TL 04.1MM RN,S, SLACTIVE TIZR 6, 8, 10, 12, 14, 16MM DENTAL IMPLANTS SLActive Roxolid TL (REF 033.532G)Recalls & reported problems

Institut Straumann AG · Standard Implant, D 4.1 RN, SLActive® 10mm, Roxolid®, guided · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
12
Deaths reported
0
Injuries reported
12
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 07630031715869), received between Jan 13, 2020 and Aug 14, 2026 — classified as 12 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 2 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate7
  • Loss of Osseointegration4
  • Difficult to Remove1

Patient problems

  • Fibrosis7
  • No Clinical Signs, Symptoms or Conditions3
  • Pain3
  • Swelling/ Edema2
  • Unspecified Infection2
  • Failure of Implant1
  • Hemorrhage/Blood Loss/Bleeding1
  • Tissue Damage1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 14, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50259116 ↗
  • InjuryAug 14, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2020 in FDI 24. On (b)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50258852 ↗
  • InjuryNov 10, 2023Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2020 IN FDI 33. ON (B)(6) 2023, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 18115706 ↗
  • InjuryJun 13, 2022Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2022 IN ADA 13. ON (B)(6) 2022, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 14678230 ↗
  • InjuryJun 13, 2022Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2022 IN ADA 13. ON 2022-05-05, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 14677102 ↗
  • InjuryApr 27, 2022Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2021 IN FDI 36. ON (B)(6) 2022, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 14220967 ↗
  • InjuryApr 27, 2022Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2021 IN FDI 36. ON (B)(6) 2022, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 14220955 ↗
  • InjuryNov 22, 2021Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2021 IN FDI 33. ON (B)(6) 2021, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain

    MAUDE report 12855610 ↗
  • InjuryOct 21, 2021Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2018 IN THE PATIENT'S MOUTH. DETAILS OF SURGERY: BONE OVERHEATING. ON (B)(6) 2021, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH POOR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFECTION, PERI-IMPLANTITIS, PAIN, MOBILITY, SWELLING AND BLEEDING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Pain; Swelling/ Edema; Unspecified Infection

    MAUDE report 12672880 ↗
  • InjuryMay 6, 2021Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2015 IN THE PATIENT'S MOUTH. DETAILS OF SURGERY: BONE OVERHEATING. ON (B)(6) 2021, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS, PAIN, MOBILITY AND SWELLING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Pain; Swelling/ Edema; Unspecified Infection

    MAUDE report 11785222 ↗
  • InjuryOct 26, 2020Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2020 IN FDI 44. ON (B)(6) 2020, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: Failure of Implant; No Clinical Signs, Symptoms or Conditions

    MAUDE report 10735526 ↗
  • InjuryJan 13, 2020Difficult to Remove; Difficult to Remove

    THE CLINICIAN REPORTS THAT THE DAY THE IMPLANT WAS PLACED IN FDI 14, FAILURE OCCURRED UPON INSERTION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: Tissue Damage

    MAUDE report 9584112 ↗

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices

All IMPLANT, ENDOSSEOUS, ROOT-FORM devices →

Sources and updates