Skip to content

MODIFICATION TO ITI DENTAL IMPLANT SYSTEM SLA Titanium TL (REF 043.939S)Recalls & reported problems

Institut Straumann AG · Standard Plus Implant, D 4.8mm RN, SLA® 6mm, Ti, Loxim™ · Class 2 · Implant

All models of MODIFICATION TO ITI DENTAL IMPLANT SYSTEM (42)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Straumann ITI dental implant is placed in the jawbone. Its 824 reports are all injuries, with no deaths. Most describe the implant failing to bond with the bone or losing that bond, and some describe tissue hardening around it. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
831
Deaths reported
0
Injuries reported
831
Recalls naming it
0

Summary

The FDA's MAUDE database contains 831 adverse event reports that list this exact device (DI 07630031719737), received between Sep 10, 2019 and Sep 30, 2026 — classified as 831 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (681 reports).

1 other recall cover devices cleared under the same FDA submission (K033922); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 13 reportsDec 2023: 16 reportsJan 2024: 11 reportsFeb 2024: 17 reportsMar 2024: 5 reportsApr 2024: 16 reportsMay 2024: 8 reportsJun 2024: 6 reportsJul 2024: 9 reportsAug 2024: 9 reportsSep 2024: 17 reportsOct 2024: 9 reportsNov 2024: 17 reportsDec 2024: 12 reportsJan 2025: 11 reportsFeb 2025: 4 reportsMar 2025: 15 reportsApr 2025: 7 reportsMay 2025: 0 reportsJun 2025: 49 reportsJul 2025: 13 reportsAug 2025: 21 reportsSep 2025: 12 reportsOct 2025: 12 reportsNov 2025: 23 reportsDec 2025: 18 reportsJan 2026: 24 reportsFeb 2026: 37 reportsMar 2026: 22 reportsApr 2026: 24 reportsMay 2026: 7 reportsJun 2026: 23 reportsJul 2026: 15 reportsAug 2026: 10 reportsSep 2026: 10 reportsOct 2026: 0 reports20242025202649
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate681
  • Loss of Osseointegration144
  • Adverse Event Without Identified Device or Use Problem7
  • Fracture2
  • Difficult to Insert1

Patient problems

  • No Clinical Signs, Symptoms or Conditions581
  • Fibrosis158
  • Pain83
  • Unspecified Infection55
  • Tissue Damage26
  • Inflammation20
  • Swelling/ Edema17
  • Hemorrhage/Blood Loss/Bleeding11

Recent adverse event reports

The 25 most recent of 831 reports, in the reporter's own words as published by the FDA.

  • InjurySep 30, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2025(b)(6) in FDI 17. On 2026(b)(6), non-osseointegration was verified. Patient presented with fair oral hygiene. The device was forwarded to the manufacturer. At the event the patient experienced: Peri-implantitis, Mobility and Bleeding. No further patient complications were reported.

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Unspecified Infection

    MAUDE report 50950227 ↗
  • InjurySep 18, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50825772 ↗
  • InjurySep 17, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50806627 ↗
  • InjurySep 14, 2026Failure to Osseointegrate

    Patient problems: Inflammation

    MAUDE report 50760123 ↗
  • InjurySep 14, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50758733 ↗
  • InjurySep 14, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50758452 ↗
  • InjurySep 14, 2026Loss of Osseointegration

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50758408 ↗
  • InjurySep 14, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 22. Details of surgery: Primary stability not achieved and Implant surface not completely covered with bone. On (b)(6) 2026, non-osseointegration was verified. Patient presented with fair oral hygiene. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50758346 ↗
  • InjurySep 10, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50720326 ↗
  • InjurySep 7, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50653580 ↗
  • InjuryAug 27, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (B)(6) 2026 in FDI 36. On (B)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50480664 ↗
  • InjuryAug 24, 2026Loss of Osseointegration

    Patient problems: Fibrosis

    MAUDE report 50416527 ↗
  • InjuryAug 17, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (B)(6) 2026 in the patient's mouth. Details of surgery: Primary stability not achieved. On (B)(6) 2026 non-osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50283176 ↗
  • InjuryAug 17, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (B)(6) 2026 in FDI 16. On (B)(6) 2026, non-osseointegration was verified. Patient presented with bruxism. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50282657 ↗
  • InjuryAug 17, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50282450 ↗
  • InjuryAug 13, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 36. Details of surgery: Primary stability not achieved and Implant surface not completely covered with bone. On (b)(6) 2026, non-osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50238744 ↗
  • InjuryAug 10, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2025 in FDI 37. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50198903 ↗
  • InjuryAug 10, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 37. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50198853 ↗
  • InjuryAug 6, 2026Failure to Osseointegrate

    Patient problems: Fibrosis

    MAUDE report 50162565 ↗
  • InjuryAug 4, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2026(b)(6) in the patient's mouth. Details of surgery: Implant surface not completely covered with bone. On 2026(b)(6), non-osseointegration was verified. No product was returned to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50132050 ↗
  • InjuryJul 31, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 36. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50103570 ↗
  • InjuryJul 30, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2025 in FDI 27. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50088606 ↗
  • InjuryJul 30, 2026Loss of Osseointegration

    The clinician reports the implant was inserted 2025(b)(6) in FDI 46. On 2026(b)(6), loss of osseointegration was verified. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50088130 ↗
  • InjuryJul 27, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 46. On 2026-06-22, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50043758 ↗
  • InjuryJul 27, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2025-(b)(6) in FDI 17. On 2026-(b)(6), non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50041172 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K033922. One submission often covers many models, so check the product and lot details in each record.

  • Z-1351-2012 · Mar 12, 2012 · Terminated
    Straumann Standard Plus Implant 04.8mm Wide Neck SLA 10mm Article Number: 043.652S. Dental implants are intended for surgical placement in maxillary and/or mandibular arches to p

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices

All IMPLANT, ENDOSSEOUS, ROOT-FORM devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page