frequently reported specific device problem is “Failure to Osseointegrate” (404 reports).
2 other recalls cover devices cleared under the same FDA submission (K150938); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Osseointegrate404
- Loss of Osseointegration172
- Adverse Event Without Identified Device or Use Problem54
- Difficult to Insert12
- Fracture12
- Difficult to Remove4
- Activation, Positioning or Separation Problem2
- Defective Device2
Patient problems
- Fibrosis218
- No Clinical Signs, Symptoms or Conditions207
- Pain159
- Tissue Damage115
- Unspecified Infection110
- Inflammation31
- Hemorrhage/Blood Loss/Bleeding30
- Swelling/ Edema24
Recent adverse event reports
The 25 most recent of 656 reports, in the reporter's own words as published by the FDA.
InjurySep 10, 2026Failure to Osseointegrate
Patient problems: Hyperresponsive to stimuli
MAUDE report 50719923 ↗InjuryAug 24, 2026Loss of Osseointegration
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50415090 ↗InjuryAug 13, 2026Loss of Osseointegration
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50239265 ↗InjuryAug 3, 2026Loss of Osseointegration
The clinician reports the implant was inserted (b)(6) 2004 in FDI 31. On (b)(6) 2026, loss of osseointegration was verified. No product was returned to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.
Patient problems: Fibrosis
MAUDE report 50120279 ↗InjuryJul 27, 2026Loss of Osseointegration
The clinician reports the implant was inserted (b)(6) 2020 in FDI 12. Details of surgery: Implant surface not completely covered with bone and Immediate extraction site. On (b)(6) 2026, loss of osseointegration was verified. Patient presented with bone type IV and fair oral hygiene. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50010432 ↗InjuryJun 26, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN ADA 11. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25639591 ↗InjuryJun 26, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN ADA 11. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25639112 ↗InjuryJun 25, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN THE PATIENT'S MOUTH. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fibrosis
MAUDE report 25626169 ↗InjuryJun 22, 2026Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2018 IN FDI 21. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE AND SINUS AUGMENTATION. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fibrosis
MAUDE report 25582334 ↗InjuryJun 22, 2026Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2019 IN ADA 25. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25582073 ↗InjuryJun 22, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN FDI 13. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25581661 ↗InjuryJun 22, 2026Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2019 IN ADA 25. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25580750 ↗InjuryJun 19, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN FDI 22. DETAILS OF SURGERY: IMMEDIATE EXTRACTION SITE. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE AND PREVIOUS BONE AUGMENTATION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25564740 ↗InjuryJun 15, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 22. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Pain
MAUDE report 25507601 ↗InjuryJun 8, 2026Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 44. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Pain
MAUDE report 25431943 ↗InjuryJun 8, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN FDI 11. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25431870 ↗InjuryJun 3, 2026Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN THE PATIENT'S MOUTH. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25379474 ↗InjuryJun 3, 2026Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN THE PATIENT'S MOUTH. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25379459 ↗InjuryMay 18, 2026Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN THE PATIENT'S MOUTH. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25207844 ↗InjuryMay 18, 2026Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2018 IN THE PATIENT'S MOUTH. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Unspecified Infection
MAUDE report 25207801 ↗InjuryMay 6, 2026Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN FDI 33. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fibrosis
MAUDE report 25100848 ↗InjuryMay 6, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED AND IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN, MOBILITY AND HYPERSENSITIVITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fibrosis; Hypersensitivity/Allergic reaction; Pain
MAUDE report 25100823 ↗InjuryApr 16, 2026Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2018 IN THE PATIENT'S MOUTH. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24898074 ↗InjuryApr 9, 2026Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2022 IN THE PATIENT'S MOUTH. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24834068 ↗InjuryApr 7, 2026Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2021 IN THE PATIENT'S MOUTH. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24807304 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K150938. One submission often covers many models, so check the product and lot details in each record.
- Z-2534-2016 · Jul 14, 2016 · TerminatedStraumann Bone Level Implant, 4.1mm RC, SLA 10mm, TiZr, NTP Article Number: 021.4510
- Z-2535-2016 · Jul 14, 2016 · TerminatedStraumann Bone Level Implant, 4.1mm RC, SLA 12mm, TiZr, NTP Article Number: 021.4512
Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM43,616 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM40,404 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM27,400 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,621 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLA Titanium BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,298 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM23,448 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,320 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,267 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page