Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BL (REF 021.6512)Recalls & reported problems

Institut Straumann AG · Bone Level Implant, D 4.8mm RC, SLA® 12mm, Roxolid®, Loxim™ · Class 2 · Implant

All models of Straumann Dental Implant System Roxolid SLA Impl. (80)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Straumann bone level implant is a dental implant placed into the jawbone. Its 2,104 adverse event reports, all classed as injuries, mostly describe the implant failing to fuse with the bone or losing its bond with it, and some mention pain, infection or tissue damage. The reports include no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
2,107
Deaths reported
0
Injuries reported
2,107
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2,107 adverse event reports that list this exact device (DI 07630031707376), received between Sep 9, 2019 and Sep 14, 2026 — classified as 2,107 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (1,308 reports).

2 other recalls cover devices cleared under the same FDA submission (K150938); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 13 reportsDec 2023: 11 reportsJan 2024: 30 reportsFeb 2024: 16 reportsMar 2024: 15 reportsApr 2024: 20 reportsMay 2024: 14 reportsJun 2024: 17 reportsJul 2024: 39 reportsAug 2024: 23 reportsSep 2024: 18 reportsOct 2024: 107 reportsNov 2024: 52 reportsDec 2024: 19 reportsJan 2025: 30 reportsFeb 2025: 19 reportsMar 2025: 24 reportsApr 2025: 14 reportsMay 2025: 0 reportsJun 2025: 60 reportsJul 2025: 16 reportsAug 2025: 25 reportsSep 2025: 5 reportsOct 2025: 11 reportsNov 2025: 18 reportsDec 2025: 17 reportsJan 2026: 11 reportsFeb 2026: 18 reportsMar 2026: 15 reportsApr 2026: 20 reportsMay 2026: 29 reportsJun 2026: 23 reportsJul 2026: 15 reportsAug 2026: 35 reportsSep 2026: 8 reportsOct 2026: 0 reports202420252026107
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate1,308
  • Loss of Osseointegration518
  • Adverse Event Without Identified Device or Use Problem212
  • Difficult to Insert59
  • Separation Failure9
  • Fracture8
  • Unintended Movement3
  • Therapeutic or Diagnostic Output Failure2

Patient problems

  • Fibrosis709
  • No Clinical Signs, Symptoms or Conditions667
  • Pain507
  • Unspecified Infection344
  • Tissue Damage288
  • Inflammation150
  • Hemorrhage/Blood Loss/Bleeding145
  • Swelling/ Edema99

Recent adverse event reports

The 25 most recent of 2,107 reports, in the reporter's own words as published by the FDA.

  • InjurySep 14, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50760588 ↗
  • InjurySep 14, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50760585 ↗
  • InjurySep 14, 2026Loss of Osseointegration

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50760565 ↗
  • InjurySep 14, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50759604 ↗
  • InjurySep 14, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50759452 ↗
  • InjurySep 14, 2026Loss of Osseointegration

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Pain; Unspecified Infection

    MAUDE report 50759149 ↗
  • InjurySep 14, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50759046 ↗
  • InjurySep 7, 2026Loss of Osseointegration

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50654661 ↗
  • InjuryAug 27, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50482056 ↗
  • InjuryAug 27, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2025 in FDI 27. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50481922 ↗
  • InjuryAug 27, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (B)(6) 2022 in FDI 17. On (B)(6) 2026, loss of osseointegration was verified. Patient presented with bone type III and inadequate bone quality/quantity. The device was forwarded to the manufacturer. At the event the patient experienced: Pain. No further patient complications were reported.

    Patient problems: Pain

    MAUDE report 50480814 ↗
  • InjuryAug 27, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2022 in FDI 16. On (b)(6) 2026, loss of osseointegration was verified. Patient presented with bone type III and inadequate bone quality/quantity. The device was forwarded to the manufacturer. At the event the patient experienced: Pain. No further patient complications were reported.

    Patient problems: Pain

    MAUDE report 50480813 ↗
  • InjuryAug 27, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2022 in FDI 46. On (b)(6) 2026, loss of osseointegration was verified. Patient presented with bone type III and inadequate bone quality/quantity. The device was forwarded to the manufacturer. At the event the patient experienced: Pain. No further patient complications were reported.

    Patient problems: Pain

    MAUDE report 50480807 ↗
  • InjuryAug 27, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 46. On (b)(6) 2026, loss of osseointegration was verified. Patient presented with fair oral hygiene and inadequate bone quality/quantity. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50480765 ↗
  • InjuryAug 27, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (B)(6) 2026 in FDI 47. On (B)(6) 2026, loss of osseointegration was verified. Patient presented with fair oral hygiene and inadequate bone quality/quantity. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50480763 ↗
  • InjuryAug 27, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (B)(6) 2021 in FDI 26. On (B)(6) 2026, loss of osseointegration was verified. Patient presented with inadequate bone quality/quantity. The device was forwarded to the manufacturer. At the event the patient experienced: Pain. No further patient complications were reported.

    Patient problems: Pain

    MAUDE report 50480762 ↗
  • InjuryAug 27, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (B)(6) 2021 in FDI 27. On (B)(6) 2026, loss of osseointegration was verified. Patient presented with inadequate bone quality/quantity. The device was forwarded to the manufacturer. At the event the patient experienced: Pain. No further patient complications were reported.

    Patient problems: Pain

    MAUDE report 50480760 ↗
  • InjuryAug 27, 2026Loss of Osseointegration

    The clinician reports the implant was inserted(B)(6) 2024 in the patient's mouth. On (B)(6) 2026, loss of osseointegration was verified. Patient presented with fair oral hygiene. The device was forwarded to the manufacturer. At the event the patient experienced: Infection, Mobility, Abscess and Inflammation. No further patient complications were reported.

    Patient problems: Abscess; Fibrosis; Inflammation; Unspecified Infection

    MAUDE report 50480596 ↗
  • InjuryAug 27, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (B)(6) 2024 in FDI 46. On (B)(6) 2026, loss of osseointegration was verified. Patient presented with fair oral hygiene. The device was forwarded to the manufacturer. At the event the patient experienced: Pain and Mobility. No further patient complications were reported.

    Patient problems: Fibrosis; Pain

    MAUDE report 50480594 ↗
  • InjuryAug 27, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (B)(6) 2025 in FDI 15. On (B)(6) 2026, non-osseointegration was verified. Patient presented with fair oral hygiene and inadequate bone quality/quantity. The device was forwarded to the manufacturer. At the event the patient experienced: Pain and Mobility. No further patient complications were reported.

    Patient problems: Fibrosis; Pain

    MAUDE report 50480589 ↗
  • InjuryAug 27, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (B)(6) 2024 in FDI 26. On (B)(6) 2026, loss of osseointegration was verified. Patient presented with fair oral hygiene and inadequate bone quality/quantity. The device was forwarded to the manufacturer. At the event the patient experienced: Pain and Mobility. No further patient complications were reported.

    Patient problems: Fibrosis; Pain

    MAUDE report 50480579 ↗
  • InjuryAug 27, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (B)(6) 2025 in FDI 46. On (B)(6) 2026, non-osseointegration was verified. Patient presented with fair oral hygiene and inadequate bone quality/quantity. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50480568 ↗
  • InjuryAug 24, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2024 in FDI 25. On 2026-07-16, loss of osseointegration was verified. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50415246 ↗
  • InjuryAug 24, 2026Failure to Osseointegrate

    Patient problems: Fibrosis

    MAUDE report 50414670 ↗
  • InjuryAug 24, 2026Loss of Osseointegration

    The clinician reports the implant was inserted 2025(b)(6) in FDI 11. On 2026(b)(6) loss of osseointegration was verified. Patient presented with inadequate bone quality/quantity. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50414263 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K150938. One submission often covers many models, so check the product and lot details in each record.

  • Z-2535-2016 · Jul 14, 2016 · Terminated
    Straumann Bone Level Implant, 4.1mm RC, SLA 12mm, TiZr, NTP Article Number: 021.4512
  • Z-2534-2016 · Jul 14, 2016 · Terminated
    Straumann Bone Level Implant, 4.1mm RC, SLA 10mm, TiZr, NTP Article Number: 021.4510

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices

All IMPLANT, ENDOSSEOUS, ROOT-FORM devices →

Sources and updates