MODIFICATION TO ITI DENTAL IMPLANT SYSTEM SLA Titanium TL (REF 043.944S)Recalls & reported problems
Institut Straumann AG · Standard Implant, D 4.8mm WN, SLA® 6mm, Ti, Loxim™ · Class 2 · Implant
All models of MODIFICATION TO ITI DENTAL IMPLANT SYSTEM (42)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Straumann SLA dental implant is placed in the jaw to support a replacement tooth. The 87 reports mostly describe implants that never fused with the jawbone or later lost that bond, sometimes leaving the implant loose in the mouth. Reports also mention scarring, pain, tissue damage, and infections. There were no deaths and 87 injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 86
- Deaths reported
- 0
- Injuries reported
- 86
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 86 adverse event reports that list this exact device (DI 07630031719782), received between Jan 31, 2020 and Sep 17, 2026 — classified as 86 injuries.
The most frequently reported specific device problem is “Failure to Osseointegrate” (51 reports).
1 other recall cover devices cleared under the same FDA submission (K033922); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Osseointegrate51
- Loss of Osseointegration34
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions48
- Fibrosis25
- Pain12
- Tissue Damage6
- Unspecified Infection5
- Abscess2
- Hemorrhage/Blood Loss/Bleeding2
- Inflammation2
Recent adverse event reports
The 25 most recent of 86 reports, in the reporter's own words as published by the FDA.
InjurySep 17, 2026Loss of Osseointegration
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50806860 ↗InjuryJun 15, 2026Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III AND INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fibrosis
MAUDE report 25506800 ↗InjuryMay 28, 2026Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN FDI 27. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fibrosis
MAUDE report 25320640 ↗InjuryApr 27, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 47. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25005336 ↗InjuryApr 7, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN THE PATIENT'S MOUTH. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24804537 ↗InjuryApr 7, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN THE PATIENT'S MOUTH. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24804116 ↗InjuryFeb 16, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN FDI 26. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED AND IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24348028 ↗InjuryJan 29, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 27. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24213849 ↗InjuryJan 8, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24009645 ↗InjuryDec 1, 2025Loss of Osseointegration; Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2010 IN THE PATIENT'S MOUTH. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23685017 ↗InjuryNov 17, 2025Loss of Osseointegration; Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2017 IN FDI 16. ON (B)(6) 2023, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23570603 ↗InjuryNov 10, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23516334 ↗InjurySep 5, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22980905 ↗InjurySep 2, 2025Loss of Osseointegration; Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2023 IN FDI 37. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22933850 ↗InjuryJul 21, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN FDI 23. ON (B)(6) 2024, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22557133 ↗InjuryJul 7, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED AND IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22418717 ↗InjuryJul 7, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN FDI 14. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED AND IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2024, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFECTION. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Unspecified Infection
MAUDE report 22417706 ↗InjuryJun 18, 2025Loss of Osseointegration; Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2023 IN FDI 37. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fibrosis
MAUDE report 22278281 ↗InjuryJun 18, 2025Loss of Osseointegration; Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2017 IN ADA 2. ON (B)(6) 2024, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III AND INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN AND MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fibrosis; Pain
MAUDE report 22272213 ↗InjuryJun 18, 2025Loss of Osseointegration; Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2017 IN ADA 2. ON (B)(6) 2024, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III AND INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN AND MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fibrosis; Pain
MAUDE report 22269976 ↗InjuryJun 18, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2023 IN FDI 26. ON (B)(6) 2024, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22263109 ↗InjuryMar 10, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN THE PATIENT'S MOUTH. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21565738 ↗InjuryJan 23, 2025Loss of Osseointegration; Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2020 IN FDI 37. ON (B)(6) 2024, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21221760 ↗InjuryDec 16, 2024Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN THE PATIENT'S MOUTH. ON (B)(6) 2024, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20942545 ↗InjuryOct 9, 2024Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN THE PATIENT'S MOUTH. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED AND IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2024, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fibrosis
MAUDE report 20409728 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K033922. One submission often covers many models, so check the product and lot details in each record.
- Z-1351-2012 · Mar 12, 2012 · TerminatedStraumann Standard Plus Implant 04.8mm Wide Neck SLA 10mm Article Number: 043.652S. Dental implants are intended for surgical placement in maxillary and/or mandibular arches to p
Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM43,616 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM40,404 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM27,400 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,621 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLA Titanium BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,298 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM23,448 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,320 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,267 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page