Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BL (REF 021.6508)Recalls & reported problems

Institut Straumann AG · Bone Level Implant, D 4.8mm RC, SLA® 8mm, Roxolid®, Loxim™ · Class 2 · Implant

All models of Straumann Dental Implant System Roxolid SLA Impl. (80)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Straumann Roxolid SLA dental implant is a bone-level implant placed in the jawbone. Reports most often describe the implant not fusing with the bone or losing that bond, with pain, scar-like tissue, infection and tissue damage mentioned. The device appears in 5,253 reports, with 5,253 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
5,280
Deaths reported
0
Injuries reported
5,280
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5,280 adverse event reports that list this exact device (DI 07630031707352), received between Sep 10, 2019 and Sep 28, 2026 — classified as 5,280 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (3,222 reports).

2 other recalls cover devices cleared under the same FDA submission (K150938); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 60 reportsJan 2024: 72 reportsFeb 2024: 54 reportsMar 2024: 57 reportsApr 2024: 75 reportsMay 2024: 79 reportsJun 2024: 57 reportsJul 2024: 70 reportsAug 2024: 61 reportsSep 2024: 74 reportsOct 2024: 110 reportsNov 2024: 75 reportsDec 2024: 55 reportsJan 2025: 59 reportsFeb 2025: 42 reportsMar 2025: 89 reportsApr 2025: 18 reportsMay 2025: 0 reportsJun 2025: 180 reportsJul 2025: 77 reportsAug 2025: 77 reportsSep 2025: 35 reportsOct 2025: 46 reportsNov 2025: 56 reportsDec 2025: 71 reportsJan 2026: 63 reportsFeb 2026: 49 reportsMar 2026: 62 reportsApr 2026: 71 reportsMay 2026: 66 reportsJun 2026: 52 reportsJul 2026: 59 reportsAug 2026: 39 reportsSep 2026: 35 reportsOct 2026: 0 reportsNov 2026: 0 reports202420252026180
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate3,222
  • Loss of Osseointegration1,598
  • Adverse Event Without Identified Device or Use Problem385
  • Difficult to Insert64
  • Fracture16
  • Separation Failure9
  • Unintended Movement7
  • Difficult to Remove6

Patient problems

  • Fibrosis1,914
  • No Clinical Signs, Symptoms or Conditions1,780
  • Pain1,144
  • Unspecified Infection823
  • Tissue Damage658
  • Inflammation399
  • Hemorrhage/Blood Loss/Bleeding310
  • Swelling/ Edema284

Recent adverse event reports

The 25 most recent of 5,280 reports, in the reporter's own words as published by the FDA.

  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 34. Details of surgery: Primary stability not achieved and Implant surface not completely covered with bone. On 2026-08-21, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50921927 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6)2026 in ADA 3. On (b)(6)2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50921140 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 13. Details of surgery: Implant surface not completely covered with bone. On (b)(6) 2026, non-osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. At the event the patient experienced: Infection, Mobility, Bleeding and Inflammation. No further patient complications were reported.

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Inflammation; Unspecified Infection

    MAUDE report 50920099 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 3. Details of surgery: Implant surface not completely covered with bone and Sinus augmentation. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50919934 ↗
  • InjurySep 28, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2025 in FDI 26. On (b)(6) 2026, loss of osseointegration was verified. Patient presented with fair oral hygiene. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility and Inflammation. No further patient complications were reported.

    Patient problems: Fibrosis; Inflammation

    MAUDE report 50919863 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 31. On 2026-08-19, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50919654 ↗
  • InjurySep 28, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6)2026 in the patient's mouth. On (b)(6)2026, loss of osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50919402 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 3. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50918728 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2026(b)(6) in ADA 13. Details of surgery: Implant surface not completely covered with bone. On 2026(b)(6), non-osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. At the event the patient experienced: Infection, Mobility, Bleeding and Inflammation. No further patient complications were reported.

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Inflammation; Unspecified Infection

    MAUDE report 50918534 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2026(b)(6) in ADA 3. Details of surgery: Implant surface not completely covered with bone and Sinus augmentation. On 2026(b)(6), non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50918469 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2026(b)(6) in ADA 31. On 2026(b)(6), non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50918323 ↗
  • InjurySep 24, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2022 in FDI 37. On (b)(6) 2026, loss of osseointegration was verified. No product was returned to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50888070 ↗
  • InjurySep 24, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 17. On (b)(6) 2026, non-osseointegration was verified. Patient presented with bone type III and fair oral hygiene. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50887998 ↗
  • InjurySep 21, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 14. Details of surgery: Primary stability not achieved and Sinus augmentation. On (b)(6) 2026, non-osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50844969 ↗
  • InjurySep 21, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2025 in the patient's mouth. On (b)(6) 2025, non-osseointegration was verified. No product was returned to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50844265 ↗
  • InjurySep 17, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2026(b)(6) in the patient's mouth. On 2026(b)(6), non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50807429 ↗
  • InjurySep 17, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (B)(6) 2025 in ADA 14. On (B)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50807105 ↗
  • InjurySep 17, 2026Loss of Osseointegration

    The clinician reports the implant was inserted on (b)(6) 2025 in ADA 14. On (b)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50806349 ↗
  • InjurySep 14, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 26. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50759238 ↗
  • InjurySep 14, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6)2026 in FDI 16. On (b)(6)2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50759237 ↗
  • InjurySep 14, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 22. On (b)(6) 2026, loss of osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. At the event the patient experienced: Peri-implantitis. No further patient complications were reported.

    Patient problems: Unspecified Infection

    MAUDE report 50758590 ↗
  • InjurySep 11, 2026Loss of Osseointegration

    The clinician reports the implant was inserted 2025(b)(6) in ADA 2. On 2026(b)(6), loss of osseointegration was verified. Patient presented with bone type III and inadequate bone quality/quantity. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50738592 ↗
  • InjurySep 11, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2026(b)(6) in ADA 18. Details of surgery: Primary stability not achieved and Implant surface not completely covered with bone. On 2026(b)(6), non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50738408 ↗
  • InjurySep 11, 2026Loss of Osseointegration

    The clinician reports the implant was inserted 2026(b)(6) in FDI 46. On 2026(b)(6), loss of osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50738197 ↗
  • InjurySep 11, 2026Loss of Osseointegration

    The clinician reports the implant was inserted 2025(b)(6) in FDI 16. On 2026(b)(6), loss of osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Peri-implantitis and Mobility. No further patient complications were reported.

    Patient problems: Fibrosis; Unspecified Infection

    MAUDE report 50738157 ↗

FDA clearance and classification

Cleared through 510(k) K150938 on Jul 24, 2015 (decision: Substantially Equivalent). Submitted by Straumann USA, LLC as “Straumann® Dental Implant System ¿ Roxolid® SLA Implants”. The FDA review took 108 days. FDA record ↗

The FDA regulates product code DZE (implant, endosseous, root-form) under 21 CFR 872.3640 as a Class II device. 21 CFR 872.3640 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K150938. One submission often covers many models, so check the product and lot details in each record.

  • Z-2535-2016 · Jul 14, 2016 · Terminated
    Straumann Bone Level Implant, 4.1mm RC, SLA 12mm, TiZr, NTP Article Number: 021.4512
  • Z-2534-2016 · Jul 14, 2016 · Terminated
    Straumann Bone Level Implant, 4.1mm RC, SLA 10mm, TiZr, NTP Article Number: 021.4510

Common questions about the Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BL

Has the Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BL been recalled?

No FDA recall record in our data names this exact device identifier (DI 07630031707352), based on the FDA recall file dated Nov 27, 2026. However, 2 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BL?

The FDA's MAUDE database lists 5,280 adverse event reports for this device, received between Sep 10, 2019 and Sep 28, 2026. In those reports, the most common specific device problems are “Failure to Osseointegrate” (3,222), “Loss of Osseointegration” (1,598) and “Difficult to Insert” (64); the most common patient problems are “Fibrosis” (1,914), “Pain” (1,144) and “Unspecified Infection” (823). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BL?

None of the reports are classified as deaths, 5,280 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BL FDA approved?

It was cleared through 510(k) K150938 on Jul 24, 2015. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BL MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BL, and is it still sold?

The FDA's device database lists Institut Straumann AG as the company responsible for it (brand Straumann® Dental Implant System Roxolid® SLA Impl., model SLA Roxolid BL). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices

All IMPLANT, ENDOSSEOUS, ROOT-FORM devices →

Sources and updates