STRAUMANN DENTAL IMPLANT SYSTEM SLA Titanium BL (REF 021.2612)Recalls & reported problems

Institut Straumann AG · Bone Level Implant, D 3.3mm NC, SLA® 12mm, Ti, Loxim™ · Class 2 · Implant

All models of STRAUMANN DENTAL IMPLANT SYSTEM (128)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Straumann dental implant is a titanium screw placed in the jawbone to hold a replacement tooth. Most reports describe implants that did not bond with the surrounding bone, and some mention scar tissue, infection, pain or inflammation at the site. The records include 1,999 reports and 1,999 injuries, with no deaths, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
2,014
Deaths reported
0
Injuries reported
2,014
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2,014 adverse event reports that list this exact device (DI 07630031706591), received between Sep 9, 2019 and Sep 28, 2026 — classified as 2,014 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (1,778 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 56 reportsDec 2023: 62 reportsJan 2024: 28 reportsFeb 2024: 46 reportsMar 2024: 38 reportsApr 2024: 45 reportsMay 2024: 35 reportsJun 2024: 49 reportsJul 2024: 45 reportsAug 2024: 26 reportsSep 2024: 50 reportsOct 2024: 28 reportsNov 2024: 42 reportsDec 2024: 23 reportsJan 2025: 35 reportsFeb 2025: 18 reportsMar 2025: 48 reportsApr 2025: 13 reportsMay 2025: 0 reportsJun 2025: 145 reportsJul 2025: 35 reportsAug 2025: 42 reportsSep 2025: 27 reportsOct 2025: 15 reportsNov 2025: 44 reportsDec 2025: 32 reportsJan 2026: 31 reportsFeb 2026: 27 reportsMar 2026: 9 reportsApr 2026: 40 reportsMay 2026: 16 reportsJun 2026: 23 reportsJul 2026: 42 reportsAug 2026: 21 reportsSep 2026: 10 reportsOct 2026: 0 reports202420252026145
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate1,778
  • Loss of Osseointegration173
  • Adverse Event Without Identified Device or Use Problem37
  • Fracture34
  • Difficult to Insert2
  • Separation Failure2
  • Unstable2
  • Therapeutic or Diagnostic Output Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1,402
  • Fibrosis275
  • Unspecified Infection167
  • Pain140
  • Inflammation133
  • Tissue Damage110
  • Swelling/ Edema50
  • Hemorrhage/Blood Loss/Bleeding43

Recent adverse event reports

The 25 most recent of 2,014 reports, in the reporter's own words as published by the FDA.

  • InjurySep 28, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50921959 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50920943 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50920939 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50920932 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50920868 ↗
  • InjurySep 17, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50807499 ↗
  • InjurySep 17, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 32. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50806684 ↗
  • InjurySep 14, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50760667 ↗
  • InjurySep 14, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50760134 ↗
  • InjurySep 7, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50654070 ↗
  • InjuryAug 27, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6)2025 in FDI 25. Details of surgery: Primary stability not achieved and Implant surface not completely covered with bone. On (b)(6)2026, non-osseointegration was verified. Patient presented with fair oral hygiene. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50481838 ↗
  • InjuryAug 24, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 23. Details of surgery: Implant surface not completely covered with bone. On (b)(6) 2026, non-osseointegration was verified. Patient presented with fair oral hygiene. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50416911 ↗
  • InjuryAug 24, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50416591 ↗
  • InjuryAug 24, 2026Failure to Osseointegrate

    Patient problems: Hypersensitivity/Allergic reaction

    MAUDE report 50415596 ↗
  • InjuryAug 20, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6)2026 in FDI 22. On (b)(6)2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50353216 ↗
  • InjuryAug 17, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50284298 ↗
  • InjuryAug 17, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 14. Details of surgery: Primary stability not achieved and Implant surface not completely covered with bone. On (b)(6) 2026, non-osseointegration was verified. Patient presented with fair oral hygiene. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50282597 ↗
  • InjuryAug 14, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2026(b)(6) in FDI 23. Details of surgery: Implant surface not completely covered with bone. On 2026(b)(6), non-osseointegration was verified. Patient presented with fair oral hygiene. The device was forwarded to the manufacturer. At the event the patient experienced: Numbness. No further patient complications were reported.

    Patient problems: Numbness

    MAUDE report 50259554 ↗
  • InjuryAug 14, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2026(b)(6) in FDI 43. Details of surgery: Implant surface not completely covered with bone. On 2026(b)(6), non-osseointegration was verified. Patient presented with fair oral hygiene. The device was forwarded to the manufacturer. At the event the patient experienced: Inflammation. No further patient complications were reported.

    Patient problems: Inflammation

    MAUDE report 50259549 ↗
  • InjuryAug 14, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50259277 ↗
  • InjuryAug 14, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2026(b)(6) in FDI 11. On 2026(b)(6), non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50259273 ↗
  • InjuryAug 14, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 15. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50259226 ↗
  • InjuryAug 13, 2026Failure to Osseointegrate

    Patient problems: Fibrosis

    MAUDE report 50239613 ↗
  • InjuryAug 13, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 21. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50239555 ↗
  • InjuryAug 13, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 11. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50239553 ↗

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices

All IMPLANT, ENDOSSEOUS, ROOT-FORM devices →