TL 04.1MM RN,S, SLACTIVE TIZR 6, 8, 10, 12, 14, 16MM DENTAL IMPLANTS SLActive Roxolid TL (REF 033.531S)Recalls & reported problems

Institut Straumann AG · Standard Implant, D 4.1mm RN, SLActive® 8mm, Roxolid®, Loxim™ · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Straumann SLActive dental implant is placed into bone. Most of the 1,292 reports describe the implant failing to bond with the bone, and some describe pain or infection. The reports record 1,292 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,301
Deaths reported
0
Injuries reported
1,301
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1,301 adverse event reports that list this exact device (DI 07630031715852), received between Sep 9, 2019 and Sep 28, 2026 — classified as 1,301 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (800 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 19 reportsDec 2023: 14 reportsJan 2024: 13 reportsFeb 2024: 11 reportsMar 2024: 17 reportsApr 2024: 21 reportsMay 2024: 18 reportsJun 2024: 27 reportsJul 2024: 15 reportsAug 2024: 13 reportsSep 2024: 22 reportsOct 2024: 20 reportsNov 2024: 23 reportsDec 2024: 11 reportsJan 2025: 13 reportsFeb 2025: 13 reportsMar 2025: 12 reportsApr 2025: 4 reportsMay 2025: 0 reportsJun 2025: 35 reportsJul 2025: 17 reportsAug 2025: 19 reportsSep 2025: 19 reportsOct 2025: 14 reportsNov 2025: 19 reportsDec 2025: 21 reportsJan 2026: 16 reportsFeb 2026: 9 reportsMar 2026: 12 reportsApr 2026: 14 reportsMay 2026: 8 reportsJun 2026: 19 reportsJul 2026: 12 reportsAug 2026: 6 reportsSep 2026: 11 reportsOct 2026: 0 reports20242025202635
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate800
  • Loss of Osseointegration475
  • Adverse Event Without Identified Device or Use Problem22
  • Difficult to Insert5
  • Fracture2
  • Separation Failure2
  • Activation, Positioning or Separation Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions607
  • Fibrosis387
  • Pain204
  • Tissue Damage179
  • Unspecified Infection135
  • Inflammation75
  • Hemorrhage/Blood Loss/Bleeding50
  • Swelling/ Edema31

Recent adverse event reports

The 25 most recent of 1,301 reports, in the reporter's own words as published by the FDA.

  • InjurySep 28, 2026Failure to Osseointegrate

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Pain

    MAUDE report 50922034 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50921963 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50921962 ↗
  • InjurySep 28, 2026Loss of Osseointegration

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50921872 ↗
  • InjurySep 28, 2026Loss of Osseointegration

    Patient problems: Fibrosis

    MAUDE report 50920641 ↗
  • InjurySep 28, 2026Loss of Osseointegration

    Patient problems: Fibrosis

    MAUDE report 50920576 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50920001 ↗
  • InjurySep 17, 2026Loss of Osseointegration

    Patient problems: Fibrosis

    MAUDE report 50807769 ↗
  • InjurySep 10, 2026Failure to Osseointegrate

    Patient problems: Fibrosis; Pain; Swelling/ Edema

    MAUDE report 50719421 ↗
  • InjurySep 10, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2025 in FDI 47. On (b)(6) 2026, loss of osseointegration was verified. Patient presented with bone type III and fair oral hygiene. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50719117 ↗
  • InjurySep 10, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2026 (b)(6) in ADA 19. On 2026 (b)(6), non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Pain, Mobility and Swelling. No further patient complications were reported.

    Patient problems: Fibrosis; Pain; Swelling/ Edema

    MAUDE report 50718641 ↗
  • InjuryAug 27, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (B)(6) 2026 in FDI 26. On (B)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50480621 ↗
  • InjuryAug 24, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50416184 ↗
  • InjuryAug 24, 2026Loss of Osseointegration

    Patient problems: Unspecified Infection

    MAUDE report 50414860 ↗
  • InjuryAug 20, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2019 in the patient's mouth. On (b)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50352848 ↗
  • InjuryAug 6, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2026(b)(6) in FDI 12. On 2026(b)(6), non-osseointegration was verified. Patient presented with inadequate bone quality/quantity and previous bone augmentation. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50163221 ↗
  • InjuryAug 6, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2026(b)(6) in FDI 22. On 2026(b)(6), non-osseointegration was verified. Patient presented with inadequate bone quality/quantity and previous bone augmentation. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50163200 ↗
  • InjuryJul 27, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6)2026 in ADA 30. Details of surgery: Ridge augmentation. On (b)(6)2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50023179 ↗
  • InjuryJul 27, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 30. Details of surgery: Ridge augmentation. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50022615 ↗
  • InjuryJul 27, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2019 in the patient's mouth. On (b)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50010355 ↗
  • InjuryJul 16, 2026Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2024 IN FDI 26. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE IV. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25844949 ↗
  • InjuryJul 16, 2026Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2016 IN FDI 36. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25844564 ↗
  • InjuryJul 13, 2026Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2020 IN FDI 27. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 25803159 ↗
  • InjuryJul 13, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 36. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 25801529 ↗
  • InjuryJul 13, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 26. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 25801489 ↗

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices

All IMPLANT, ENDOSSEOUS, ROOT-FORM devices →