STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BL (REF 021.5314)Recalls & reported problems
Institut Straumann AG · Bone Level Tapered Implant, D 4.1mm RC, SLActive® 14mm, Roxolid®, Loxim™ · Class 2 · Implant
All models of STRAUMANN BONE LEVEL TAPERED IMPLANT (39)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Straumann SLActive bone level tapered implant is placed in the jawbone to support a replacement tooth. Its 4,369 adverse event reports mostly describe implants that did not fuse with the jawbone or later lost that bond, with some describing pain, infection and scar tissue. The reports include no deaths and 4,369 injuries. 1 recall names this device: started in July 2026 and still open, it covers implants that may be 12mm long rather than their labeled length.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 4,400
- Deaths reported
- 0
- Injuries reported
- 4,400
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4,400 adverse event reports that list this exact device (DI 07630031707055), received between Sep 9, 2019 and Sep 28, 2026 — classified as 4,400 injuries.
The most frequently reported specific device problem is “Failure to Osseointegrate” (3,033 reports).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K140878); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Jul 16, 2026Class IIOpen, Classified
Bone Level Tapered (BLT) Implant, D 4.1mm RC, SLActive 14mm, Roxolid, Loxim may inadvertently contain implants that are 12mm in length instead of 14mm as the labeling would suggest.
Z-2935-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Failure to Osseointegrate3,033
- Loss of Osseointegration723
- Adverse Event Without Identified Device or Use Problem551
- Difficult to Insert70
- Fracture20
- Separation Failure9
- Difficult to Remove5
- Defective Device2
Patient problems
- No Clinical Signs, Symptoms or Conditions1,821
- Fibrosis1,286
- Pain826
- Unspecified Infection695
- Tissue Damage514
- Inflammation341
- Swelling/ Edema213
- Hemorrhage/Blood Loss/Bleeding179
Recent adverse event reports
The 25 most recent of 4,400 reports, in the reporter's own words as published by the FDA.
InjurySep 28, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. Details of surgery: Primary stability not achieved. On (b)(6) 2026, non-osseointegration was verified. No product was returned to the manufacturer. At the event the patient experienced: Infection. No further patient complications were reported.
Patient problems: Unspecified Infection
MAUDE report 50921212 ↗InjurySep 28, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6) 2025 in FDI 32. On (b)(6) 2026, non-osseointegration was verified. Patient presented with fair oral hygiene and inadequate bone quality/quantity. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.
Patient problems: Fibrosis
MAUDE report 50921190 ↗InjurySep 28, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6)2026 in ADA 11. On (b)(6)2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50920182 ↗InjurySep 28, 2026Loss of Osseointegration
The clinician reports the implant was inserted 2026(b)(6) in ADA 10. Details of surgery: Ridge augmentation and Immediate extraction site. On 2026(b)(6), loss of osseointegration was verified. Patient presented with bone type III and fair oral hygiene. The device was forwarded to the manufacturer. At the event the patient experienced: Infection, Mobility, Fistula and Inflammation. No further patient complications were reported.
Patient problems: Fibrosis; Fistula; Inflammation; Unspecified Infection
MAUDE report 50920117 ↗InjurySep 28, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility and Inflammation. No further patient complications were reported.
Patient problems: Fibrosis; Inflammation
MAUDE report 50919403 ↗InjurySep 28, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6) 2026 in ADA 11. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50918610 ↗InjurySep 28, 2026Loss of Osseointegration
The clinician reports the implant was inserted 2026(b)(6) in ADA 10. Details of surgery: Ridge augmentation and Immediate extraction site. On 2026(b)(6), loss of osseointegration was verified. Patient presented with bone type III and fair oral hygiene. The device was forwarded to the manufacturer. At the event the patient experienced: Infection, Mobility, Fistula and Inflammation. No further patient complications were reported.
Patient problems: Fibrosis; Fistula; Inflammation; Unspecified Infection
MAUDE report 50918546 ↗InjurySep 25, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. Details of surgery: Implant surface not completely covered with bone. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Bleeding. No further patient complications were reported.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 50901742 ↗InjurySep 25, 2026Loss of Osseointegration
The clinician reports the implant was inserted (b)(6) 2018 in ADA 14. Details of surgery: Primary stability not achieved. On (b)(6) 2026, loss of osseointegration was verified. Patient presented with bone type IV. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.
Patient problems: Fibrosis
MAUDE report 50901334 ↗InjurySep 25, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6) 2026 in FDI 32. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50901184 ↗InjurySep 24, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6) 2026 in ADA 11. Details of surgery: Implant surface not completely covered with bone. On(b)(6) 2026, non-osseointegration was verified. No product was returned to the manufacturer. At the event the patient experienced: Pain and Mobility. No further patient complications were reported.
Patient problems: Fibrosis; Pain
MAUDE report 50887682 ↗InjurySep 24, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6) 2026 in ADA 21. Details of surgery: Implant surface not completely covered with bone. On (b)(6) 2026, non-osseointegration was verified. No product was returned to the manufacturer. At the event the patient experienced: Pain and Mobility. No further patient complications were reported.
Patient problems: Fibrosis; Pain
MAUDE report 50887678 ↗InjurySep 21, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6) 2026 in ADA 5. Details of surgery: Immediate extraction site. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Pain. No further patient complications were reported.
Patient problems: Pain
MAUDE report 50845036 ↗InjurySep 21, 2026Failure to Osseointegrate
The clinician reports the implant was inserted 2026(b)(6) in ADA 5. Details of surgery: Immediate extraction site. On 2026(b)(6), non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Pain. No further patient complications were reported.
Patient problems: Pain
MAUDE report 50843450 ↗InjurySep 18, 2026Loss of Osseointegration
The clinician reports the implant was inserted (b)(6) 2025 in FDI 21. On (b)(6) 2026, loss of osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility, Swelling and Increased Sensitivity. No further patient complications were reported.
Patient problems: Fibrosis; Hyperresponsive to stimuli; Swelling/ Edema
MAUDE report 50826449 ↗InjurySep 17, 2026Adverse Event Without Identified Device or Use Problem
The clinician reports that the day the implant was placed in the patient's mouth, failure occurred upon insertion. Details of surgery: Primary stability not achieved. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50806989 ↗InjurySep 14, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6) 2026 in FDI 27. On (b)(6) 2026, non-osseointegration was verified. Patient presented with bone type III and fair oral hygiene. The device was forwarded to the manufacturer. At the event the patient experienced: Inflammation. No further patient complications were reported.
Patient problems: Inflammation
MAUDE report 50760158 ↗InjurySep 14, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6) 2026 in FDI 12. On (b)(6) 2026, non-osseointegration was verified. No product was returned to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.
Patient problems: Fibrosis
MAUDE report 50759136 ↗InjurySep 11, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6) 2026 in ADA 13. Details of surgery: Ridge augmentation. On (b)(6) 2026, non-osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. At the event the patient experienced: Infection and Peri-implantitis. No further patient complications were reported.
Patient problems: Unspecified Infection
MAUDE report 50738102 ↗InjurySep 11, 2026Failure to Osseointegrate
The clinician reports the implant was inserted 2026 (b)(6) in ADA 13. Details of surgery: Ridge augmentation. On 2026 (b)(6), non-osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. At the event the patient experienced: Infection and Peri-implantitis. No further patient complications were reported.
Patient problems: Unspecified Infection
MAUDE report 50737412 ↗InjurySep 10, 2026Adverse Event Without Identified Device or Use Problem
The clinician reports that the day the implant was placed in ADA 10, failure occurred upon insertion. Details of surgery: Primary stability not achieved. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50720347 ↗InjurySep 10, 2026Loss of Osseointegration
The clinician reports the implant was inserted (B)(6) 2026 in FDI 15. On (B)(6) 2026, loss of osseointegration was verified. Patient presented with bone type III and poor oral hygiene. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50719283 ↗InjurySep 10, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6) 2026 in ADA 28. On (b)(6) 2026, non-osseointegration was verified. Patient presented with bone type III. No product was returned to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.
Patient problems: Fibrosis
MAUDE report 50718933 ↗InjurySep 10, 2026Loss of Osseointegration
The clinician reports the implant was inserted (b)(6) 2020 in FDI 21. On (b)(6) 2026, loss of osseointegration was verified. Patient presented with fair oral hygiene. The device was forwarded to the manufacturer. At the event the patient experienced: Peri-implantitis. No further patient complications were reported.
Patient problems: Unspecified Infection
MAUDE report 50718924 ↗InjurySep 7, 2026Failure to Osseointegrate
The clinician reports the implant was inserted 2026(b)(6) in FDI 25. On 2026(b)(6), non-osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50655296 ↗
FDA clearance and classification
Cleared through 510(k) K140878 on Aug 20, 2014 (decision: Substantially Equivalent). Submitted by Straumann USA, LLC as “Straumann Bone Level Tapered Implant”. The FDA review took 135 days. FDA record ↗
The FDA regulates product code DZE (implant, endosseous, root-form) under 21 CFR 872.3640 as a Class II device. 21 CFR 872.3640 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K140878. One submission often covers many models, so check the product and lot details in each record.
- Z-0117-2015 · Sep 19, 2014 · TerminatedStraumann Bone Level Tapered Implant 04.1 mm RC, SLActive 8mm, Roxolid, Loxim Article Number: 021 .5308 Dental implants are intended for surgical placement in maxillary and/or
Common questions about the STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BL
Has the STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BL been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2935-2026 (Class II), began on Jul 16, 2026: Bone Level Tapered (BLT) Implant, D 4.1mm RC, SLActive 14mm, Roxolid, Loxim may inadvertently contain implants that are 12mm in length instead of 14mm as the labeling would suggest. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BL?
The FDA's MAUDE database lists 4,400 adverse event reports for this device, received between Sep 9, 2019 and Sep 28, 2026. In those reports, the most common specific device problems are “Failure to Osseointegrate” (3,033), “Loss of Osseointegration” (723) and “Difficult to Insert” (70); the most common patient problems are “Fibrosis” (1,286), “Pain” (826) and “Unspecified Infection” (695). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BL?
None of the reports are classified as deaths, 4,400 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BL FDA approved?
It was cleared through 510(k) K140878 on Aug 20, 2014. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BL MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BL, and is it still sold?
The FDA's device database lists Institut Straumann AG as the company responsible for it (brand STRAUMANN BONE LEVEL TAPERED IMPLANT, model SLActive Roxolid BL). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM43,942 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM40,651 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM27,576 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,756 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLA Titanium BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,548 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM23,631 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,501 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,423 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page