PURE Ceramic Implants Ceramic TL (REF 031.003S)Recalls & reported problems

Institut Straumann AG · Straumann® PURE Ceramic Implant Monotype, Ø 3.3mm ND, ZLA 12mm, AH 4mm, ZrO2 · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Straumann PURE Ceramic Implant is a small tooth-root implant used to replace a missing tooth. Most of the 64 reports describe implants that fail to bond with the surrounding bone, or lose that bond after placement, leaving the implant loose; some also mention bone growth around the implant, pain, bleeding, and a few fractured implants. All 64 involve injuries, and no deaths were reported. No recalls name this device, and reports have dropped from 9 to 4 over the last 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
64
Deaths reported
0
Injuries reported
64
Recalls naming it
0

Summary

The FDA's MAUDE database contains 64 adverse event reports that list this exact device (DI 07630031713803), received between Sep 13, 2019 and Jul 6, 2026 — classified as 64 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (37 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 4 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate37
  • Loss of Osseointegration14
  • Adverse Event Without Identified Device or Use Problem10
  • Fracture3
  • Difficult to Insert1

Patient problems

  • Fibrosis31
  • No Clinical Signs, Symptoms or Conditions18
  • Pain12
  • Tissue Damage9
  • Unspecified Infection9
  • Hemorrhage/Blood Loss/Bleeding8
  • Hyperresponsive to stimuli7
  • Inflammation4

Recent adverse event reports

The 25 most recent of 64 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 6, 2026Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN FDI 25. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 25723014 ↗
  • InjuryApr 17, 2026Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2021 IN FDI 12. DETAILS OF SURGERY: RIDGE AUGMENTATION. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS, MOBILITY, BLEEDING AND INCREASED SENSITIVITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Hyperresponsive to stimuli; Unspecified Infection

    MAUDE report 24914752 ↗
  • InjuryMar 13, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 45. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY AND BLEEDING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding

    MAUDE report 24593005 ↗
  • InjuryNov 17, 2025Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2021 IN FDI 41. DETAILS OF SURGERY: RIDGE AUGMENTATION. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS, MOBILITY, BLEEDING, INCREASED SENSITIVITY AND INFLAMMATION. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Hyperresponsive to stimuli; Inflammation; Unspecified Infection

    MAUDE report 23574788 ↗
  • InjuryAug 25, 2025Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN FDI 15. DETAILS OF SURGERY: SINUS AUGMENTATION. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 22888165 ↗
  • InjuryJul 25, 2025Fracture; Fracture

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2020 IN FDI 43. ON (B)(6) 2025, FRACTURE OF THE IMPLANT WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22612025 ↗
  • InjuryJun 18, 2025Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 45. DETAILS OF SURGERY: BONE OVERHEATING. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN AND MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Pain

    MAUDE report 22276506 ↗
  • InjuryJun 18, 2025Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2024 IN FDI 41. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22271552 ↗
  • InjuryJun 18, 2025Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN FDI 26. DETAILS OF SURGERY: SINUS AUGMENTATION. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22261881 ↗
  • InjuryJun 17, 2025Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2024 IN FDI 14. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22218679 ↗
  • InjuryFeb 20, 2025Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN FDI 35. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2024, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN AND MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Pain

    MAUDE report 21424787 ↗
  • InjuryJan 16, 2025Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6)2024 IN FDI 31. ON (B)(6)2024, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN, SWELLING AND INCREASED SENSITIVITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Hyperresponsive to stimuli; Pain; Swelling/ Edema

    MAUDE report 21171501 ↗
  • InjuryDec 5, 2024Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6)2024 IN THE PATIENT'S MOUTH. ON (B)(6)2024, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN AND INCREASED SENSITIVITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Hyperresponsive to stimuli; Pain

    MAUDE report 20858229 ↗
  • InjuryOct 24, 2024Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN ADA 12. ON (B)(6) 2024, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN, MOBILITY, BLEEDING AND INCREASED SENSITIVITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Hyperresponsive to stimuli; Pain

    MAUDE report 20525859 ↗
  • InjuryOct 24, 2024Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN ADA 12. ON 2024-09-17, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN, MOBILITY, BLEEDING AND INCREASED SENSITIVITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Hyperresponsive to stimuli; Pain

    MAUDE report 20525622 ↗
  • InjuryFeb 26, 2024Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2023 IN FDI 45. ON (B)(6) 2023, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 18781222 ↗
  • InjuryFeb 5, 2024Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2020 IN FDI 43. ON (B)(6) 2023, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 18643505 ↗
  • InjuryDec 4, 2023Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2023 IN FDI 22. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED AND IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2023, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 18261981 ↗
  • InjuryNov 13, 2023Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2023 IN ADA 12. DETAILS OF SURGERY: RIDGE AUGMENTATION AND IMMEDIATE EXTRACTION SITE. ON (B)(6) 2023, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III AND INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 18125581 ↗
  • InjuryNov 13, 2023Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2023 IN ADA 12. DETAILS OF SURGERY: RIDGE AUGMENTATION AND IMMEDIATE EXTRACTION SITE. ON (B)(6) 2023, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III AND INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 18121398 ↗
  • InjuryAug 28, 2023Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2022 IN THE PATIENT'S MOUTH. ON (B)(6) 2023, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFECTION, MOBILITY AND INFLAMMATION. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Inflammation; Unspecified Infection

    MAUDE report 17635201 ↗
  • InjuryJul 27, 2023Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2023 IN FDI 24. ON (B)(6) 2023, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFECTION, PAIN, MOBILITY AND SWELLING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Pain; Swelling/ Edema; Unspecified Infection

    MAUDE report 17407586 ↗
  • InjuryJul 7, 2023Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2023 IN ADA 6. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE AND IMMEDIATE EXTRACTION SITE. ON (B)(6) 2023, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III AND INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17281066 ↗
  • InjuryJul 7, 2023Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED 2023-01-27 IN ADA 6. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE AND IMMEDIATE EXTRACTION SITE. ON (B)(6) 2023, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III AND INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17280335 ↗
  • InjuryJun 12, 2023Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2017 IN FDI 41. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE AND IMMEDIATE EXTRACTION SITE. ON (B)(6) 2023, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III AND FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 17103740 ↗

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices