ITI DENTAL IMPLANT SYSTEM SLA Titanium TL (REF 043.921S)Recalls & reported problems

Institut Straumann AG · Standard Implant, D 4.1mm RN, SLA® 10mm, Ti, Loxim™ · Class 2 · Implant

All models of ITI DENTAL IMPLANT SYSTEM (60)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This is the ITI Dental Implant System, a titanium implant placed in the jawbone. The 1,697 reports mostly describe the implant failing to bond with the surrounding bone, and some mention pain, tissue damage, or infection. The records hold 1,697 injury reports and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,704
Deaths reported
0
Injuries reported
1,704
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1,704 adverse event reports that list this exact device (DI 07630031719553), received between Sep 9, 2019 and Sep 14, 2026 — classified as 1,704 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (1,430 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 20 reportsDec 2023: 22 reportsJan 2024: 33 reportsFeb 2024: 28 reportsMar 2024: 16 reportsApr 2024: 28 reportsMay 2024: 27 reportsJun 2024: 3 reportsJul 2024: 20 reportsAug 2024: 28 reportsSep 2024: 27 reportsOct 2024: 14 reportsNov 2024: 13 reportsDec 2024: 17 reportsJan 2025: 20 reportsFeb 2025: 10 reportsMar 2025: 25 reportsApr 2025: 4 reportsMay 2025: 0 reportsJun 2025: 54 reportsJul 2025: 23 reportsAug 2025: 30 reportsSep 2025: 36 reportsOct 2025: 9 reportsNov 2025: 15 reportsDec 2025: 37 reportsJan 2026: 33 reportsFeb 2026: 24 reportsMar 2026: 8 reportsApr 2026: 16 reportsMay 2026: 22 reportsJun 2026: 31 reportsJul 2026: 14 reportsAug 2026: 11 reportsSep 2026: 5 reportsOct 2026: 0 reports20242025202654
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate1,430
  • Loss of Osseointegration266
  • Adverse Event Without Identified Device or Use Problem6
  • Fracture4
  • Separation Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions936
  • Fibrosis338
  • Tissue Damage241
  • Pain234
  • Unspecified Infection101
  • Inflammation69
  • Hemorrhage/Blood Loss/Bleeding30
  • Swelling/ Edema30

Recent adverse event reports

The 25 most recent of 1,704 reports, in the reporter's own words as published by the FDA.

  • InjurySep 14, 2026Failure to Osseointegrate

    Patient problems: Fibrosis

    MAUDE report 50761067 ↗
  • InjurySep 14, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6)2021 in FDI 46. On (b)(6)2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50758530 ↗
  • InjurySep 11, 2026Loss of Osseointegration

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50738756 ↗
  • InjurySep 7, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50654206 ↗
  • InjurySep 7, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50654014 ↗
  • InjuryAug 24, 2026Failure to Osseointegrate

    Patient problems: Fibrosis

    MAUDE report 50416924 ↗
  • InjuryAug 24, 2026Failure to Osseointegrate

    Patient problems: Fibrosis

    MAUDE report 50416419 ↗
  • InjuryAug 24, 2026Loss of Osseointegration

    Patient problems: Fibrosis; Pain

    MAUDE report 50415282 ↗
  • InjuryAug 20, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. Details of surgery: Primary stability not achieved. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50352161 ↗
  • InjuryAug 17, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2026 (b)(6) in FDI 37. On 2026 (b)(6), non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50282032 ↗
  • InjuryAug 17, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 15. Details of surgery: Implant surface not completely covered with bone. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Inflammation. No further patient complications were reported.

    Patient problems: Inflammation

    MAUDE report 50281906 ↗
  • InjuryAug 12, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2025 in FDI 13. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Pain. No further patient complications were reported.

    Patient problems: Pain

    MAUDE report 50225422 ↗
  • InjuryAug 10, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2010 in the patient's mouth. On (b)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Pain and Mobility. No further patient complications were reported.

    Patient problems: Fibrosis; Pain

    MAUDE report 50198391 ↗
  • InjuryAug 6, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50162588 ↗
  • InjuryAug 3, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50119756 ↗
  • InjuryAug 3, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50119302 ↗
  • InjuryJul 31, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2025 in FDI 24. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50103107 ↗
  • InjuryJul 31, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2025 in FDI 46. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50102685 ↗
  • InjuryJul 31, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2025 in FDI 25. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50102625 ↗
  • InjuryJul 29, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted(b)(6)2025 in FDI 36. On (b)(6)2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50072648 ↗
  • InjuryJul 27, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6)2026 in FDI 17. On (b)(6)2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50043190 ↗
  • InjuryJul 27, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50023032 ↗
  • InjuryJul 27, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6)2026 in the patient's mouth. On (b)(6)2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50023031 ↗
  • InjuryJul 17, 2026Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2017 IN FDI 13. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 25858883 ↗
  • InjuryJul 13, 2026Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2023 IN FDI 15. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25800349 ↗

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices

All IMPLANT, ENDOSSEOUS, ROOT-FORM devices →