Skip to content

ITI DENTAL IMPLANT SYSTEM SLA Titanium TL (REF 043.920S)Recalls & reported problems

Institut Straumann AG · Standard Implant, D 4.1mm RN, SLA® 8mm, Ti, Loxim™ · Class 2 · Implant

All models of ITI DENTAL IMPLANT SYSTEM (60)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Straumann ITI dental implant appears in 881 adverse event reports, with 881 injuries and no deaths. Most describe the implant failing to fuse with the surrounding bone, and some mention the implant moving, pain and tissue damage. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
895
Deaths reported
0
Injuries reported
895
Recalls naming it
0

Summary

The FDA's MAUDE database contains 895 adverse event reports that list this exact device (DI 07630031719546), received between Aug 28, 2019 and Sep 28, 2026 — classified as 895 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (769 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 10 reportsDec 2023: 8 reportsJan 2024: 19 reportsFeb 2024: 20 reportsMar 2024: 9 reportsApr 2024: 23 reportsMay 2024: 17 reportsJun 2024: 6 reportsJul 2024: 16 reportsAug 2024: 14 reportsSep 2024: 12 reportsOct 2024: 8 reportsNov 2024: 8 reportsDec 2024: 11 reportsJan 2025: 19 reportsFeb 2025: 16 reportsMar 2025: 16 reportsApr 2025: 3 reportsMay 2025: 0 reportsJun 2025: 43 reportsJul 2025: 12 reportsAug 2025: 15 reportsSep 2025: 13 reportsOct 2025: 6 reportsNov 2025: 18 reportsDec 2025: 19 reportsJan 2026: 14 reportsFeb 2026: 12 reportsMar 2026: 15 reportsApr 2026: 17 reportsMay 2026: 4 reportsJun 2026: 13 reportsJul 2026: 15 reportsAug 2026: 6 reportsSep 2026: 11 reportsOct 2026: 0 reports20242025202643
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate769
  • Loss of Osseointegration119
  • Fracture3
  • Adverse Event Without Identified Device or Use Problem2
  • Patient-Device Incompatibility1
  • Separation Failure1
  • Unintended Movement1

Patient problems

  • No Clinical Signs, Symptoms or Conditions581
  • Fibrosis158
  • Pain103
  • Tissue Damage86
  • Unspecified Infection38
  • Inflammation17
  • Swelling/ Edema11
  • Failure of Implant8

Recent adverse event reports

The 25 most recent of 895 reports, in the reporter's own words as published by the FDA.

  • InjurySep 28, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50920838 ↗
  • InjurySep 28, 2026Loss of Osseointegration

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50920833 ↗
  • InjurySep 18, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50826231 ↗
  • InjurySep 17, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50807268 ↗
  • InjurySep 14, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2026 (b)(6) in the patient's mouth. Details of surgery: Primary stability not achieved and Implant surface not completely covered with bone. On 2026 (b)(6), non-osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50761094 ↗
  • InjurySep 14, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50761076 ↗
  • InjurySep 14, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50758774 ↗
  • InjurySep 11, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50738936 ↗
  • InjurySep 11, 2026Loss of Osseointegration

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50738259 ↗
  • InjurySep 10, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 15. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50719337 ↗
  • InjurySep 7, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50654015 ↗
  • InjuryAug 24, 2026Failure to Osseointegrate

    Patient problems: Fibrosis

    MAUDE report 50416422 ↗
  • InjuryAug 17, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. Details of surgery: Primary stability not achieved and Implant surface not completely covered with bone. On (b)(6) 2026, non-osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50283116 ↗
  • InjuryAug 14, 2026Failure to Osseointegrate

    Patient problems: Swelling/ Edema

    MAUDE report 50259546 ↗
  • InjuryAug 12, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2025 in FDI 16. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50225415 ↗
  • InjuryAug 10, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2003 in FDI 45. On (b)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Pain and Mobility. No further patient complications were reported.

    Patient problems: Fibrosis; Pain

    MAUDE report 50198996 ↗
  • InjuryAug 3, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50119735 ↗
  • InjuryJul 31, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 36. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50103086 ↗
  • InjuryJul 30, 2026Loss of Osseointegration

    The clinician reports the implant was inserted 2019(b)(6) in FDI 15. On 2026(b)(6), loss of osseointegration was verified. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50088377 ↗
  • InjuryJul 30, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2026(b)(6) in FDI 17. On 2026(b)(6) non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50087970 ↗
  • InjuryJul 27, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (B)(6) 2026 in FDI 16. On (B)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50011102 ↗
  • InjuryJul 16, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 26. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25844632 ↗
  • InjuryJul 16, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 16. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25844608 ↗
  • InjuryJul 16, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2024 IN FDI 27. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25844011 ↗
  • InjuryJul 13, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN THE PATIENT'S MOUTH. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25800010 ↗

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices

All IMPLANT, ENDOSSEOUS, ROOT-FORM devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page