MODIFICATION TO ITI DENTAL IMPLANT SYSTEM SLA Titanium TL (REF 043.913S)Recalls & reported problems

Institut Straumann AG · Standard Plus Implant, D 3.3mm RN, SLA® 12mm, Ti, Loxim™ · Class 2 · Implant

All models of MODIFICATION TO ITI DENTAL IMPLANT SYSTEM (42)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Straumann ITI Standard Plus implant appears in 672 adverse event reports, with no deaths and 672 injuries. Reports mostly describe the implant failing to fuse with the jawbone or losing that bond, some describing it not being firmly anchored at placement, with patients reporting pain, fibrous tissue build-up and infection. There are no recalls that name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
676
Deaths reported
0
Injuries reported
676
Recalls naming it
0

Summary

The FDA's MAUDE database contains 676 adverse event reports that list this exact device (DI 07630031719478), received between Sep 9, 2019 and Sep 28, 2026 — classified as 676 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (469 reports).

1 other recall cover devices cleared under the same FDA submission (K033922); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 10 reportsDec 2023: 14 reportsJan 2024: 7 reportsFeb 2024: 12 reportsMar 2024: 4 reportsApr 2024: 10 reportsMay 2024: 10 reportsJun 2024: 6 reportsJul 2024: 17 reportsAug 2024: 5 reportsSep 2024: 14 reportsOct 2024: 14 reportsNov 2024: 13 reportsDec 2024: 9 reportsJan 2025: 9 reportsFeb 2025: 4 reportsMar 2025: 9 reportsApr 2025: 3 reportsMay 2025: 0 reportsJun 2025: 12 reportsJul 2025: 9 reportsAug 2025: 14 reportsSep 2025: 12 reportsOct 2025: 4 reportsNov 2025: 7 reportsDec 2025: 6 reportsJan 2026: 9 reportsFeb 2026: 17 reportsMar 2026: 7 reportsApr 2026: 6 reportsMay 2026: 5 reportsJun 2026: 6 reportsJul 2026: 6 reportsAug 2026: 7 reportsSep 2026: 4 reportsOct 2026: 0 reports20242025202617
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate469
  • Loss of Osseointegration143
  • Fracture55
  • Adverse Event Without Identified Device or Use Problem11
  • Difficult to Insert3
  • Separation Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions398
  • Fibrosis149
  • Pain91
  • Unspecified Infection62
  • Tissue Damage55
  • Hemorrhage/Blood Loss/Bleeding25
  • Inflammation24
  • Swelling/ Edema20

Recent adverse event reports

The 25 most recent of 676 reports, in the reporter's own words as published by the FDA.

  • InjurySep 28, 2026Fracture

    The clinician reports the implant was inserted (b)(6) 2021 in the patient's mouth. On (b)(6) 2026, fracture of the implant was verified. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50922018 ↗
  • InjurySep 18, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50826248 ↗
  • InjurySep 14, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50758596 ↗
  • InjurySep 7, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50654898 ↗
  • InjuryAug 24, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50416484 ↗
  • InjuryAug 24, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50416473 ↗
  • InjuryAug 20, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2025 in FDI 22. Details of surgery: Primary stability not achieved and Implant surface not completely covered with bone. On (b)(6) 2026, non-osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50353048 ↗
  • InjuryAug 13, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2024 in FDI 46. Details of surgery: Primary stability not achieved and Implant surface not completely covered with bone. On (b)(6) 2026, loss of osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50238809 ↗
  • InjuryAug 3, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50119699 ↗
  • InjuryAug 3, 2026Loss of Osseointegration

    Patient problems: Fibrosis; Inflammation; Unspecified Infection

    MAUDE report 50119210 ↗
  • InjuryAug 3, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (B)(6) 2024 in the patient's mouth. On (B)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Pain. No further patient complications were reported.

    Patient problems: Pain

    MAUDE report 50118959 ↗
  • InjuryJul 27, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted on (b)(6) 2026 in FDI 35. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50043508 ↗
  • InjuryJul 27, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 15. On (b)(6) 2026 non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50010860 ↗
  • InjuryJul 13, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN THE PATIENT'S MOUTH. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED AND IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25800019 ↗
  • InjuryJul 13, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN THE PATIENT'S MOUTH. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED AND IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25799994 ↗
  • InjuryJul 13, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN THE PATIENT'S MOUTH. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED AND IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25799992 ↗
  • InjuryJul 13, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN THE PATIENT'S MOUTH. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED AND IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25799991 ↗
  • InjuryJun 29, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25656863 ↗
  • InjuryJun 29, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25656744 ↗
  • InjuryJun 26, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2025 IN FDI 13. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25639818 ↗
  • InjuryJun 26, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 23. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25639792 ↗
  • InjuryJun 25, 2026Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2011 IN THE PATIENT'S MOUTH. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Unspecified Infection

    MAUDE report 25624847 ↗
  • InjuryJun 3, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN FDI 32. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25380542 ↗
  • InjuryMay 22, 2026Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2005 IN FDI 34. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Unspecified Infection

    MAUDE report 25258880 ↗
  • InjuryMay 18, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2024 IN FDI 44. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 25205383 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K033922. One submission often covers many models, so check the product and lot details in each record.

  • Z-1351-2012 · Mar 12, 2012 · Terminated
    Straumann Standard Plus Implant 04.8mm Wide Neck SLA 10mm Article Number: 043.652S. Dental implants are intended for surgical placement in maxillary and/or mandibular arches to p

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices

All IMPLANT, ENDOSSEOUS, ROOT-FORM devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page