STRAUMANN BONE LEVEL TAPERED IMPLANT SLA Titanium BL (REF 021.5416)Recalls & reported problems
Institut Straumann AG · Bone Level Tapered Implant, D 4.1mm RC, SLA® 16mm, Ti, Loxim™ · Class 2 · Implant
All models of STRAUMANN BONE LEVEL TAPERED IMPLANT (39)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Straumann Bone Level tapered dental implant is a post placed in the jawbone to hold a tooth replacement. Its 333 reports mostly describe the implant failing to bond with the bone, with pain, infection and scar tissue mentioned in some. The reports mention 333 injuries and no deaths. Reports fell from 64 in the prior 12 months to 37 in the latest 12 months. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 339
- Deaths reported
- 0
- Injuries reported
- 339
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 339 adverse event reports that list this exact device (DI 07630031707116), received between Sep 10, 2019 and Sep 28, 2026 — classified as 339 injuries.
The most frequently reported specific device problem is “Failure to Osseointegrate” (279 reports).
2 other recalls cover devices cleared under the same FDA submission (K140878); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Osseointegrate279
- Loss of Osseointegration55
- Adverse Event Without Identified Device or Use Problem5
- Fracture1
Patient problems
- No Clinical Signs, Symptoms or Conditions211
- Fibrosis59
- Pain54
- Unspecified Infection32
- Inflammation19
- Tissue Damage15
- Swelling/ Edema13
- Abscess9
Recent adverse event reports
The 25 most recent of 339 reports, in the reporter's own words as published by the FDA.
InjurySep 28, 2026Failure to Osseointegrate
Patient problems: Inflammation
MAUDE report 50921058 ↗InjurySep 28, 2026Failure to Osseointegrate
Patient problems: Inflammation
MAUDE report 50921054 ↗InjurySep 10, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6) 2026 in FDI 21. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50719941 ↗InjurySep 7, 2026Failure to Osseointegrate
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50654722 ↗InjuryAug 27, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6) 2026 in FDI 46. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50481685 ↗InjuryAug 24, 2026Failure to Osseointegrate
Patient problems: Pain
MAUDE report 50415564 ↗InjuryAug 17, 2026Failure to Osseointegrate
The clinician reports the implant was inserted 2026 (b)(6) in FDI 21. On 2026 (b)(6), non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50282028 ↗InjuryAug 10, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6) 2026 in FDI 14. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50199087 ↗InjuryAug 6, 2026Failure to Osseointegrate
The clinician reports the implant was inserted 2026(b)(6) in FDI 15. On 2026(b)(6), non-osseointegration was verified. Patient presented with fair oral hygiene. The device was forwarded to the manufacturer. At the event the patient experienced: Inflammation. No further patient complications were reported.
Patient problems: Inflammation
MAUDE report 50162888 ↗InjuryAug 3, 2026Failure to Osseointegrate
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50120992 ↗InjuryJul 30, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6)2026 in FDI 15. Details of surgery: Primary stability not achieved and Implant surface not completely covered with bone. On (b)(6)2026, non-osseointegration was verified. Patient presented with fair oral hygiene. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50088375 ↗InjuryJul 30, 2026Failure to Osseointegrate
The clinician reports the implant was inserted 2026(b)(6) in FDI 16. On 2026(b)(6), non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50088247 ↗InjuryJul 27, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6)2025 in FDI 47. Details of surgery: Primary stability not achieved and Implant surface not completely covered with bone. On 2026-03-27, non-osseointegration was verified. Patient presented with fair oral hygiene. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50044012 ↗InjuryJul 27, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6) 2025 in FDI 16. Details of surgery: Primary stability not achieved and Implant surface not completely covered with bone. On (b)(6) 2026, non-osseointegration was verified. Patient presented with fair oral hygiene. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50040574 ↗InjuryJul 27, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (B)(6) 2025 in FDI 46. On (B)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50023403 ↗InjuryJul 13, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN FDI 26. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25800894 ↗InjuryJul 9, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN FDI 25. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25767540 ↗InjuryJul 9, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN FDI 15. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25767534 ↗InjuryJun 26, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 43. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25639557 ↗InjuryJun 25, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN FDI 12. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS, PAIN AND INFLAMMATION. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Inflammation; Pain; Unspecified Infection
MAUDE report 25624630 ↗InjuryJun 22, 2026Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2022 IN ADA 22. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Unspecified Infection
MAUDE report 25584151 ↗InjuryJun 22, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2025 IN FDI 15. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED AND IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25582749 ↗InjuryJun 18, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 21. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25550781 ↗InjuryJun 3, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 25. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED AND IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25378746 ↗InjuryMay 15, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 16. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25189759 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K140878. One submission often covers many models, so check the product and lot details in each record.
- Z-2935-2026 · Jul 16, 2026 · Open, ClassifiedStraumann Bone Level Tapered (BLT) Implant, D 4.1mm RC, SLActive 14mm, Roxolid, Loxim (Article 021.5314); implant, endosseous, root-form
- Z-0117-2015 · Sep 19, 2014 · TerminatedStraumann Bone Level Tapered Implant 04.1 mm RC, SLActive 8mm, Roxolid, Loxim Article Number: 021 .5308 Dental implants are intended for surgical placement in maxillary and/or man
Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM43,616 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM40,404 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM27,400 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,621 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLA Titanium BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,298 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM23,448 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,320 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,267 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗