ITI DENTAL IMPLANT SYSTEM SLA Titanium TL (REF 043.907S)Recalls & reported problems
Institut Straumann AG · Standard Implant, D 3.3mm RN, SLA® 10mm, Ti, Loxim™ · Class 2 · Implant
All models of ITI DENTAL IMPLANT SYSTEM (60)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Straumann ITI dental implant has 269 reports, with no deaths and 269 injuries. Most describe the implant failing to bond with the jawbone or losing that bond, and some describe damage to the surrounding tissue, pain or infection. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 270
- Deaths reported
- 0
- Injuries reported
- 270
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 270 adverse event reports that list this exact device (DI 07630031719416), received between Sep 13, 2019 and Aug 27, 2026 — classified as 270 injuries.
The most frequently reported specific device problem is “Failure to Osseointegrate” (203 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Osseointegrate203
- Loss of Osseointegration55
- Fracture9
- Adverse Event Without Identified Device or Use Problem3
Patient problems
- No Clinical Signs, Symptoms or Conditions148
- Fibrosis57
- Tissue Damage48
- Pain29
- Unspecified Infection18
- Inflammation6
- Hemorrhage/Blood Loss/Bleeding4
- Swelling/ Edema4
Recent adverse event reports
The 25 most recent of 270 reports, in the reporter's own words as published by the FDA.
InjuryAug 27, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. On (b)(6) 2026, non-osseointegration was verified. Patient presented with poor oral hygiene. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50481762 ↗InjuryAug 27, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50481732 ↗InjuryAug 27, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6) 2026 in FDI 47. On (b)(6) 2026, non-osseointegration was verified. Patient presented with poor oral hygiene. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50481730 ↗InjuryAug 27, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6) 2026 in FDI 45. On (b)(6) 2026, non-osseointegration was verified. Patient presented with fair oral hygiene. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50481727 ↗InjuryAug 24, 2026Failure to Osseointegrate
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50416587 ↗InjuryAug 20, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6) 2026 in FDI 44. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50352420 ↗InjuryAug 6, 2026Failure to Osseointegrate
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50162600 ↗InjuryJun 15, 2026Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE IV, FAIR ORAL HYGIENE AND INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fibrosis
MAUDE report 25506825 ↗InjuryJun 15, 2026Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE IV, FAIR ORAL HYGIENE AND INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fibrosis
MAUDE report 25506809 ↗InjuryJun 15, 2026Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE IV, FAIR ORAL HYGIENE AND INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fibrosis
MAUDE report 25506797 ↗InjuryMay 22, 2026Fracture
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2022 IN THE PATIENT'S MOUTH. DETAILS OF SURGERY: SINUS AUGMENTATION. ON (B)(6) 2026, FRACTURE OF THE IMPLANT WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fibrosis
MAUDE report 25258820 ↗InjuryApr 10, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 46. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24840350 ↗InjuryFeb 26, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN FDI 37. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24459648 ↗InjuryFeb 26, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 34. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24459594 ↗InjuryFeb 26, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN FDI 46. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24459573 ↗InjuryFeb 23, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 46. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24419270 ↗InjuryFeb 23, 2026Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2023 IN FDI 16. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24417689 ↗InjuryFeb 20, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24404307 ↗InjuryJan 15, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN FDI 23. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24083555 ↗InjuryJan 5, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 22. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23964015 ↗InjuryDec 29, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 45. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23913922 ↗InjuryDec 8, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23748138 ↗InjuryDec 8, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2025 IN THE PATIENT'S MOUTH. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23748135 ↗InjuryOct 16, 2025Loss of Osseointegration; Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2003 IN FDI 21. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fibrosis
MAUDE report 23313821 ↗InjuryOct 15, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN FDI 36. ON (B)(6) 2025 , NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23283032 ↗
Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM43,616 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM40,404 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM27,400 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,621 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLA Titanium BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,298 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM23,448 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,320 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,267 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗