STRAUMANN DENTAL IMPLANT SYSTEM SLActive Titanium TL (REF 033.251S)Recalls & reported problems
Institut Straumann AG · Standard Plus Implant D 4.8mm, RN, SLActive® 8mm, Ti · Class 2 · Implant
All models of STRAUMANN DENTAL IMPLANT SYSTEM (128)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Straumann SLActive dental implant is a screw-shaped device placed into bone during dental surgery. Most of the 314 reports describe the implant losing its bond with the bone or never forming one, and some mention pain, infection or tissue damage. All 314 reports are injury reports, with no deaths, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 315
- Deaths reported
- 0
- Injuries reported
- 315
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 315 adverse event reports that list this exact device (DI 07630031714923), received between Sep 10, 2019 and Sep 18, 2026 — classified as 315 injuries.
The most frequently reported specific device problem is “Loss of Osseointegration” (198 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Loss of Osseointegration198
- Failure to Osseointegrate114
- Adverse Event Without Identified Device or Use Problem7
- Difficult to Remove1
- Fracture1
Patient problems
- No Clinical Signs, Symptoms or Conditions115
- Fibrosis80
- Unspecified Infection69
- Tissue Damage59
- Pain58
- Inflammation33
- Hemorrhage/Blood Loss/Bleeding31
- Swelling/ Edema29
Recent adverse event reports
The 25 most recent of 315 reports, in the reporter's own words as published by the FDA.
InjurySep 18, 2026Loss of Osseointegration
Patient problems: Fibrosis; Inflammation; Unspecified Infection
MAUDE report 50826356 ↗InjuryAug 24, 2026Loss of Osseointegration
Patient problems: Fibrosis; Inflammation; Unspecified Infection
MAUDE report 50414755 ↗InjuryAug 3, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6)2025 in the patient's mouth. On (b)(6)2026, non-osseointegration was verified. No product was returned to the manufacturer. At the event the patient experienced: Peri-implantitis, Pain and Mobility. No further patient complications were reported.
Patient problems: Fibrosis; Pain; Unspecified Infection
MAUDE report 50120395 ↗InjuryJul 30, 2026Loss of Osseointegration
The clinician reports the implant was inserted (b)(6)2018 in FDI 36. On (b)(6)2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50088385 ↗InjuryJul 27, 2026Loss of Osseointegration
The clinician reports the implant was inserted on (b)(6)2007 in ADA 4. On (b)(6)2026, loss of osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. At the event the patient experienced: Peri-implantitis and Mobility. No further patient complications were reported.
Patient problems: Fibrosis; Unspecified Infection
MAUDE report 50041880 ↗InjuryJul 27, 2026Loss of Osseointegration
The clinician reports the implant was inserted 2007 (b)(6) in ADA 4. On 2026 (b)(6), loss of osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. At the event the patient experienced: Peri-implantitis and Mobility. No further patient complications were reported.
Patient problems: Fibrosis; Unspecified Infection
MAUDE report 50039756 ↗InjuryJun 22, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN THE PATIENT'S MOUTH. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED AND IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25584113 ↗InjuryJun 22, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN THE PATIENT'S MOUTH. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED AND IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25584112 ↗InjuryJun 22, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN THE PATIENT'S MOUTH. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED AND IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25584053 ↗InjuryJun 22, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN THE PATIENT'S MOUTH. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED AND IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25584008 ↗InjuryJun 22, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN THE PATIENT'S MOUTH. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED AND IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25584003 ↗InjuryJun 4, 2026Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2008 IN FDI 36. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFLAMMATION. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Inflammation
MAUDE report 25397540 ↗InjuryJun 4, 2026Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2013 IN THE PATIENT'S MOUTH. DETAILS OF SURGERY: SINUS AUGMENTATION. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BRUXISM. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INCREASED SENSITIVITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Hyperresponsive to stimuli
MAUDE report 25397053 ↗InjuryJun 3, 2026Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2021 IN FDI 16. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25380066 ↗InjuryMay 22, 2026Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2013 IN FDI 26. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFECTION, PERI-IMPLANTITIS, PAIN AND SWELLING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Pain; Swelling/ Edema; Unspecified Infection
MAUDE report 25259139 ↗InjuryMay 15, 2026Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2016 IN FDI 46. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fibrosis
MAUDE report 25188374 ↗InjuryApr 29, 2026Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2016 IN FDI 36. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: ABSCESS. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Abscess
MAUDE report 25033971 ↗InjuryJan 30, 2026Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2008 IN FDI 36. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24224811 ↗InjuryJan 26, 2026Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2011 IN FDI 16. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN AND INFLAMMATION. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Inflammation; Pain
MAUDE report 24181672 ↗InjuryJan 8, 2026Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2021 IN FDI 17. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24009832 ↗InjuryDec 1, 2025Loss of Osseointegration; Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2007 IN FDI 26. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY AND INFLAMMATION. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fibrosis; Inflammation
MAUDE report 23683565 ↗InjuryNov 10, 2025Loss of Osseointegration; Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2014 IN ADA 3. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fibrosis
MAUDE report 23516720 ↗InjuryNov 10, 2025Loss of Osseointegration; Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2014 IN ADA 3. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fibrosis
MAUDE report 23514778 ↗InjuryOct 27, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN THE PATIENT'S MOUTH. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS AND MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fibrosis; Unspecified Infection
MAUDE report 23395910 ↗InjuryAug 4, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 35. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN AND BLEEDING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Hemorrhage/Blood Loss/Bleeding; Pain
MAUDE report 22706173 ↗
Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM43,616 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM40,404 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM27,400 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,621 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLA Titanium BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,298 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM23,448 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,320 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,267 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page