PURE Ceramic Implants Ceramic TL (REF 031.002S)Recalls & reported problems

Institut Straumann AG · Straumann® PURE Ceramic Implant Monotype, Ø 3.3mm ND, ZLA 10mm, AH 4mm, ZrO2 · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Straumann PURE Ceramic dental implant is a ceramic screw-shaped implant placed in the jawbone to support a replacement tooth. Most of the 112 reports describe the implant failing to bond with the jawbone, and some describe pain, bleeding, or infection. The reports include 112 injuries and no deaths. Reports fell to 2 in the last 12 months from 11 in the previous 12 months. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
113
Deaths reported
0
Injuries reported
113
Recalls naming it
0

Summary

The FDA's MAUDE database contains 113 adverse event reports that list this exact device (DI 07630031713797), received between Aug 5, 2019 and Sep 17, 2026 — classified as 113 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (68 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 3 reportsJan 2024: 3 reportsFeb 2024: 2 reportsMar 2024: 2 reportsApr 2024: 2 reportsMay 2024: 1 reportsJun 2024: 2 reportsJul 2024: 1 reportsAug 2024: 2 reportsSep 2024: 4 reportsOct 2024: 1 reportsNov 2024: 3 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 3 reportsMar 2025: 3 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate68
  • Loss of Osseointegration34
  • Adverse Event Without Identified Device or Use Problem9
  • Difficult to Insert3
  • Separation Failure2
  • Fracture1

Patient problems

  • Fibrosis40
  • No Clinical Signs, Symptoms or Conditions35
  • Pain34
  • Hemorrhage/Blood Loss/Bleeding22
  • Tissue Damage15
  • Unspecified Infection14
  • Inflammation10
  • Swelling/ Edema6

Recent adverse event reports

The 25 most recent of 113 reports, in the reporter's own words as published by the FDA.

  • InjurySep 17, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (B)(6) 2026 in the patient's mouth. On (B)(6) 2026, non-osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. At the event the patient experienced: Pain. No further patient complications were reported.

    Patient problems: Pain

    MAUDE report 50807145 ↗
  • InjuryMar 16, 2026Failure to Osseointegrate; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN ADA 24. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 24610680 ↗
  • InjuryJan 15, 2026Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2025 IN FDI 31. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE AND IMMEDIATE EXTRACTION SITE. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: FISTULA. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fistula

    MAUDE report 24083749 ↗
  • InjurySep 2, 2025Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFECTION, PAIN AND MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Pain; Unspecified Infection

    MAUDE report 22933765 ↗
  • InjuryJun 18, 2025Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2025 IN THE PATIENT'S MOUTH. DETAILS OF SURGERY: RIDGE AUGMENTATION. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22262537 ↗
  • InjuryMar 28, 2025Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 15. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE IV. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN AND MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Pain

    MAUDE report 21723078 ↗
  • InjuryMar 3, 2025Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN ADA 7. ON (B)(6) 2024 , NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21507798 ↗
  • InjuryMar 3, 2025Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN ADA 7. ON (B)(6) 2024, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21506763 ↗
  • InjuryFeb 20, 2025Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN FDI 35. ON (B)(6) 2024, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE IV AND INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN AND MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Pain

    MAUDE report 21424493 ↗
  • InjuryFeb 10, 2025Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2024 IN FDI 45. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21347801 ↗
  • InjuryFeb 3, 2025Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN THE PATIENT'S MOUTH. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED. ON (B)(6) 2024, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21296037 ↗
  • InjuryNov 21, 2024Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2022 IN ADA 15. ON (B)(6) 2024, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 20749216 ↗
  • InjuryNov 15, 2024Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN ADA 4. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE AND SINUS PERFORATION. ON (B)(6) 2024, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE IV. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: FISTULA. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fistula; Sinus Perforation

    MAUDE report 20697334 ↗
  • InjuryNov 15, 2024Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN ADA 4. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE AND SINUS PERFORATION. ON (B)(6) 2024 NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE IV. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: FISTULA. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fistula; Sinus Perforation

    MAUDE report 20696638 ↗
  • InjuryOct 7, 2024Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN FDI 31. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED, IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE AND IMMEDIATE EXTRACTION SITE. ON (B)(6) 2024, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20388172 ↗
  • InjurySep 27, 2024Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN ADA 22. ON (B)(6) 2024 NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20324159 ↗
  • InjurySep 27, 2024Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN ADA 22. ON (B)(6) 2024, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20323963 ↗
  • InjurySep 6, 2024Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2023 IN ADA 10. ON (B)(6) 2024, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III AND PREVIOUS BONE AUGMENTATION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 20160090 ↗
  • InjurySep 6, 2024Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2023 IN ADA 10. ON (B)(6) 2024, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III AND PREVIOUS BONE AUGMENTATION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 20159406 ↗
  • InjuryAug 20, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE CLINICIAN REPORTS THAT THE DAY THE IMPLANT WAS PLACED IN ADA 8, FAILURE OCCURRED UPON INSERTION. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED, IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE AND RIDGE AUGMENTATION. PATIENT PRESENTED WITH BONE TYPE IV, INADEQUATE BONE QUALITY/QUANTITY AND PREVIOUS BONE AUGMENTATION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20005171 ↗
  • InjuryAug 20, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE CLINICIAN REPORTS THAT THE DAY THE IMPLANT WAS PLACED IN ADA 8, FAILURE OCCURRED UPON INSERTION. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED, IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE AND RIDGE AUGMENTATION. PATIENT PRESENTED WITH BONE TYPE IV, INADEQUATE BONE QUALITY/QUANTITY AND PREVIOUS BONE AUGMENTATION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20001173 ↗
  • InjuryJul 16, 2024Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2023 IN ADA 2. ON (B)(6) 2024, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFECTION, PAIN, MOBILITY AND ABSCESS. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Abscess; Fibrosis; Pain; Unspecified Infection

    MAUDE report 19754830 ↗
  • InjuryJun 21, 2024Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2017 IN FDI 21. ON (B)(6) 2024, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 19586517 ↗
  • InjuryJun 17, 2024Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN FDI 34. ON (B)(6) 2024, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain

    MAUDE report 19551491 ↗
  • InjuryMay 6, 2024Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2023 IN THE PATIENT'S MOUTH. ON (B)(6) 2024, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19251262 ↗

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices