Skip to content

Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid TL (REF 043.256S)Recalls & reported problems

Institut Straumann AG · Standard Implant, D 3.3mm RN, SLA® 10mm, Roxolid®, Loxim™ · Class 2 · Implant

All models of Straumann Dental Implant System Roxolid SLA Impl. (80)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Straumann Roxolid SLA implant is placed into the jawbone. Most of its 960 reports describe the implant failing to fuse with the bone or losing that fusion, and some describe fractures, pain and infection. The reports mention no deaths and 960 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
967
Deaths reported
0
Injuries reported
967
Recalls naming it
0

Summary

The FDA's MAUDE database contains 967 adverse event reports that list this exact device (DI 07630031718938), received between Aug 29, 2019 and Sep 28, 2026 — classified as 967 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (547 reports).

2 other recalls cover devices cleared under the same FDA submission (K150938); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 9 reportsDec 2023: 10 reportsJan 2024: 12 reportsFeb 2024: 12 reportsMar 2024: 13 reportsApr 2024: 15 reportsMay 2024: 5 reportsJun 2024: 13 reportsJul 2024: 8 reportsAug 2024: 8 reportsSep 2024: 4 reportsOct 2024: 7 reportsNov 2024: 8 reportsDec 2024: 3 reportsJan 2025: 4 reportsFeb 2025: 5 reportsMar 2025: 9 reportsApr 2025: 4 reportsMay 2025: 0 reportsJun 2025: 35 reportsJul 2025: 9 reportsAug 2025: 8 reportsSep 2025: 3 reportsOct 2025: 10 reportsNov 2025: 10 reportsDec 2025: 20 reportsJan 2026: 8 reportsFeb 2026: 18 reportsMar 2026: 12 reportsApr 2026: 10 reportsMay 2026: 7 reportsJun 2026: 6 reportsJul 2026: 10 reportsAug 2026: 11 reportsSep 2026: 8 reportsOct 2026: 0 reports20242025202635
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate547
  • Loss of Osseointegration271
  • Fracture136
  • Adverse Event Without Identified Device or Use Problem15
  • Difficult to Insert2
  • Insufficient Device Problem Information1
  • Separation Failure1
  • Unintended Movement1

Patient problems

  • No Clinical Signs, Symptoms or Conditions440
  • Fibrosis244
  • Pain210
  • Tissue Damage136
  • Unspecified Infection128
  • Inflammation54
  • Hemorrhage/Blood Loss/Bleeding51
  • Failure of Implant38

Recent adverse event reports

The 25 most recent of 967 reports, in the reporter's own words as published by the FDA.

  • InjurySep 28, 2026Loss of Osseointegration

    Patient problems: Fibrosis; Pain; Unspecified Infection

    MAUDE report 50919534 ↗
  • InjurySep 14, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50760273 ↗
  • InjurySep 14, 2026Loss of Osseointegration

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50759117 ↗
  • InjurySep 14, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50759104 ↗
  • InjurySep 10, 2026Loss of Osseointegration

    Patient problems: Fibrosis; Pain; Unspecified Infection

    MAUDE report 50720388 ↗
  • InjurySep 7, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50655404 ↗
  • InjurySep 7, 2026Failure to Osseointegrate

    Patient problems: Fibrosis

    MAUDE report 50655343 ↗
  • InjurySep 7, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50654872 ↗
  • InjuryAug 24, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50416929 ↗
  • InjuryAug 24, 2026Loss of Osseointegration

    Patient problems: Unspecified Infection

    MAUDE report 50416659 ↗
  • InjuryAug 24, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50416652 ↗
  • InjuryAug 24, 2026Failure to Osseointegrate

    Patient problems: Pain

    MAUDE report 50416311 ↗
  • InjuryAug 17, 2026Loss of Osseointegration

    The clinician reports the implant was inserted on (b)(6)2015 in FDI 43. On (b)(6)2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Peri-implantitis and Mobility. No further patient complications were reported.

    Patient problems: Fibrosis; Unspecified Infection

    MAUDE report 50283886 ↗
  • InjuryAug 14, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2026(b)(6) in ADA 29. On 2026(b)(6), non-osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. At the event the patient experienced: Peri-implantitis, Mobility and Inflammation. No further patient complications were reported.

    Patient problems: Fibrosis; Inflammation; Unspecified Infection

    MAUDE report 50258758 ↗
  • InjuryAug 14, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 29. On (b)(6) 2026, non-osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. At the event the patient experienced: Peri-implantitis, Mobility and Inflammation. No further patient complications were reported.

    Patient problems: Fibrosis; Inflammation; Unspecified Infection

    MAUDE report 50258028 ↗
  • InjuryAug 10, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2025 in FDI 25. Details of surgery: Primary stability not achieved and Implant surface not completely covered with bone. On (b)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Peri-implantitis, Pain and Mobility. No further patient complications were reported.

    Patient problems: Fibrosis; Pain; Unspecified Infection

    MAUDE report 50198657 ↗
  • InjuryAug 7, 2026Loss of Osseointegration

    The clinician reports the implant was inserted 2021 (b)(6) in the patient's mouth. On 2026 (b)(6), loss of osseointegration was verified. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50179691 ↗
  • InjuryAug 6, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2026(b)(6) in FDI 34. On 2026(b)(6) non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Peri-implantitis and Mobility. No further patient complications were reported.

    Patient problems: Fibrosis; Unspecified Infection

    MAUDE report 50162834 ↗
  • InjuryAug 3, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6)2026 in FDI 24. On (b)(6)2026, non-osseointegration was verified. Patient presented with bone type IV. No product was returned to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50120273 ↗
  • InjuryJul 30, 2026Loss of Osseointegration

    The clinician reports the implant was inserted 2021(b)(6) in FDI 14. On 2026(b)(6), loss of osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50089077 ↗
  • InjuryJul 17, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN THE PATIENT'S MOUTH. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25859883 ↗
  • InjuryJul 17, 2026Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2008 IN THE PATIENT'S MOUTH. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN AND MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Pain

    MAUDE report 25859833 ↗
  • InjuryJul 17, 2026Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2008 IN THE PATIENT'S MOUTH. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN AND MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Pain

    MAUDE report 25858717 ↗
  • InjuryJul 13, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN ADA 23. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25800406 ↗
  • InjuryJul 6, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN FDI 22. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFECTION. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Unspecified Infection

    MAUDE report 25721168 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K150938. One submission often covers many models, so check the product and lot details in each record.

  • Z-2534-2016 · Jul 14, 2016 · Terminated
    Straumann Bone Level Implant, 4.1mm RC, SLA 10mm, TiZr, NTP Article Number: 021.4510
  • Z-2535-2016 · Jul 14, 2016 · Terminated
    Straumann Bone Level Implant, 4.1mm RC, SLA 12mm, TiZr, NTP Article Number: 021.4512

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices

All IMPLANT, ENDOSSEOUS, ROOT-FORM devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page